General principle and optimization of magneto‐acousto‐electrical tomography based on image distortion evaluation

Peixia Li, Wei Chen, Gepu Guo, Juan Tu, Dong Zhang, Qingyu Ma

Medical Physics · 2023

DOI: 10.1002/mp.16317 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, 2023. Verfasst von Peixia Li und 5 weiteren Autor:innen. Beteiligte Einrichtung: School of Computer and Electronic Information, Nanjing Normal University, Nanjing, China. Hauptthemen laut MEDLINE: Algorithms, Tomography. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 37 Literaturangaben. Gefördert von: National Natural Science Foundation of China, Natural Science Foundation of Jiangsu Province, Innovation Project of Postgraduates in Jiangsu Province.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

As a novel non‐invasive multi‐physics imaging methodology, the magneto‐acousto‐electrical (MAE) technology is capable of detecting electric conductivity changes for biological tissues, exhibiting prosperous perspectives in medical applications. However, the acoustic beam was often simplified to a straight line or a focused one, being perpendicular to layered boundaries of tissues in previous studies. Linear‐scanning measurements were carried out to reconstruct B‐mode MAE images for layered models without considering the radiation pattern of transducers. Obvious image distortions in both shape and brightness were observed in experiments without any reasonable explanation.

Purpose

This study aims to establish a general physical model for MAE measurements and solve the problem of B‐mode image distortion, and hence provide theoretical and technical supports for the improvement of MAE imaging in practical applications.

Methods

By considering the radiation pattern of actual transducers and the inclined angle of electric conductivity boundaries, a general principal of MAE measurements applicable for objects of arbitrary shapes is proposed based on the theories of acoustic radiation, Hall Effect and electrical detection. The influences of inclined conductivity boundaries and transducer directivities are numerically analyzed with Matlab programming and also demonstrated by experimental measurements. To evaluate the degree of B‐mode image distortion, the deformation length (3 dB amplitude decrease) of approximate straight lines for a circular model is defined as L = d tan( β m /2), with d and β m being the measurement distance and the half radiation angle of the main‐lobe, respectively. The rotary‐scanning‐based MAE tomography (MAET) is employed to reduce the image distortion, and the rotation angle step is further optimized based on the criterion of the boundary radius fluctuation coefficient

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

AlgorithmsTomographyElectricityElectric ConductivityAcousticsPhantoms, Imaging

Stichwörter der Autor:innen

deformation lengthimage distortionincident anglemagneto-acousto-electrical tomographyrotation angle step optimization

Autor:innen und Institutionen

  • School of Computer and Electronic Information, Nanjing Normal University, Nanjing, China.
  • Institute of Acoustics, Nanjing University, Nanjing, China.

Förderung

  • National Natural Science Foundation of China
  • Natural Science Foundation of Jiangsu Province
  • Innovation Project of Postgraduates in Jiangsu Province

Zitieren

APA

Li, P., Chen, W., Guo, G., Tu, J., Zhang, D., & Ma, Q. (2023). General principle and optimization of magneto‐acousto‐electrical tomography based on image distortion evaluation. Medical Physics. https://doi.org/10.1002/mp.16317

Vancouver

Li P, Chen W, Guo G, Tu J, Zhang D, Ma Q. General principle and optimization of magneto‐acousto‐electrical tomography based on image distortion evaluation. Medical Physics. 2023. doi:10.1002/mp.16317

BibTeX
@article{Li2023,
  author = {Li, Peixia and Chen, Wei and Guo, Gepu and Tu, Juan and Zhang, Dong and Ma, Qingyu},
  title = {General principle and optimization of magneto‐acousto‐electrical tomography based on image distortion evaluation},
  journal = {Medical Physics},
  year = {2023},
  doi = {10.1002/mp.16317}
}
Kostenlos starten →