High three‐dimensional accuracy of component placement and lower limb alignment using a robotic arm‐assisted system and gap‐balancing instrument in total knee arthroplasty
Knee Surgery, Sports Traumatology, Arthroscopy · 2024-03 · 32 (3) · 685-692
DOI: 10.1002/ksa.12088 ↗Kurz gefasst
Fachartikel in der Zeitschrift Knee Surgery, Sports Traumatology, Arthroscopy, online erschienen am 28.02.2024. Verfasst von Emanuele Diquattro und 4 weiteren Autor:innen. Beteiligte Einrichtung: Orthopaedic-Traumatology and Prosthetic Surgery and Revisions of Hip and Knee Implants, IRCCS Istituto Ortopedico Rizzoli, Bologna, Italy. Hauptthemen laut MEDLINE: Arthroplasty, Replacement, Knee, Robotic Surgical Procedures, Knee Prosthesis, Osteoarthritis, Knee. Die Arbeit wurde bisher 37-mal zitiert (Crossref) und nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: None.
Abstract (Originalsprache)
Purpose
It was hypothesized that robotic arm‐assisted total knee arthroplasty (RA‐TKA) using additionally a gap‐balancing instrumentation will show high accuracy in executing the planning in femoral and tibial component placement throughout the range of knee motion (ROM) during TKA surgery.
Methods
Prospectively collected data were analysed for patients undergoing RA‐TKA. A cruciate retaining cemented design was implanted using the MAKO® robotic system. Lower limb alignment at 0°, 30°, 45°, 60° and 90° of flexion was recorded at the beginning of surgery and finally after implantation of the components. A ligament tensioner was inserted after tibial precut to measure the extension and flexion gap, and final component placement was planned based on 3D CT images. Femoral and tibial component placement was measured in all three planes.
Results
A total of 104 patients were included (mean age 69.4 ± 9 years; 44 male, 60 female). The difference in component placement after planning and final implantation showed less valgus of 0.7° ± 1.4° ( p < 0.001), less external rotation of 0.6° ± 1.9° ( p = 0.001) and less flexion of 0.9° ± 1.8° ( p < 0.001) for the femoral component. The tibial component was placed in more varus of 0.2° ± 0.9° ( p = 0.056) and more posterior slope of 0.5° ± 0.9° ( p < 0.001). The lower limb alignment in extension was 4.4° ± 5.2° of varus of the native knee and changed to 1.2° ± 1.9° of varus after TKA ( p < 0.01).
Conclusion
Robotic‐assisted TKA helps to achieve the target of alignment and component placement very close to the planning. It allows optimal component placement of off‐the‐shelf implants respecting patient's specific anatomy. Level of Evidence Level II.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Orthopaedic-Traumatology and Prosthetic Surgery and Revisions of Hip and Knee Implants, IRCCS Istituto Ortopedico Rizzoli, Bologna, Italy.
- Department of Orthopaedics and Traumatology, University Hospital Brandenburg/Havel, Brandenburg Medical School Theodor Fontane, Brandenburg an der Havel, Germany.
- Faculty of Health Sciences, Brandenburg Medical School Theodor Fontane, Brandenburg an der Havel, Germany.
Förderung
- None
Zitieren
APA
Diquattro, E., Prill, R., Salzmann, M., Traina, F., & Becker, R. (2024). High three‐dimensional accuracy of component placement and lower limb alignment using a robotic arm‐assisted system and gap‐balancing instrument in total knee arthroplasty. Knee Surgery, Sports Traumatology, Arthroscopy, 32(3), 685-692. https://doi.org/10.1002/ksa.12088
Vancouver
Diquattro E, Prill R, Salzmann M, Traina F, Becker R. High three‐dimensional accuracy of component placement and lower limb alignment using a robotic arm‐assisted system and gap‐balancing instrument in total knee arthroplasty. Knee Surgery, Sports Traumatology, Arthroscopy. 2024;32(3):685-692. doi:10.1002/ksa.12088
BibTeX
@article{Diquattro2024,
author = {Diquattro, Emanuele and Prill, Robert and Salzmann, Mikhail and Traina, Francesco and Becker, Roland},
title = {High three‐dimensional accuracy of component placement and lower limb alignment using a robotic arm‐assisted system and gap‐balancing instrument in total knee arthroplasty},
journal = {Knee Surgery, Sports Traumatology, Arthroscopy},
year = {2024},
volume = {32},
number = {3},
pages = {685--692},
doi = {10.1002/ksa.12088}
}
EUVIMED