Fachartikel in der Zeitschrift Medical Physics, online erschienen am 26.12.2022. Verfasst von Ahtesham Ullah Khan und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Physics, School of Medicine and Public Health, University of Wisconsin-Madison, Madison, Wisconsin, USA. Hauptthemen laut MEDLINE: Linear Energy Transfer, Radiometry. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 53 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: UW-Madison, Advanced Computing Initiative, Wisconsin Alumni Research Foundation.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

In a recent study, we reported beam quality correction factors, f Q , in carbon ion beams using Monte Carlo (MC) methods for a cylindrical and a parallel‐plate ionization chamber (IC). A non‐negligible perturbation effect was observed; however, the magnitude of the perturbation correction due to the specific IC subcomponents was not included. Furthermore, the stopping power data presented in the International Commission on Radiation Units and Measurements (ICRU) report 73 were used, whereas the latest stopping power data have been reported in the ICRU report 90.

Purpose

The aim of this study was to extend our previous work by computing f Q correction factors using the ICRU 90 stopping power data and by reporting IC‐specific perturbation correction factors. Possible energy or linear energy transfer (LET) dependence of the f Q correction factor was investigated by simulating both pristine beams and spread‐out Bragg peaks (SOBPs).

Methods

The TOol for PArticle Simulation (TOPAS)/GEANT4 MC code was used in this study. A 30 × 30 × 50 cm 3 water phantom was simulated with a uniform 10 × 10 cm 2 parallel beam incident on the surface. A Farmer‐type cylindrical IC (Exradin A12) and two parallel‐plate ICs (Exradin P11 and A11) were simulated in TOPAS using the manufacturer‐provided geometrical drawings. The f Q correction factor was calculated in pristine carbon ion beams in the 150–450 MeV/u energy range at 2 cm depth and in the middle of the flat region of four SOBPs. The k Q correction factor was calculated by simulating the f Qo correction factor in a 60 Co beam at 5 cm depth. The perturbation correction factors due to the presence of the individual IC subcomponents, such as the displacement effect in the air cavity, collecting electrode, chamber wall, and chamber stem, were calculated at 2 cm depth for monoenergetic beams only. Additionally, the mean dose‐averaged and track‐averaged LET was calculated at the depths at which the f Q was calculated.

Results

The ICRU 90 f Q correction factors were reported. The p dis correction factor was found to be significant for the cylindrical IC with magnitudes up to 1.70%. The individual perturbation corrections for the parallel‐plate ICs were 1.00% and displayed differences up to 0.87% between pristine beams and SOBPs. Conversely, the f Q for A11 and A12 displayed a minimal energy dependence of

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Linear Energy TransferRadiometryRelative Biological EffectivenessCarbonMonte Carlo MethodCobalt Radioisotopes

Stichwörter der Autor:innen

Ion chambersLET dependenceMonte Carlobeam qualitycarbon ion therapycomputational dosimetryfQ correction factorkQ correction factorperturbation correction factor

Substanzen

CarbonCobalt-60Cobalt Radioisotopes

Autor:innen und Institutionen

  • Department of Medical Physics, School of Medicine and Public Health, University of Wisconsin-Madison, Madison, Wisconsin, USA.

Förderung

  • UW-Madison
  • Advanced Computing Initiative
  • Wisconsin Alumni Research Foundation
  • Wisconsin Institutes for Discovery
  • US Department of Energy's Office of Science
  • National Science Foundation

Zitieren

APA

Khan, A. U., Nelson, N. P., Culberson, W. S., & DeWerd, L. A. (2023). On the perturbation effect and LET dependence of beam quality correction factors in carbon ion beams. Medical Physics, 50(2), 1105-1120. https://doi.org/10.1002/mp.16089

Vancouver

Khan AU, Nelson NP, Culberson WS, DeWerd LA. On the perturbation effect and LET dependence of beam quality correction factors in carbon ion beams. Medical Physics. 2023;50(2):1105-1120. doi:10.1002/mp.16089

BibTeX
@article{Khan2023,
  author = {Khan, Ahtesham Ullah and Nelson, Nicholas P and Culberson, Wesley S and DeWerd, Larry A},
  title = {On the perturbation effect and LET dependence of beam quality correction factors in carbon ion beams},
  journal = {Medical Physics},
  year = {2023},
  volume = {50},
  number = {2},
  pages = {1105--1120},
  doi = {10.1002/mp.16089}
}
Kostenlos starten →