Design and performance evaluation of a slit‐slat camera for 2D prompt gamma imaging in proton therapy monitoring: A Monte Carlo simulation study

Etesam Malekzadeh, Hossein Rajabi, Mohammad Ali Tajik-Mansoury, Pouya Sabouri, Elisa Fiorina, Faraz Kalantari

Medical Physics · 2023

DOI: 10.1002/mp.16259 ↗

At a glance

Journal article in Medical Physics, 2023. Written by Etesam Malekzadeh and 5 co-authors. Institution: Department of Medical Physics, Faculty of Medical Sciences, Tarbiat Modares University, Tehran, Iran. Major MEDLINE topics: Proton Therapy. Cited 2 times so far (Crossref); 60 references.

AbstraktSubjectsBibliografické údajeCiteRelated studies

Abstract (original language)

Purpose

We investigated the design of a prompt gamma camera for real‐time dose delivery verification and the partial mitigation of range uncertainties.

Methods

A slit slat (SS) camera was optimized using the trade‐off between the signal‐to‐noise ratio and spatial resolution. Then, using the GATE Monte Carlo package, the camera performances were estimated by means of target shifts, beam position quantification, changing the camera distance from the beam, and air cavity inserting. A homogeneous PMMA phantom and the air gaps induced PMMA phantom were used. The air gaps ranged from 5 mm to 30 mm by 5 mm increments were positioned in the middle of the beam range. To reduce the simulation time, phase space scoring was used. The batch method with five realizations was used for stochastic error calculations.

Results

The system's detection efficiency was PGs/proton) for a 10 × 20 cm 2 detector (source‐to‐collimator distance = 15.0 cm). Axial and transaxial resolutions were 23 mm and 18 mm, respectively. The SS camera estimated the range as 69.0 ± 3.4 (relative stochastic error 1‐sigma is 5%) and 67.6 ± 1.8 mm (2.6%) for the real range of 67.0 mm for 10 7 and 10 8 protons of 100 MeV, respectively. Considering 160 MeV, these values are 155.5 ± 3.1 (2%) and 152.2 ± 2.0 mm (1.3%) for the real range of 152.0 mm for 10 7 and 10 8 protons, respectively. Considering phantom shift, for a 100 MeV beam, the precision of the quantification (1‐sigma) in the axial and lateral phantom shift estimation is 2.6 mm and 1 mm, respectively. Accordingly, the axial and lateral quantification precisions were 1.3 mm and 1 mm for a 160 MeV beam, respectively. Furthermore, the quantification of an air gap formulated as , where and gap real are the estimated and real air gap, respectively. The precision of the air gap quantification is 1.6 mm (1 sigma). Moreover, 2D PG images show the trajectory of the proton beam through the phantom.

Conclusion

The proposed slit‐slat imaging systems can potentially provide a real‐time, in‐vivo, and non‐invasive treatment monitoring method for proton therapy.

Subjects

MeSH terms (MEDLINE)

Proton TherapyProtonsMonte Carlo MethodPolymethyl MethacrylateDiagnostic ImagingPhantoms, Imaging

Author keywords

Monte Carloproton therapyrange verification

Substances

ProtonsPolymethyl Methacrylate

Authors and institutions

  • Department of Medical Physics, Faculty of Medical Sciences, Tarbiat Modares University, Tehran, Iran.
  • Biomedical Engineering and Medical Physics Department, School of Medicine, Shahid Beheshti University of Medical Sciences, Tehran, Iran.
  • Department of Radiation Oncology, University of Arkansas for Medical Sciences, Little Rock, Arkansas, USA.
  • National Institute of Nuclear Physics INFN, Section of Torino, Torino, Italy.
  • Clinical Department, Fondazione CNAO, Pavia, Italy.

Cite

APA

Malekzadeh, E., Rajabi, H., Tajik-Mansoury, M. A., Sabouri, P., Fiorina, E., & Kalantari, F. (2023). Design and performance evaluation of a slit‐slat camera for 2D prompt gamma imaging in proton therapy monitoring: A Monte Carlo simulation study. Medical Physics. https://doi.org/10.1002/mp.16259

Vancouver

Malekzadeh E, Rajabi H, Tajik-Mansoury MA, Sabouri P, Fiorina E, Kalantari F. Design and performance evaluation of a slit‐slat camera for 2D prompt gamma imaging in proton therapy monitoring: A Monte Carlo simulation study. Medical Physics. 2023. doi:10.1002/mp.16259

BibTeX
@article{Malekzadeh2023,
  author = {Malekzadeh, Etesam and Rajabi, Hossein and Tajik-Mansoury, Mohammad Ali and Sabouri, Pouya and Fiorina, Elisa and Kalantari, Faraz},
  title = {Design and performance evaluation of a slit‐slat camera for 2D prompt gamma imaging in proton therapy monitoring: A Monte Carlo simulation study},
  journal = {Medical Physics},
  year = {2023},
  doi = {10.1002/mp.16259}
}
Start for free →