Simulation studies of a full‐ring, CZT SPECT system for whole‐body imaging of 99m Tc and 177 Lu

Yoonsuk Huh, Javier Caravaca, Jaehyuk Kim, Yonggang Cui, Qiu Huang, Grant Gullberg, Youngho Seo

Medical Physics · 2023-06 · 50 (6) · 3726-3737

DOI: 10.1002/mp.16360 ↗

At a glance

Journal article in Medical Physics, published online on 2023-03-22. Written by Yoonsuk Huh and 6 co-authors. Institution: Department of Radiology and Biomedical Imaging, University of California, San Francisco, California, USA. Major MEDLINE topics: Cadmium, Whole Body Imaging. Cited 14 times so far (Crossref); 31 references. The full text is freely available in PubMed Central. Funded by: NIBIB NIH HHS, NHLBI NIH HHS, National Institute of Biomedical Imaging and Bioengineering.

AbstractSubjectsDati bibliograficiCiteRelated studies

Abstract (original language)

Background

Single photon emission computed tomography (SPECT) is an imaging modality that has demonstrated its utility in a number of clinical indications. Despite this progress, a high sensitivity, high spatial resolution, multi‐tracer SPECT with a large field of view suitable for whole‐body imaging of a broad range of radiotracers for theranostics is not available.

Purpose

With the goal of filling this technological gap, we have designed a cadmium zinc telluride (CZT) full‐ring SPECT scanner instrumented with a broad‐energy tungsten collimator. The final purpose is to provide a multi‐tracer solution for brain and whole‐body imaging. Our static SPECT does not rely on the dual‐ and the triple‐head rotational SPECT standard paradigm, enabling a larger effective area in each scan to increase the sensitivity. We provide a demonstration of the performance of our design using a realistic model of our detector with simulated body‐sized phantoms filled with 99m Tc and 177 Lu.

Methods

We create a realistic model of our detector by using a combination of a Geant4 Application for Tomographic Emission (GATE) Monte Carlo simulation and a finite element model for the CZT response, accounting for low‐energy tail effects in CZT that affects the sensitivity and the scatter correction. We implement a modified dual‐energy‐window scatter correction adapted for CZT. Other corrections for attenuation, detector and collimator response, and detector gaps and edges are also included. The images are reconstructed using the maximum‐likelihood expectation‐maximization. Detector and reconstruction performance are characterized with point sources, Derenzo phantoms, and a body‐sized National Electrical Manufacturers Association (NEMA) Image Quality (IQ) phantom for both 99m Tc and 177 Lu.

Results

Our SPECT design can resolve 7.9 mm rods for 99m Tc (140 keV) and 9.5 mm for 177 Lu (208 keV) in a hot‐rod Derenzo phantom with a 3‐min exposure and reach an image contrast of 78% for 99m Tc and 57% for 177 Lu using the NEMA IQ phantom with a 6‐min exposure. Our modified scatter correction shows an improved contrast‐recovery ratio compared to a standard correction.

Conclusions

In this paper, we demonstrate the good performance of our design for whole‐body imaging purposes. This adds to our previous demonstration of improved qualitative and quantitative 99m Tc imaging of brain perfusion and 123 I imaging of dopamine transport with respect to state‐of‐the‐art NaI dual‐head cameras. We show that our design provides similar IQ and contrast to the commercial full‐ring SPECT VERITON for 99m Tc. Regarding 177 Lu imaging of the 208 keV emissions, our design provides similar contrast to that of other state‐of‐the‐art SPECTs with a significant reduction in exposure. The high sensitivity and extended energy range up to 250 keV makes our SPECT design a promising alternative for clinical imaging and theranostics of emerging radionuclides.

Subjects

MeSH terms (MEDLINE)

CadmiumWhole Body ImagingTomography, Emission-Computed, Single-PhotonPhantoms, ImagingIodine RadioisotopesZincTellurium

Author keywords

CZTMonte Carlo simulationSPECTfinite element methodfull-ring

Substances

CadmiumIodine RadioisotopesZincCdZnTeTelluriumIodine-123

Authors and institutions

  • Department of Radiology and Biomedical Imaging, University of California, San Francisco, California, USA.
  • Princess Margaret Cancer Centre, University Health Network, Toronto, Ontario, Canada.
  • Department of Nonproliferation and National Security, Brookhaven National Laboratory, Upton, New York, USA.
  • School of Biomedical Engineering, Shanghai Jiao Tong University, Shanghai, China.
  • Molecular Biophysics and Integrated Bioimaging Division, Lawrence Berkeley National Laboratory, Berkeley, California, USA.
  • Joint Graduate Group in Bioengineering, University of California, San Francisco, Berkeley, California, USA.

Funding

  • NIBIB NIH HHS
  • NHLBI NIH HHS
  • National Institute of Biomedical Imaging and Bioengineering
  • National Heart, Lung, and Blood Institute

Cite

APA

Huh, Y., Caravaca, J., Kim, J., Cui, Y., Huang, Q., Gullberg, G., & Seo, Y. (2023). Simulation studies of a full‐ring, CZT SPECT system for whole‐body imaging of 99m Tc and 177 Lu. Medical Physics, 50(6), 3726-3737. https://doi.org/10.1002/mp.16360

Vancouver

Huh Y, Caravaca J, Kim J, Cui Y, Huang Q, Gullberg G, et al. Simulation studies of a full‐ring, CZT SPECT system for whole‐body imaging of 99m Tc and 177 Lu. Medical Physics. 2023;50(6):3726-3737. doi:10.1002/mp.16360

BibTeX
@article{Huh2023,
  author = {Huh, Yoonsuk and Caravaca, Javier and Kim, Jaehyuk and Cui, Yonggang and Huang, Qiu and Gullberg, Grant and Seo, Youngho},
  title = {Simulation studies of a full‐ring, CZT SPECT system for whole‐body imaging of 99m Tc and 177 Lu},
  journal = {Medical Physics},
  year = {2023},
  volume = {50},
  number = {6},
  pages = {3726--3737},
  doi = {10.1002/mp.16360}
}
Start for free →