EUVIMEDSkopri EUVIMED
Open access · CC BY-NC-ND 4.0

Global sensitivity study for irreversible electroporation: Towards treatment planning under uncertainty

Prashanth Lakshmi Narasimhan, Zoi Tokoutsi, Davide Baroli, Marco Baragona, Karen Veroy, Ralph Maessen, Andreas Ritter

Medical Physics · 2023

DOI: 10.1002/mp.16220 ↗

At a glance

Journal article in Medical Physics, 2023. Written by Prashanth Lakshmi Narasimhan and 6 co-authors. Institution: Philips Research, Eindhoven, AE, The Netherlands. Major MEDLINE topics: Liver Neoplasms, Ablation Techniques. Cited 7 times so far (Crossref); 69 references. The full text is freely available (open access, CC BY-NC-ND 4.0). Funded by: H2020 Marie Skodowska-Curie Actions.

AstrattSubjectsDettalji biblijografiċiCiteRelated studies

Abstract (original language)

Background

Electroporation‐based cancer treatments are minimally invasive, nonthermal interventional techniques that leverage cell permeabilization to ablate the target tumor. However, the amount of permeabilization is susceptible to the numerous uncertainties during treatment, such as patient‐specific variations in the tissue, type of the tumor, and the resolution of imaging equipment. These uncertainties can reduce the extent of ablation in the tissue, thereby affecting the effectiveness of the treatment.

Purpose

The aim of this work is to understand the effect of these treatment uncertainties on the treatment outcome for irreversible electroporation (IRE) in the case of colorectal liver metastasis (CRLM). Understanding the nature and extent of these effects can help us identify the influential treatment parameters and build better models for predicting the treatment outcome.

Methods

This is an in silico study using a static computational model with a custom applicator design, spherical tissue, and tumor geometry. A nonlinear electrical conductivity, dependent on the local electric field, is considered. Morris analysis is used to identify the influential treatment parameters on the treatment outcome. Seven treatment parameters pertaining to the relative tumor location with respect to the applicator, the tumor growth pattern, and the electrical conductivity of tissue are analyzed. The treatment outcome is measured in terms of the relative tumor ablation with respect to the target ablation volume and total ablation volume.

Results

The Morris analysis was performed with 800 model evaluations, sampled from the seven dimensional input parameter space. Electrical properties of the tissue, especially the electrical conductivity of the tumor before ablation, were found to be the most influential parameter for relative tumor ablation and total ablation volume. This parameter was found to be about 4–15 times more influential than the least influential parameter, depending on the tumor size. The tumor border configuration was identified as the least important parameter for treatment effectiveness. The most desired treatment outcome is obtained by a combination of high healthy liver conductivity and low tumor conductivity. This information can be used to tackle worst‐case scenarios in treatment planning. Finally, when the safety margins used in the clinical applications are accounted for, the effects of uncertainties in the treatment parameters reduce drastically.

Conclusions

The results of this work can be used to create an efficient surrogate estimator for uncertainty quantification in the treatment outcome, that can be utilized in optimal real‐time treatment planning solutions.

Subjects

MeSH terms (MEDLINE)

Liver NeoplasmsAblation TechniquesHumansUncertaintyElectroporationElectric Conductivity

Author keywords

Morris methodglobal sensitivity analysisirreversible electroporationtreatment planning

Authors and institutions

  • Philips Research, Eindhoven, AE, The Netherlands.
  • Centre for Analysis, Scientific Computing, and Applications, Eindhoven University of Technology, Eindhoven, AZ, The Netherlands.
  • Euler Institute (Instituto Eulero) and Cardiocentro Ticino, Faculty of Informatics, Universitá della Svizzera italiana, Viganello-Lugano, Switzerland.
  • Department of Diagnostic and Interventional Radiology, University Hospital RWTH Aachen, Aachen, Germany.

Funding

  • H2020 Marie Skodowska-Curie Actions

Cite

APA

Lakshmi Narasimhan, P., Tokoutsi, Z., Baroli, D., Baragona, M., Veroy, K., Maessen, R., & Ritter, A. (2023). Global sensitivity study for irreversible electroporation: Towards treatment planning under uncertainty. Medical Physics. https://doi.org/10.1002/mp.16220

Vancouver

Lakshmi Narasimhan P, Tokoutsi Z, Baroli D, Baragona M, Veroy K, Maessen R, et al. Global sensitivity study for irreversible electroporation: Towards treatment planning under uncertainty. Medical Physics. 2023. doi:10.1002/mp.16220

BibTeX
@article{LakshmiNarasimhan2023,
  author = {Lakshmi Narasimhan, Prashanth and Tokoutsi, Zoi and Baroli, Davide and Baragona, Marco and Veroy, Karen and Maessen, Ralph and Ritter, Andreas},
  title = {Global sensitivity study for irreversible electroporation: Towards treatment planning under uncertainty},
  journal = {Medical Physics},
  year = {2023},
  doi = {10.1002/mp.16220}
}
Start for free →