Tracheal Anastomosis Leaks Across Suture Techniques and Tensions: A Biomechanical Ex Vivo Study
The Laryngoscope · 2024
DOI: 10.1002/lary.31652 ↗Kurz gefasst
Fachartikel in der Zeitschrift The Laryngoscope, 2024. Verfasst von Alex Hu und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Otolaryngology-Head and Neck Surgery, University of Cincinnati College of Medicine, Cincinnati, Ohio, U.S.A. Hauptthemen laut MEDLINE: Suture Techniques, Trachea, Anastomosis, Surgical, Anastomotic Leak. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 15 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Background
Anastomotic leak after tracheal resection may occur while coughing in the early postoperative period. We investigated the varying effects of suturing technique, stretch, and tension on anastomotic leaks during simulated coughs.
Methods
End‐to‐end anastomoses were performed using continuous or interrupted sutures on excised porcine larynges. Tracheas were secured to a pressurized system simulating cough forces, submerged in a water bath, and stretched to 1, 2, and 3 cm above baseline. Peak pressure, incomplete cough generation, and observed leakages were recorded. Parameters were analyzed using Analysis of Variance (ANOVA), multiple linear regression, and logistic regression modeling.
Results
Peak tension ( B = −0.660, p < 0.001) and stretch lengths ( B = −0.329, p = 0.006) were associated with variance in peak pressure ( R 2 = 0.77, F (3,294) = 8.182, p < 0.001). Incomplete coughs increased with higher peak tension (odds ratio [OR] = 15.627, p < 0.001) and stretching to 3 cm above baseline (OR = 4.335, p < 0.007). Similarly, leak occurrences, primarily from the posterior tracheal wall, increased with higher peak tension (OR = 1.787, p < 0.001) and stretching to 3 cm (OR = 2.613, p = 0.017). No significance was identified with suturing technique.
Conclusion
Interrupted and continuous suture techniques do not differ in anastomotic strength during simulated coughs. Increased peak tracheal tension is associated with a weaker anastomosis, and tracheal stretch to 3 cm was associated with a weaker anastomosis. Our study supports the commonly held clinical belief that, to create a stronger anastomosis, tension should be minimized, and particular attention should be placed at the posterior tracheal wall during closure. Level of Evidence NA, Benchtop study Laryngoscope , 134:4998–5005, 2024
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Otolaryngology-Head and Neck Surgery, University of Cincinnati College of Medicine, Cincinnati, Ohio, U.S.A.
- Department of Otolaryngology-Head and Neck Surgery, University of South Florida Morsani College of Medicine, Tampa, Florida, U.S.A.
ORCID-Profile: Yael Bensoussan, Liran Oren, Gregory R Dion
Zitieren
APA
Hu, A., Awad, N., Lee, J., Patel, R., Cohen, O., Bensoussan, Y., Oren, L., Friedman, A. D., & Dion, G. R. (2024). Tracheal Anastomosis Leaks Across Suture Techniques and Tensions: A Biomechanical Ex Vivo Study. The Laryngoscope. https://doi.org/10.1002/lary.31652
Vancouver
Hu A, Awad N, Lee J, Patel R, Cohen O, Bensoussan Y, et al. Tracheal Anastomosis Leaks Across Suture Techniques and Tensions: A Biomechanical Ex Vivo Study. The Laryngoscope. 2024. doi:10.1002/lary.31652
BibTeX
@article{Hu2024,
author = {Hu, Alex and Awad, Nour and Lee, Joonsue and Patel, Rahul and Cohen, Ohad and Bensoussan, Yael and Oren, Liran and Friedman, Aaron D and Dion, Gregory R},
title = {Tracheal Anastomosis Leaks Across Suture Techniques and Tensions: A Biomechanical Ex Vivo Study},
journal = {The Laryngoscope},
year = {2024},
doi = {10.1002/lary.31652}
}
EUVIMED