EUVIMEDEUVIMED entdecken
Multizentrische StudieOpen Access · CC BY 4.0

Multi‐Frequency Electrocochleography Results in Fewer Drop Alarms During Cochlear Implant Insertion

Mana Espahbodi, Ali Syed, Alex Tu, Michael D Seidman iD, Christopher J Danner, Kyle P Allen, Loren J Bartels, Syed F Ahsan, Courtney C J Voelker, Nicholas L Deep iD, Jacob B Hunter iD, Rebecca C Chiffer und 16 weitere

The Laryngoscope · 2026

DOI: 10.1002/lary.70592 ↗

Kurz gefasst

Multizentrische Studie in der Zeitschrift The Laryngoscope, 2026. Verfasst von Mana Espahbodi und 27 weiteren Autor:innen. Beteiligte Einrichtung: Department of Otolaryngology-Head and Neck Surgery, University of Utah, Salt Lake City, Utah, USA. Hauptthemen laut MEDLINE: Cochlear Implantation, Audiometry, Evoked Response, Cochlear Implants, Hearing Loss, Sensorineural. Die Arbeit nennt 23 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Advanced Bionics, LLC.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

To evaluate intracochlear electrocochleography (ECochG) amplitude parameters during cochlear implantation (CI) using a novel multi‐frequency ECochG algorithm.

Methods

A multi‐institutional, prospective cohort study was performed at 18 high‐volume CI centers. The inclusion criteria were adults with sensorineural hearing loss and audiometric thresholds of ≤ 90 dB hearing level at 500 Hz undergoing CI with Advanced Bionics (Valencia, CA) Ultra 3D devices between 2024 and 2025. ECochG recordings were performed with simultaneous multi‐frequency stimulation of four frequencies between 125 and 4000 Hz during cochlear implant insertion. Concurrent multi‐frequency recording allowed extraction of amplitude and phase of each frequency individually. Post hoc analysis was performed to determine the difference in the number of drop alarms between single‐ and multi‐frequency ECochG. An ECochG amplitude drop of 6 dB was defined as a drop alarm. Insertion track patterns were compared between single‐ and multi‐frequency ECochG.

Results

One hundred ninety‐five ears were included. Mean number of drop alarms for the single‐frequency algorithm was 1.72 (95% CI: 1.52, 1.92; median 1) compared to 0.42 (95% CI: 0.31, 0.53; median 0) for multi‐frequency; p < 0.001. The number of Type C patterns (rise in amplitude during insertion followed by a drop) decreased with the multi‐frequency ECochG algorithm compared to the single‐frequency ECochG algorithm. The number of Type D patterns (no‐response) decreased, indicating that multi‐frequency ECochG generated more responses across the cochlea than single‐frequency ECochG.

Conclusions

A novel multi‐frequency ECochG algorithm during CI is associated with fewer drop alarms and altered insertion track patterns, which may provide a more accurate assessment of the cochlear microenvironment. Level of Evidence 3.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Cochlear ImplantationAudiometry, Evoked ResponseCochlear ImplantsHearing Loss, SensorineuralClinical AlarmsHumansProspective StudiesAlgorithmsFemale

Stichwörter der Autor:innen

ECochGcochlear implantationmulti‐frequency electrocochleography

Autor:innen und Institutionen

  • Department of Otolaryngology-Head and Neck Surgery, University of Utah, Salt Lake City, Utah, USA.
  • Department of Otolaryngology & Communication Sciences, Medical College of Wisconsin, Milwaukee, Wisconsin, USA.
  • AdventHealth; University of Central Florida, Orlando, Florida, USA.
  • Tampa Bay General Hospital, Tampa Bay, Florida, USA.
  • Head & Neck Surgery Department, Southern California Kaiser Permanente Medical Group, Anaheim, California, USA.
  • Otology/Neurotology-Lateral Skull Base Surgery, Pacific Neuroscience Institute, Los Angeles, California, USA.

ORCID-Profile: Michael D Seidman, Nicholas L Deep, Jacob B Hunter, Hamid R Djalilian, Katrina R Stidham, Richard K Gurgel, Karl W Doerfer, Michael S Harris

Förderung

  • Advanced Bionics, LLC

Zitieren

APA

Espahbodi, M., Syed, A., Tu, A., Seidman, M. D., Danner, C. J., Allen, K. P., Bartels, L. J., Ahsan, S. F., Voelker, C. C. J., Deep, N. L., Hunter, J. B., Chiffer, R. C., Michaelides, E. M., Elrakhawy, M., Miller, M. W., Lin, H. W., Djalilian, H. R., Vivas, E. X., Angeli, S. I., …, & Harris, M. S. (2026). Multi‐Frequency Electrocochleography Results in Fewer Drop Alarms During Cochlear Implant Insertion. The Laryngoscope. https://doi.org/10.1002/lary.70592

Vancouver

Espahbodi M, Syed A, Tu A, Seidman MD, Danner CJ, Allen KP, et al. Multi‐Frequency Electrocochleography Results in Fewer Drop Alarms During Cochlear Implant Insertion. The Laryngoscope. 2026. doi:10.1002/lary.70592

BibTeX
@article{Espahbodi2026,
  author = {Espahbodi, Mana and Syed, Ali and Tu, Alex and Seidman, Michael D and Danner, Christopher J and Allen, Kyle P and Bartels, Loren J and Ahsan, Syed F and Voelker, Courtney C J and Deep, Nicholas L and Hunter, Jacob B and Chiffer, Rebecca C and Michaelides, Elias M and Elrakhawy, Mohamed and Miller, Matthew W and Lin, Harrison W and Djalilian, Hamid R and Vivas, Esther X and Angeli, Simon I and Hoa, Michael and Kim, H Jeffrey and Widick, Mark H and Stidham, Katrina R and Patel, Neil S and Gurgel, Richard K and Doerfer, Karl W and Koka, Kanthaiah and Harris, Michael S},
  title = {Multi‐Frequency Electrocochleography Results in Fewer Drop Alarms During Cochlear Implant Insertion},
  journal = {The Laryngoscope},
  year = {2026},
  doi = {10.1002/lary.70592}
}
Kostenlos starten →