Technical note: The contribution of internal bremsstrahlung to the 90 Y dose point kernel
Medical Physics · 2023
DOI: 10.1002/mp.16171 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, 2023. Verfasst von Lucrezia Auditore und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical and Dental Sciences and of Morphologic and Functional Imaging (BIOMORF), University of Messina, Italy. Hauptthemen laut MEDLINE: Radiometry, Nuclear Medicine. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 38 Literaturangaben.
Abstract (Originalsprache)
Background
Internal dosimetry has an increasing role in the planning and verification of nuclear medicine therapies with radiopharmaceuticals. Dose Point Kernels (DPKs), quantifying the energy deposition all around a point source, in a homogenous medium, are extensively used for 3D dosimetry and nowadays are mostly evaluated by Monte Carlo (MC) simulation. To our knowledge, DPK for beta emitters is estimated neglecting the continuous photon emission due to the Internal Bremsstrahlung (IB), whose contribution to the absorbed dose can be relevant beyond the maximum range of betas, as evidenced in recent works.
Purpose
Aim of this study was to investigate and quantify, by means of MC simulations, the contribution of IB photons to DPK calculated for 90 Y and provide the updated 90 Y DPK.
Methods
The overall radiation due to the decay of a 90 Y point source, placed at the centre of concentric water shells of increasing radii from 0.02 cm to 20 cm, was simulated with GAMOS, including the IB source term whose spectral distribution was described by an analytical model. Energy deposition was scored in the shells as a function of the distance from the source, R , and DPK was estimated in terms of the scaled absorbed dose fraction, F(R/X 90 ) , where X 90 is the range within which the beta particles deposit 90% of their energy.
Results
A comparison between the two simulated absorbed dose distributions, calculated with or without IB, clearly shows that the latter (incomplete) choice is consistent with the findings of other Authors and systematically underestimates the absorbed dose imparted to the tissue. 90 Y DPK values currently used are underestimated by 20%‐34% for R>2X 90 .
Conclusions
The revised values provided in this work suggest that the inclusion of IB emission in DPK evaluations is advisable for pure beta emitters.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Biomedical and Dental Sciences and of Morphologic and Functional Imaging (BIOMORF), University of Messina, Italy.
- INFN, National Institute for Nuclear Physics, Section of Catania, Italy.
- Department of Mathematical and Computer Science, Physical Sciences and Earth Sciences (MIFT), University of Messina, Italy.
Zitieren
APA
Auditore, L., Amato, E., Pistone, D., & Italiano, A. (2023). Technical note: The contribution of internal bremsstrahlung to the 90 Y dose point kernel. Medical Physics. https://doi.org/10.1002/mp.16171
Vancouver
Auditore L, Amato E, Pistone D, Italiano A. Technical note: The contribution of internal bremsstrahlung to the 90 Y dose point kernel. Medical Physics. 2023. doi:10.1002/mp.16171
BibTeX
@article{Auditore2023,
author = {Auditore, Lucrezia and Amato, Ernesto and Pistone, Daniele and Italiano, Antonio},
title = {Technical note: The contribution of internal bremsstrahlung to the 90 Y dose point kernel},
journal = {Medical Physics},
year = {2023},
doi = {10.1002/mp.16171}
}
EUVIMED