Optimizing CI Systems for Better Recognition of Soft Speech —the Concept of Broad‐Range Mapping
Laryngoscope Investigative Otolaryngology · 2025-10 · 10 (5)
DOI: 10.1002/lio2.70273 ↗At a glance
Journal article in Laryngoscope Investigative Otolaryngology, published online on 2025-10-07. Written by Matthias Hey and 1 co-authors. Institution: Department of Otorhinolaryngology, Head and Neck Surgery, Audiology University Hospital Schleswig-Holstein (UKSH), Campus Kiel Kiel Germany. Cited 2 times so far (Crossref); 60 references. The full text is freely available (open access, CC BY 4.0). Funded by: Cochlear.
Abstract (original language)
Objectives
The individual mapping of cochlear implants (CIs) aims to optimize the user's speech understanding. Recent investigations have shown the importance of soft speech: (1) According to Datalog studies, a large proportion of speech components lies in the range below 60 dB, and (2) soft speech represents a separate category in CI outcome, in addition to supra‐threshold speech and speech in noise. Soft‐speech understanding can be influenced by optimizing T‐values or by global parameters (loudness growth and TSPL in the Nucleus system). This study focussed on improving soft speech below 60 dB by optimizing loudness growth.
Methods
Speech understanding with varying loudness growth in the speech processor CP11 (Cochlear Ltd.) was compared in 20 experienced adult CI users. The mean soft‐speech score based on monosyllabic words at 40 and 50 dB was introduced for quantification.
Results
Six of the 20 patients studied showed significant individual improvement for soft speech when loudness growth was optimized, while none showed a significant decrease under quiet or noisy test conditions.
Conclusion
Actual CI systems offer a broad loudness range of speech understanding. In addition to suprathreshold speech understanding, additional attention should be paid to soft speech, and the result should therefore be confirmed by speech audiometry at low levels. Levels of Evidence 2.
Subjects
Author keywords
Authors and institutions
- Department of Otorhinolaryngology, Head and Neck Surgery, Audiology University Hospital Schleswig-Holstein (UKSH), Campus Kiel Kiel Germany.
- Cochlear Hannover Germany.
ORCID profiles: Matthias Hey
Funding
- Cochlear
Cite
APA
Hey, M., & Hocke, T. (2025). Optimizing CI Systems for Better Recognition of Soft Speech —the Concept of Broad‐Range Mapping. Laryngoscope Investigative Otolaryngology, 10(5). https://doi.org/10.1002/lio2.70273
Vancouver
Hey M, Hocke T. Optimizing CI Systems for Better Recognition of Soft Speech —the Concept of Broad‐Range Mapping. Laryngoscope Investigative Otolaryngology. 2025;10(5). doi:10.1002/lio2.70273
BibTeX
@article{Hey2025,
author = {Hey, Matthias and Hocke, Thomas},
title = {Optimizing CI Systems for Better Recognition of Soft Speech —the Concept of Broad‐Range Mapping},
journal = {Laryngoscope Investigative Otolaryngology},
year = {2025},
volume = {10},
number = {5},
doi = {10.1002/lio2.70273}
}
EUVIMED