Local Diagnostic Reference Levels in Digital Breast Tomosynthesis
Journal of Medical Radiation Sciences · 2025
DOI: 10.1002/jmrs.892 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Medical Radiation Sciences, 2025. Verfasst von Peter Barnes und 3 weiteren Autor:innen. Beteiligte Einrichtung: Austin Health, Repatriation Hospital Campus, Heidelberg, Victoria, Australia. Hauptthemen laut MEDLINE: Mammography, Breast Neoplasms. Die Arbeit nennt 36 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
Introduction
Diagnostic Reference Levels (DRLs) are a guideline to indicate whether the radiation administered during medical procedures performed under routine conditions is unusually low or high. Currently, there are no national DRLs in Australia for Digital Breast Tomosynthesis Mammography (DBT). Given the radiosensitivity of breast tissue, establishing DRLs for DBT is of great importance. This research aims to develop Local Diagnostic Reference Levels (LDRLs) for DBT that can be used as a guide for monitoring radiation delivery levels, and a Quality Assurance tool to assist in monitoring the performance of the mammography unit.
Methods
Average Glandular Dose (AGD) was collected from patient data through OpenREM software. Other data collected included breast density, compression force, compressed breast thickness (CBT), operative and treatment status. LDRLs were set using the 75th percentile (third quartile AGD values), categorised into four CBT ranges and compared against the median AGD values.
Results
LDRLs were similar for both craniocaudal and mediolateral oblique views. LDRLs for CBT ranges of 13–49 mm, 50–74 mm, 75–99 mm and 100–118 mm were calculated to be 1.5, 2.70, 3.90 and 4.70 mGy, respectively. There was no statistically significant correlation between breast density and AGD.
Conclusion
It was found that CBT had a larger impact on dose than density. Significant differences were found between the LDRLs for each of the CBTs; therefore, setting a single LDRL for all CBTs may result in missing unacceptably high or low breast doses. LDRLs should be reviewed frequently to ensure the radiation dose patients receive is within optimal levels.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Austin Health, Repatriation Hospital Campus, Heidelberg, Victoria, Australia.
ORCID-Profile: Michelle Kostidis
Zitieren
APA
Barnes, P., Kostidis, M., Nguyen, J., & O'Donohue, M. (2025). Local Diagnostic Reference Levels in Digital Breast Tomosynthesis. Journal of Medical Radiation Sciences. https://doi.org/10.1002/jmrs.892
Vancouver
Barnes P, Kostidis M, Nguyen J, O'Donohue M. Local Diagnostic Reference Levels in Digital Breast Tomosynthesis. Journal of Medical Radiation Sciences. 2025. doi:10.1002/jmrs.892
BibTeX
@article{Barnes2025,
author = {Barnes, Peter and Kostidis, Michelle and Nguyen, Julie and O'Donohue, Molly},
title = {Local Diagnostic Reference Levels in Digital Breast Tomosynthesis},
journal = {Journal of Medical Radiation Sciences},
year = {2025},
doi = {10.1002/jmrs.892}
}