Comparison of 3‐Dimensional Ultrasound and Computed Tomography for Imaging of Abdominal Aortic Aneurysms
Journal of Ultrasound in Medicine · 2026-12 · 45 (12) · 2859-2869
DOI: 10.1002/jum.70399 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Journal of Ultrasound in Medicine, online erschienen am 03.08.2026. Verfasst von Rianne E van Rijswijk und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Vascular Surgery, Rijnstate, Arnhem, The Netherlands. Hauptthemen laut MEDLINE: Aortic Aneurysm, Abdominal, Imaging, Three-Dimensional, Tomography, X-Ray Computed. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Stichting Lijf en Leven.
Abstract (Originalsprache)
Objectives
Three‐dimensional ultrasound (3D US) is emerging as a novel imaging method for abdominal aortic aneurysms (AAAs) as it creates a low‐cost 3D AAA overview without harmful radiation and nephrotoxic contrast agents, as opposed to conventional imaging. 3D US can be used with a non‐nephrotoxic contrast agent (3D‐CEUS) for lumen delineation. This study aims to compare 3D US and 3D‐CEUS using a commercially available 3D US system against contrast‐enhanced computed tomography (CTA).
Methods
For twenty AAA patients, the total AAA was segmented for all 3 modalities, and the lumen was segmented in the 3D‐CEUS and CTA scans. The CTA segmentations were registered onto each 3D (CE)US segmentation, and evaluation metrics were computed, including the volumes, average surface distance, sensitivity, precision, and Dice score.
Results
3D US and 3D‐CEUS scans were feasible in all patients. Lumen volumes were significantly larger on 3D‐CEUS compared to CTA (94 versus 70 mL, p < .001). The total AAA volumes were significantly smaller on 3D US compared to CTA (142 versus 165 mL, p = .002), and comparable on 3D‐CEUS (176 versus 165 mL, p = .433). The Dice scores were overall acceptable, ranging from 0.82 to 0.88, and the average surface distances were 2.5 mm.
Conclusions
This study demonstrated the feasibility of 3D US and 3D‐CEUS for AAA imaging with good correspondence to CTA. Addition of the contrast agent increased agreement of volume measurements. 3D US facilitated retrospective offline AAA assessment, which is a major added value compared to 2D US. For now, 3D US already seems a promising imaging modality, which has to be proven in a large‐scale study.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Vascular Surgery, Rijnstate, Arnhem, The Netherlands.
- Multi-Modality Medical Imaging group, Techmed Centre, University of Twente, Enschede, The Netherlands.
- Department of Applied Mathematics, Technical Medical Centre, University of Twente, Enschede, The Netherlands.
ORCID-Profile: Rianne E van Rijswijk, Jelmer M Wolterink, Michel M P J Reijnen, Erik Groot Jebbink
Förderung
- Stichting Lijf en Leven
Zitieren
APA
van Rijswijk, R. E., Everink, S. R. T., van Nie, H. S., Wolterink, J. M., Reijnen, M. M. P. J., & Groot Jebbink, E. (2026). Comparison of 3‐Dimensional Ultrasound and Computed Tomography for Imaging of Abdominal Aortic Aneurysms. Journal of Ultrasound in Medicine, 45(12), 2859-2869. https://doi.org/10.1002/jum.70399
Vancouver
van Rijswijk RE, Everink SRT, van Nie HS, Wolterink JM, Reijnen MMPJ, Groot Jebbink E. Comparison of 3‐Dimensional Ultrasound and Computed Tomography for Imaging of Abdominal Aortic Aneurysms. Journal of Ultrasound in Medicine. 2026;45(12):2859-2869. doi:10.1002/jum.70399
BibTeX
@article{vanRijswijk2026,
author = {van Rijswijk, Rianne E and Everink, Suzan R T and van Nie, Heleen S and Wolterink, Jelmer M and Reijnen, Michel M P J and Groot Jebbink, Erik},
title = {Comparison of 3‐Dimensional Ultrasound and Computed Tomography for Imaging of Abdominal Aortic Aneurysms},
journal = {Journal of Ultrasound in Medicine},
year = {2026},
volume = {45},
number = {12},
pages = {2859--2869},
doi = {10.1002/jum.70399}
}