Real‐time laser speckle contrast imaging measurement during normothermic machine perfusion in pretransplant kidney assessment
Lasers in Surgery and Medicine · 2023-10 · 55 (8) · 784-793
DOI: 10.1002/lsm.23715 ↗Kurz gefasst
Fachartikel in der Zeitschrift Lasers in Surgery and Medicine, online erschienen am 09.08.2023. Verfasst von Yitian Fang und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Surgery, Division of HPB and Transplant Surgery, Transplant Institute, Erasmus Medical Center, Rotterdam, the Netherlands. Hauptthemen laut MEDLINE: Laser Speckle Contrast Imaging, Kidney. Die Arbeit wurde bisher 17-mal zitiert (Crossref) und nennt 23 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Objectives
Normothermic machine perfusion (NMP) provides a platform for pre‐transplant kidney quality assessment that is essential for the use of marginal donor kidneys. Laser speckle contrast imaging (LSCI) presents distinct advantages as a real‐time and noncontact imaging technique for measuring microcirculation. In this study, we aimed to assess the value of LSCI in visualizing renal cortical perfusion and investigate the additional value of dual‐side LSCI measurements compared to single aspect measurement during NMP.
Methods
Porcine kidneys were obtained from a slaughterhouse and then underwent NMP. LSCI was used to measure one‐sided cortical perfusion in the first 100 min of NMP. Thereafter, the inferior renal artery branch was occluded to induce partial ischemia and LSCI measurements on both ventral and dorsal sides were performed.
Results
LSCI fluxes correlated linearly with the renal blood flow ( R 2 = 0.90, p < 0.001). After renal artery branch occlusion, absence of renal cortical perfusion could be visualized and semiquantified by LSCI. The overall ischemic area percentage of the ventral and dorsal sides was comparable (median interquartile range [IQR], 38 [24−43]% vs. 29 [17−46]%, p = 0.43), but heterogenous patterns between the two aspects were observed. There was a significant difference in oxygen consumption (mean ± standard deviation [SD], 2.57 ± 0.63 vs. 1.83 ± 0.49 mLO 2 /min/100 g, p < 0.001), urine output (median [IQR], 1.3 [1.1−1.7] vs. 0.8 [0.6−1.3] mL/min, p < 0.05), lactate dehydrogenase (mean ± SD, 768 ± 370 vs. 905 ± 401 U/L, p < 0.05) and AST (mean ± SD, 352 ± 285 vs. 462 ± 383 U/L, p < 0.01) before and after renal artery occlusion, while no significant difference was found in creatinine clearance, fractional excretion of sodium, total sodium reabsorption and histological damage.
Conclusions
LSCI fluxes correlated linearly with renal blood flow during NMP. Renal cortical microcirculation and absent perfusion can be visualized and semiquantified by LSCI. It provides a relative understanding of perfusion levels, allowing for a qualitative comparison between regions in the kidney. Dual‐side LSCI measurements are of added value compared to single aspect measurement and renal function markers.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Surgery, Division of HPB and Transplant Surgery, Transplant Institute, Erasmus Medical Center, Rotterdam, the Netherlands.
- Department of Biomechanical Engineering, Delft University of Technology, Delft, the Netherlands.
- Department of Cardiology, Erasmus Medical Center, Rotterdam, the Netherlands.
- Department of Pathology and Clinical Bioinformatics, Erasmus Medical Center, Rotterdam, the Netherlands.
- Institute of Experimental and Systems Biology, RWTH Aachen University, Aachen, Germany.
ORCID-Profile: Yitian Fang
Zitieren
APA
Fang, Y., van Ooijen, L., Ambagtsheer, G., Nikolaev, A. V., Clahsen-van Groningen, M. C., Dankelman, J., de Bruin, R. W. F., & Minnee, R. C. (2023). Real‐time laser speckle contrast imaging measurement during normothermic machine perfusion in pretransplant kidney assessment. Lasers in Surgery and Medicine, 55(8), 784-793. https://doi.org/10.1002/lsm.23715
Vancouver
Fang Y, van Ooijen L, Ambagtsheer G, Nikolaev AV, Clahsen-van Groningen MC, Dankelman J, et al. Real‐time laser speckle contrast imaging measurement during normothermic machine perfusion in pretransplant kidney assessment. Lasers in Surgery and Medicine. 2023;55(8):784-793. doi:10.1002/lsm.23715
BibTeX
@article{Fang2023,
author = {Fang, Yitian and van Ooijen, Lisanne and Ambagtsheer, Gisela and Nikolaev, Anton V and Clahsen-van Groningen, Marian C and Dankelman, Jenny and de Bruin, Ron W F and Minnee, Robert C},
title = {Real‐time laser speckle contrast imaging measurement during normothermic machine perfusion in pretransplant kidney assessment},
journal = {Lasers in Surgery and Medicine},
year = {2023},
volume = {55},
number = {8},
pages = {784--793},
doi = {10.1002/lsm.23715}
}
EUVIMED