Advanced Testing Protocols Simulate Failures and Validate Antioxidant Polyethylene in Ankle Implants
Journal of Orthopaedic Research · 2026-01 · 44 (1)
DOI: 10.1002/jor.26103 ↗Kurz gefasst
Validierungsstudie in der Zeitschrift Journal of Orthopaedic Research, online erschienen am 19.05.2025. Verfasst von Ryan Siskey und 5 weiteren Autor:innen. Beteiligte Einrichtung: Exponent Inc., Philadelphia, Pennsylvania, USA. Hauptthemen laut MEDLINE: Polyethylenes, Materials Testing, Prosthesis Failure, Antioxidants. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: This study was internally funded by Enovis..
Abstract (Originalsprache)
Total ankle replacement (TAR) has become an effective treatment for end‐stage ankle osteoarthritis. Multiple factors, including patient characteristics, surgical technique, alignment, and bearing surfaces, influence TAR survivorship. Polyethylene (PE) fatigue is a key consideration in improving outcomes. This study establishes a novel, clinically relevant testing protocol incorporating varus‐valgus rotation to simulate polyethylene fatigue failures observed in mobile‐bearing total ankle replacements. Using this robust methodology, we evaluated the impact of oxidation and antioxidant stabilization on ultrahigh‐molecular‐weight polyethylene (UHMWPE) performance in a mobile bearing implant application. A six‐degree‐of‐freedom simulator was used to iteratively adjust loading parameters (1500–3000 N, −4° to +8° flexion‐extension, ±5° axial rotation, and ±3° or ±8° varus‐valgus rotation at 37 ± 3°C in 20 g/L bovine serum) until clinically observed midline fractures were replicated. Oxidation levels were measured by Fourier‐transform infrared spectroscopy per ASTM F2102. This validated loading protocol was then applied to conventional (25 kGy GUR 1020) and vitamin E‐stabilized (75 kGy GUR 1020‐E) UHMWPE inserts and tested to visible fracture or a 3‐million‐cycle runout. Post‐test fractographic analysis identified crack initiation sites. Conventional aged UHMWPE demonstrated fatigue failure under varus‐valgus rotation (OI = 2.59 ± 1.11) but no failure without rotation. Vitamin E‐stabilized UHMWPE showed no fatigue failure after 3 million cycles, even under varus‐valgus rotation (OI = 0.23 ± 0.02). Fractography revealed fractures originating at the trough and propagating with cyclic loading. Oxidation significantly reduces polyethylene fatigue life, and varus‐valgus rotation exacerbates this effect in mobile bearing TAR implants. Antioxidant‐stabilized UHMWPE showed promising resistance to fatigue and oxidation. These findings support the role of antioxidant stabilization in improving TAR performance, and the protocols developed here provide a framework for assessing the safety of alternative materials.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Exponent Inc., Philadelphia, Pennsylvania, USA.
- School of Biomedical Engineering, Science, and Health Systems, Drexel University, Philadelphia, Pennysylvania, USA.
- Surgical Division, Enovis, Austin, Texas, USA.
- Foot and Ankle Division, Enovis, Atlanta, Georgia, USA.
- Palomar Health Medical Group, San Diego, California, USA.
ORCID-Profile: James Johnson
Förderung
- This study was internally funded by Enovis.
Zitieren
APA
Siskey, R., Hall, P., Heckler, R., Safranski, D., Johnson, J., & Palanca, A. (2026). Advanced Testing Protocols Simulate Failures and Validate Antioxidant Polyethylene in Ankle Implants. Journal of Orthopaedic Research, 44(1). https://doi.org/10.1002/jor.26103
Vancouver
Siskey R, Hall P, Heckler R, Safranski D, Johnson J, Palanca A. Advanced Testing Protocols Simulate Failures and Validate Antioxidant Polyethylene in Ankle Implants. Journal of Orthopaedic Research. 2026;44(1). doi:10.1002/jor.26103
BibTeX
@article{Siskey2026,
author = {Siskey, Ryan and Hall, Patrick and Heckler, Ruth and Safranski, David and Johnson, James and Palanca, Ariel},
title = {Advanced Testing Protocols Simulate Failures and Validate Antioxidant Polyethylene in Ankle Implants},
journal = {Journal of Orthopaedic Research},
year = {2026},
volume = {44},
number = {1},
doi = {10.1002/jor.26103}
}
EUVIMED