Use of ultrasonography in assessing peri‐implant osseous defects: An in vitro study
Journal of Periodontology · 26.06.2026
DOI: 10.1002/jper.70160 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Periodontology, online erschienen am 26.06.2026. Verfasst von Melissa Rachel Fok und 4 weiteren Autor:innen. Beteiligte Einrichtung: Division of Periodontology & Implant Dentistry, Faculty of Dentistry, The University of Hong Kong, Hong Kong, China. Die Arbeit nennt 56 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: University of Hong Kong Periodontal Research Fund, Peak Dental Solutions HK Limited (Hong Kong SAR, China).
Abstract (Originalsprache)
Background
The purpose of this study is to investigate the feasibility and diagnostic accuracy of ultrasonography in peri‐implant defect identification and evaluation, and to compare ultrasonography (USG) with cone‐beam computed tomography (CBCT) in defect identification and evaluation in an in vitro model.
Methods
Seventy‐two implants were placed in fresh porcine rib bone models with different types of artificially created defects (including buccal dehiscence, circumferential defect, circumferential defect with buccal dehiscence), and in control sites with no defect. Both USG and CBCT were used to assess the presence, type, and linear measurements (height and depth) of the defects. Diagnostic performance metrics, including the sensitivity, specificity, accuracy, positive predictive value, and negative predictive value of USG and CBCT in characterizing different types of defects, were examined. Linear measurement results obtained by using USG and CBCT were compared with the direct measurement as the gold standard, with the computation of the intraclass correlation coefficient (ICC).
Results
USG demonstrated satisfactory diagnostic accuracy (> 95%) in defect type detection across all types. Its sensitivity in detecting circumferential defects (83.3%) and its positive predictive value for no‐defect sites (87.0%) were lower. CBCT showed perfect diagnostic accuracy (100%) for defect detection. USG illustrated high agreement with direct measurements for defect depth (ICC = 0.934, p < 0.001) but poor agreement for defect height (ICC = 0.349, p < 0.001), underlining its limitations compared with CBCT.
Conclusions
USG is a reliable, non‐ionizing diagnostic tool for characterizing peri‐implant defect types with performance comparable to CBCT. However, it underestimates defect height measurements in defects with infrabony components, where CBCT remains superior.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Division of Periodontology & Implant Dentistry, Faculty of Dentistry, The University of Hong Kong, Hong Kong, China.
- Division of Oral and Maxillofacial Radiology, Faculty of Dentistry, The University of Hong Kong, Hong Kong, China.
ORCID-Profile: Melissa Rachel Fok, Chris Tat Chuen Fok, George Pelekos
Förderung
- University of Hong Kong Periodontal Research Fund
- Peak Dental Solutions HK Limited (Hong Kong SAR, China)
Zitieren
APA
Fok, M. R., Kan, A. Y. T., Fok, C. T. C., Tanaka, R., & Pelekos, G. (2026). Use of ultrasonography in assessing peri‐implant osseous defects: An in vitro study. Journal of Periodontology. https://doi.org/10.1002/jper.70160
Vancouver
Fok MR, Kan AYT, Fok CTC, Tanaka R, Pelekos G. Use of ultrasonography in assessing peri‐implant osseous defects: An in vitro study. Journal of Periodontology. 2026. doi:10.1002/jper.70160
BibTeX
@article{Fok2026,
author = {Fok, Melissa Rachel and Kan, Alice Yee Ting and Fok, Chris Tat Chuen and Tanaka, Ray and Pelekos, George},
title = {Use of ultrasonography in assessing peri‐implant osseous defects: An in vitro study},
journal = {Journal of Periodontology},
year = {2026},
doi = {10.1002/jper.70160}
}
EUVIMED