A new cementless tibial baseplate in kinematically aligned total knee arthroplasty has low migration and stabilizes at 3 months: A cohort study using radiostereometric analysis

Rubiana Monteiro, Ali Mohammadi, Alexander J Nedopil, Stephen M Howell, Maury L Hull iD

Knee Surgery, Sports Traumatology, Arthroscopy · 17.08.2026

DOI: 10.1002/ksa.70572 ↗

Kurz gefasst

Kohortenstudie in der Zeitschrift Knee Surgery, Sports Traumatology, Arthroscopy, online erschienen am 17.08.2026. Verfasst von Rubiana Monteiro und 4 weiteren Autor:innen. Beteiligte Einrichtung: KinAlign Orthopaedics, Sacramento, California, USA. Die Arbeit nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Introduction of new components in total knee arthroplasty (TKA) should be assessed for long‐term stability. The primary objective was to determine whether migration of a new cementless tibial baseplate in kinematically aligned (KA) TKA is below the 6‐month threshold where baseplates have a low incidence of long‐term loosening. Other objectives were to determine when stability occurred, whether stability was associated with varus angles in knee and limb alignments, and whether a recut of the tibia affected migration.

Methods

New cementless components were implanted in 45 patients, of which 38% had a tibial recut. Baseplate migration was assessed using model‐based radiostereometric analysis with radiographs obtained post‐operatively and at 6 weeks, 3 months and 6 months. Mean maximum total point motion (MTPM) at 6 months was compared to a 6‐month threshold where baseplates have low incidence of long‐term loosening. An equivalence analysis determined when stability occurred. Correlation analyses related MTPM at 6 months to the medial proximal tibial angle (MPTA) and the hip–knee–ankle angle (HKAA). A two‐sample t test determined whether mean MTPM differed between recut and non‐recut groups.

Results

Mean 6‐month MTPM of 0.93 mm was significantly below the 6‐month stability threshold of 1.15 mm ( p = 0.015), and stability was achieved by 3 months. There was no significant relationship between MTPM and MPTA ( ρ = 0.050, p = 0.77) or HKAA ( ρ = 0.060, p = 0.72). Mean MTPM did not differ between recut (0.85 ± 0.56 mm) and non‐recut (0.97 ± 0.64 mm) groups ( p = 0.57).

Conclusion

The new cementless tibial baseplate has migration significantly below the 6‐month threshold where baseplates have a low incidence of long‐term loosening and migration stabilizes early at 3 months. Varus alignment of the baseplate and the limb as well as tibial recuts were not associated with increased migration. Level of Evidence N/A.

Schlagwörter

Stichwörter der Autor:innen

hip–knee–ankle anglemaximum total point motionmedial proximal tibial anglesix degree‐of‐freedom motion

Autor:innen und Institutionen

  • KinAlign Orthopaedics, Sacramento, California, USA.
  • Department of Biomedical Engineering, University of California Davis, Davis, California, USA.
  • Department of Orthopaedic Surgery, Konig-Ludwig-Haus, University of Wurzburg, Wurzburg, Germany.
  • Institute for Orthopaedic Surgery and Research, College of Medicine, California Northstate University, Elk Grove, California, USA.
  • Department of Mechanical Engineering, University of California Davis, Davis, California, USA.

ORCID-Profile: Maury L Hull

Zitieren

APA

Monteiro, R., Mohammadi, A., Nedopil, A. J., Howell, S. M., & Hull, M. L. (2026). A new cementless tibial baseplate in kinematically aligned total knee arthroplasty has low migration and stabilizes at 3 months: A cohort study using radiostereometric analysis. Knee Surgery, Sports Traumatology, Arthroscopy. https://doi.org/10.1002/ksa.70572

Vancouver

Monteiro R, Mohammadi A, Nedopil AJ, Howell SM, Hull ML. A new cementless tibial baseplate in kinematically aligned total knee arthroplasty has low migration and stabilizes at 3 months: A cohort study using radiostereometric analysis. Knee Surgery, Sports Traumatology, Arthroscopy. 2026. doi:10.1002/ksa.70572

BibTeX
@article{Monteiro2026,
  author = {Monteiro, Rubiana and Mohammadi, Ali and Nedopil, Alexander J and Howell, Stephen M and Hull, Maury L},
  title = {A new cementless tibial baseplate in kinematically aligned total knee arthroplasty has low migration and stabilizes at 3 months: A cohort study using radiostereometric analysis},
  journal = {Knee Surgery, Sports Traumatology, Arthroscopy},
  year = {2026},
  doi = {10.1002/ksa.70572}
}
Kostenlos starten →