EUVIMEDEUVIMED entdecken
Crossref

Flexible real‐time skin dosimeter based on a thin‐film copper indium gallium selenide solar cell for electron radiation therapy

Dong-Seok Shin, Tae-Ho Kim, Jeong-Eun Rah, Se Byeong Lee, Young Kyung Lim, Jonghwi Jeong, Haksoo Kim, Dongho Shin, Jaeman Son

Medical Physics · 2023

DOI: 10.1002/mp.16191 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, 2023. Verfasst von Dong-Seok Shin und 8 weiteren Autor:innen. Beteiligte Einrichtung: Proton Therapy Center, National Cancer Center, Goyang, Republic of Korea. Hauptthemen laut MEDLINE: Radiation Dosimeters, Copper. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 57 Literaturangaben. Gefördert von: National Research Foundation of Korea, Korean National Cancer Center Fund.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Various dosimeters have been proposed for skin dosimetry in electron radiotherapy. However, one main drawback of these skin dosimeters is their lack of flexibility, which could make accurate dose measurements challenging due to air gaps between a curved patient surface and dosimeter. Therefore, the purpose of this study is to suggest a novel flexible skin dosimeter based on a thin‐film copper indium gallium selenide (CIGS) solar cell, and to evaluate its dosimetric characteristics.

Methods

The CIGS solar cell dosimeter consisted of (a) a customized thin‐film CIGS solar cell and (b) a data acquisition (DAQ) system. The CIGS solar cell with a thickness of 0.33 mm was customized to a size of 10 × 10 mm 2 . This customized solar cell plays a role in converting therapeutic electron radiation into electrical signals. The DAQ system was composed of a voltage amplifier with a gain of 1000, a voltage input module, a DAQ chassis, and an in‐house software. This system converted the electrical analog signals (from solar cell) to digital signals with a sampling rate of ≤50 kHz and then quantified/visualized the digital signals in real time. We quantified the linearity/ sampling rate effect/dose rate dependence/energy dependence/field size output factor/reproducibility/curvature/bending recoverability/angular dependence of the CIGS solar cell dosimeter in therapeutic electron beams. To evaluate clinical feasibility, we measured the skin point doses by attaching the CIGS solar cell to an anthropomorphic phantom surface (for forehead, mouth, and thorax). The CIGS‐measured doses were compared with calculated doses (by treatment planning system) and measured doses (by optically stimulated luminescent dosimeter).

Results

The normalized signals of the solar cell dosimeter increased linearly as the delivered dose increased. The gradient of the linearly fitted line was 1.00 with an R‐square of 0.9999. The sampling rates (2, 10, and 50 kHz) of the solar cell dosimeter showed good performance even at low doses (

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiation DosimetersCopperHumansIndiumElectronsReproducibility of ResultsRadiometryGalliumSeleniumBenzoquinones

Stichwörter der Autor:innen

copper indium gallium selenide solar cellelectron radiotherapyflexible dosimeterskin dosimeter

Substanzen

CopperIndiumgallium selenideGalliumSeleniumDAQBenzoquinones

Autor:innen und Institutionen

  • Proton Therapy Center, National Cancer Center, Goyang, Republic of Korea.
  • Department of Radiation Oncology, Myongji Hospital, Goyang, Republic of Korea.
  • Department of Radiation Oncology, Seoul National University Hospital, Seoul, Republic of Korea.

Förderung

  • National Research Foundation of Korea
  • Korean National Cancer Center Fund

Zitieren

APA

Shin, D. S., Kim, T. H., Rah, J. E., Lee, S. B., Lim, Y. K., Jeong, J., Kim, H., Shin, D., & Son, J. (2023). Flexible real‐time skin dosimeter based on a thin‐film copper indium gallium selenide solar cell for electron radiation therapy. Medical Physics. https://doi.org/10.1002/mp.16191

Vancouver

Shin DS, Kim TH, Rah JE, Lee SB, Lim YK, Jeong J, et al. Flexible real‐time skin dosimeter based on a thin‐film copper indium gallium selenide solar cell for electron radiation therapy. Medical Physics. 2023. doi:10.1002/mp.16191

BibTeX
@article{Shin2023,
  author = {Shin, Dong-Seok and Kim, Tae-Ho and Rah, Jeong-Eun and Lee, Se Byeong and Lim, Young Kyung and Jeong, Jonghwi and Kim, Haksoo and Shin, Dongho and Son, Jaeman},
  title = {Flexible real‐time skin dosimeter based on a thin‐film copper indium gallium selenide solar cell for electron radiation therapy},
  journal = {Medical Physics},
  year = {2023},
  doi = {10.1002/mp.16191}
}
Kostenlos starten →