Phantom evaluation of electrical conductivity mapping by MRI : Comparison to vector network analyzer measurements and spatial resolution assessment
Magnetic Resonance in Medicine · 2024
DOI: 10.1002/mrm.30009 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2024. Verfasst von Zhongzheng He und 6 weiteren Autor:innen. Beteiligte Einrichtung: IADI U1254, INSERM and Université de Lorraine, Nancy, France. Hauptthemen laut MEDLINE: Brain, Image Processing, Computer-Assisted. Die Arbeit wurde bisher 11-mal zitiert (Crossref) und nennt 50 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: CPER IT2MP, Agence Nationale de la Recherche, FEDER (European Regional Development Fund).
Abstract (Originalsprache)
Purpose
To evaluate the performance of various MR electrical properties tomography (MR‐EPT) methods at 3 T in terms of absolute quantification and spatial resolution limit for electrical conductivity.
Methods
Absolute quantification as well as spatial resolution performance were evaluated on homogeneous phantoms and a phantom with holes of different sizes, respectively. Ground‐truth conductivities were measured with an open‐ended coaxial probe connected to a vector network analyzer (VNA). Four widely used MR‐EPT reconstruction methods were investigated: phase‐based Helmholtz (PB), phase‐based convection‐reaction (PB‐cr), image‐based (IB), and generalized‐image‐based (GIB). These methods were compared using the same complex images from a 1 mm‐isotropic UTE sequence. Alternative transceive phase acquisition sequences were also compared in PB and PB‐cr.
Results
In large homogeneous phantoms, all methods showed a strong correlation with ground truth conductivities ( r > 0.99); however, GIB was the best in terms of accuracy, spatial uniformity, and robustness to boundary artifacts. In the resolution phantom, the normalized root‐mean‐squared error of all methods grew rapidly (>0.40) when the hole size was below 10 mm, with simplified methods (PB and IB), or below 5 mm, with generalized methods (PB‐cr and GIB).
Conclusion
VNA measurements are essential to assess the accuracy of MR‐EPT. In this study, all tested MR‐EPT methods correlated strongly with the VNA measurements. The UTE sequence is recommended for MR‐EPT, with the GIB method providing good accuracy for structures down to 5 mm. Structures below 5 mm may still be detected in the conductivity maps, but with significantly lower accuracy.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- IADI U1254, INSERM and Université de Lorraine, Nancy, France.
- BioSerenity, Paris, France.
- Siemens Healthcare SAS, Saint Denis, France.
- CIC-IT 1433, INSERM, Université de Lorraine and CHRU Nancy, Nancy, France.
ORCID-Profile: Zhongzheng He, Paul Soullié, Freddy Odille
Förderung
- CPER IT2MP
- Agence Nationale de la Recherche
- FEDER (European Regional Development Fund)
- European Regional Development Fund
Zitieren
APA
He, Z., Lefebvre, P. M., Soullié, P., Doguet, M., Ambarki, K., Chen, B., & Odille, F. (2024). Phantom evaluation of electrical conductivity mapping by MRI : Comparison to vector network analyzer measurements and spatial resolution assessment. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30009
Vancouver
He Z, Lefebvre PM, Soullié P, Doguet M, Ambarki K, Chen B, et al. Phantom evaluation of electrical conductivity mapping by MRI : Comparison to vector network analyzer measurements and spatial resolution assessment. Magnetic Resonance in Medicine. 2024. doi:10.1002/mrm.30009
BibTeX
@article{He2024,
author = {He, Zhongzheng and Lefebvre, Pauline M and Soullié, Paul and Doguet, Martin and Ambarki, Khalid and Chen, Bailiang and Odille, Freddy},
title = {Phantom evaluation of electrical conductivity mapping by MRI : Comparison to vector network analyzer measurements and spatial resolution assessment},
journal = {Magnetic Resonance in Medicine},
year = {2024},
doi = {10.1002/mrm.30009}
}
EUVIMED