Proton beam dosimetry in the presence of magnetic fields using Farmer‐type ionization chambers of different radii
Medical Physics · 2023
DOI: 10.1002/mp.16368 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, 2023. Verfasst von Mathieu Marot und 8 weiteren Autor:innen. Beteiligte Einrichtung: German Cancer Research Center DKFZ, Medical Physics in Radiation Oncology, Heidelberg, Germany. Hauptthemen laut MEDLINE: Protons, Radius. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: DKFZ International PhD Program in Heidelberg, Participating States and from the European Union's Horizon 2020 research and innovation programme, German Federal Ministry of Education and Research: Image Guided Diagnostics and Therapy - New Directions in Intervention.
Abstract (Originalsprache)
Background
Magnetic resonance‐guided proton therapy is promising, as it combines high‐contrast imaging of soft tissue with highly conformal dose delivery. However, proton dosimetry in magnetic fields using ionization chambers is challenging since the dose distribution as well as the detector response are perturbed.
Purpose
This work investigates the effect of the magnetic field on the ionization chamber response, and on the polarity and ion recombination correction factors, which are essential for the implementation of a proton beam dosimetry protocol in the presence of magnetic fields.
Methods
Three Farmer‐type cylindrical ionization chambers, the 30013 with 3 mm inner radius (PTW, Freiburg, Germany) and two custom built chambers “R1” and “R6” with 1 and 6 mm inner radii respectively were placed at the center of an experimental electromagnet (Schwarzbeck Mess – Elektronik, Germany) 2 cm depth of an in‐house developed 3D printed water phantom. The detector response was measured for a 3 × 10 cm 2 field of mono‐energetic protons 221.05 MeV/u for the three chambers, and with an additional proton beam of 157.43 MeV/u for the chamber PTW 30013. The magnetic flux density was varied between 0.1 and 1.0 Tesla in steps of 0.1 Tesla.
Results
At both energies, the ionization chamber PTW 30013 showed a non‐linear response as a function of the magnetic field strength, with a decrease of the ionization chamber response of up to 0.27% ± 0.06% (1 SD) at 0.2 Tesla, followed by a smaller effect at higher magnetic field strength. For the chamber R1, the response decreased slightly with the magnetic field strength up to 0.45% ± 0.12% at 1 Tesla, and for the chamber R6, the response decreased up to 0.54% ± 0.13% at 0.1 Tesla, followed by a plateau up to 0.3 Tesla, and a weaker effect at higher magnetic field strength. The dependence of the polarity and recombination correction factor on the magnetic field was ⩽0.1% for the chamber PTW 30013.
Conclusions
The magnetic field has a small but significant effect on the chamber response in the low magnetic field region for the chamber PTW 30013 and for R6, and in the high magnetic field region for the chamber R1. Corrections may be necessary for ionization chamber measurements, depending on both the chamber volume and the magnetic flux density. No significant effect of the magnetic field on the polarity and recombination correction factor was detected in this work for the ionization chamber PTW 30013.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- German Cancer Research Center DKFZ, Medical Physics in Radiation Oncology, Heidelberg, Germany.
- Faculty of Medicine, University of Heidelberg, Heidelberg, Germany.
- National Center for Radiation Research in Oncology NCRO, Heidelberg Institute for Radiation Oncology HIRO, Heidelberg, Germany.
- Faculty of Physics and Astronomy, University of Heidelberg, Heidelberg, Germany.
- Department of Radiation Oncology, Heidelberg University Hospital, Heidelberg, Germany.
- Heidelberg Ion-Beam Therapy Center HIT, University Hospital Heidelberg, Heidelberg, Germany.
ORCID-Profile: Mathieu Marot, Sonja Surla, Stephan Brons, Steffen Greilich, Christian P Karger, Oliver Jäkel, Lucas N Burigo
Förderung
- DKFZ International PhD Program in Heidelberg
- Participating States and from the European Union's Horizon 2020 research and innovation programme
- German Federal Ministry of Education and Research: Image Guided Diagnostics and Therapy - New Directions in Intervention
- ARTEMIS project
Zitieren
APA
Marot, M., Surla, S., Burke, E., Brons, S., Runz, A., Greilich, S., Karger, C. P., Jäkel, O., & Burigo, L. N. (2023). Proton beam dosimetry in the presence of magnetic fields using Farmer‐type ionization chambers of different radii. Medical Physics. https://doi.org/10.1002/mp.16368
Vancouver
Marot M, Surla S, Burke E, Brons S, Runz A, Greilich S, et al. Proton beam dosimetry in the presence of magnetic fields using Farmer‐type ionization chambers of different radii. Medical Physics. 2023. doi:10.1002/mp.16368
BibTeX
@article{Marot2023,
author = {Marot, Mathieu and Surla, Sonja and Burke, Elisa and Brons, Stephan and Runz, Armin and Greilich, Steffen and Karger, Christian P and Jäkel, Oliver and Burigo, Lucas N},
title = {Proton beam dosimetry in the presence of magnetic fields using Farmer‐type ionization chambers of different radii},
journal = {Medical Physics},
year = {2023},
doi = {10.1002/mp.16368}
}
EUVIMED