Improved Accuracy in Pelvic Tumor Resections Using a Real‐Time Vision‐Guided Surgical System

Vahid Danesh iD, Paul Arauz, Maede Boroji, Andrew Zhu, Mia Cottone, Elaine Gould, Fazel A Khan, Imin Kao

Journal of Orthopaedic Research · 2025

DOI: 10.1002/jor.26111 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Orthopaedic Research, 2025. Verfasst von Vahid Danesh und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Mechanical Engineering, Stony Brook University, Stony Brook, New York, USA. Hauptthemen laut MEDLINE: Bone Neoplasms, Pelvic Bones, Surgery, Computer-Assisted. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 33 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Pelvic bone tumor resections remain significantly challenging due to complex three‐dimensional anatomy and limited surgical visualization. While accurate, current navigation systems and patient‐specific instruments present limitations, including high costs, radiation exposure, workflow disruption, long production time, and lack of reusability. This study evaluates a real‐time vision‐guided surgical system combined with modular jigs to improve accuracy in pelvic bone tumor resections. A vision‐guided surgical system combined with modular cutting jigs and real‐time optical tracking was developed and validated. Five male pelvis sawbones were used, with each hemipelvis randomly assigned to either the vision‐guided and modular jig system or the traditional freehand method. A total of 20 resection planes were analyzed for each method. Accuracy was assessed by measuring distance and angular deviations from the planned resection planes. The vision‐guided and modular jig system significantly improved resection accuracy compared to the freehand method, reducing the mean distance deviation from 2.07 ± 1.71 mm to 1.01 ± 0.78 mm ( p = 0.0193). In particular, all specimens resected using the vision‐guided system exhibited errors of less than 3 mm. Angular deviations also showed significant improvements with roll angle deviation reduced from 15.36 ± 17.57° to 4.21 ± 3.46° ( p = 0.0275), and pitch angle deviation decreased from 6.17 ± 4.58° to 1.84 ± 1.48° ( p < 0.001). The proposed vision‐guided and modular jig system significantly improves the accuracy of pelvic bone tumor resections while maintaining workflow efficiency. This cost‐effective solution provides real‐time guidance without the need for referencing external monitors, potentially improving surgical outcomes in complex pelvic bone tumor cases.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Bone NeoplasmsPelvic BonesSurgery, Computer-AssistedHumansMale

Stichwörter der Autor:innen

modular jigspelvic bone tumor resectionresection accuracysurgical navigationvision‐guided surgery

Autor:innen und Institutionen

  • Department of Mechanical Engineering, Stony Brook University, Stony Brook, New York, USA.
  • Department of Orthopedics, Renaissance School of Medicine, Stony Brook University, Stony Brook, New York, USA.
  • Department of Radiology, Renaissance School of Medicine, Stony Brook University, Stony Brook, New York, USA.

ORCID-Profile: Vahid Danesh

Zitieren

APA

Danesh, V., Arauz, P., Boroji, M., Zhu, A., Cottone, M., Gould, E., Khan, F. A., & Kao, I. (2025). Improved Accuracy in Pelvic Tumor Resections Using a Real‐Time Vision‐Guided Surgical System. Journal of Orthopaedic Research. https://doi.org/10.1002/jor.26111

Vancouver

Danesh V, Arauz P, Boroji M, Zhu A, Cottone M, Gould E, et al. Improved Accuracy in Pelvic Tumor Resections Using a Real‐Time Vision‐Guided Surgical System. Journal of Orthopaedic Research. 2025. doi:10.1002/jor.26111

BibTeX
@article{Danesh2025,
  author = {Danesh, Vahid and Arauz, Paul and Boroji, Maede and Zhu, Andrew and Cottone, Mia and Gould, Elaine and Khan, Fazel A and Kao, Imin},
  title = {Improved Accuracy in Pelvic Tumor Resections Using a Real‐Time Vision‐Guided Surgical System},
  journal = {Journal of Orthopaedic Research},
  year = {2025},
  doi = {10.1002/jor.26111}
}
Kostenlos starten →