Validation of Non‐Soundproof Tablet Audiometry in Pediatrics: Comparison With Traditional Audiometry
The Laryngoscope · 2026-09 · 136 (9) · 4056-4063
DOI: 10.1002/lary.70514 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift The Laryngoscope, online erschienen am 19.03.2026. Verfasst von Amy E Ensing und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Otolaryngology-Head and Neck Surgery, Washington University in St. Louis, St. Louis, Missouri, USA. Hauptthemen laut MEDLINE: Audiometry, Pure-Tone, Computers, Handheld, Hearing Loss, Audiometry. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 27 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NIDCD NIH HHS, NIH HHS, Denardo Education and Research Foundation.
Abstract (Originalsprache)
Objectives
Portable, tablet‐based audiometry has been proposed as a diagnostic tool for patients unable to access traditional sound booth audiometry. However, the validity of tablet‐based audiometry for diagnostic use in pediatric patients has not been established. The objective of this study is to validate self‐administered tablet‐based diagnostic audiometry in a pediatric clinic setting.
Methods
Children 4 to 17.9 years‐old were recruited from a pediatric audiology clinic. Tablet audiometry was performed with tablet‐based software, SHOEBOX Audiometry (Ottawa, ON) at six frequencies (250–8000 Hz). Hearing thresholds from traditional pure tone audiometry in a sound booth were used as a gold‐standard comparison. Hearing loss was defined as ≥ 1 hearing threshold of > 20 dB.
Results
120 children underwent tablet‐audiometry, though not all were able to complete the test. In 948 individual frequencies tested, the mean threshold difference was −0.95 dB (95% CI –1.5 to −0.42 dB) between traditional and tablet audiogram. Most tablet‐assessed thresholds (92%, n = 873) were within 10 dB of corresponding traditional audiometry thresholds. Tablet audiometry demonstrated 92% sensitivity (95% CI 84%–97%) and 85% specificity (95% CI 77%–92%) for diagnosis of hearing loss. Test–retest reliability of tablet audiometry was high, with 93% (55/59) of test–retest hearing threshold differences falling within ±5 dB.
Conclusion
Tablet audiometry is valid and reliable for children aged 5–17 years who can successfully complete testing in a non‐soundproof clinic setting. It may serve as a tool for remote hearing testing, with potential applications in schools, clinics, and research. Level of Evidence 2
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Otolaryngology-Head and Neck Surgery, Washington University in St. Louis, St. Louis, Missouri, USA.
- School of Medicine, Tulane University, New Orleans, Louisiana, USA.
ORCID-Profile: Amy E Ensing, David S Lee, Judith E C Lieu
Förderung
- NIDCD NIH HHS
- NIH HHS
- Denardo Education and Research Foundation
- National Institutes of Health
Zitieren
APA
Ensing, A. E., Mueller, L., Lee, D. S., Kallogjeri, D., & Lieu, J. E. C. (2026). Validation of Non‐Soundproof Tablet Audiometry in Pediatrics: Comparison With Traditional Audiometry. The Laryngoscope, 136(9), 4056-4063. https://doi.org/10.1002/lary.70514
Vancouver
Ensing AE, Mueller L, Lee DS, Kallogjeri D, Lieu JEC. Validation of Non‐Soundproof Tablet Audiometry in Pediatrics: Comparison With Traditional Audiometry. The Laryngoscope. 2026;136(9):4056-4063. doi:10.1002/lary.70514
BibTeX
@article{Ensing2026,
author = {Ensing, Amy E and Mueller, Lauren and Lee, David S and Kallogjeri, Dorina and Lieu, Judith E C},
title = {Validation of Non‐Soundproof Tablet Audiometry in Pediatrics: Comparison With Traditional Audiometry},
journal = {The Laryngoscope},
year = {2026},
volume = {136},
number = {9},
pages = {4056--4063},
doi = {10.1002/lary.70514}
}
EUVIMED