Size‐specific dose estimates calculated using patient size measurements from scanned projection radiograph in high‐resolution chest computed tomography
Journal of Medical Radiation Sciences · 2025
DOI: 10.1002/jmrs.830 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Medical Radiation Sciences, 2025. Verfasst von Ajit Brindhaban. Beteiligte Einrichtung: Department of Radiologic Sciences, Kuwait University, Sulaibikhat, Kuwait. Hauptthemen laut MEDLINE: Radiation Dosage, Tomography, X-Ray Computed, Radiography, Thoracic, Body Size. Die Arbeit nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Introduction
Size‐specific dose estimates (SSDE) are used to assess patient‐specific radiation exposure in Computed Tomography (CT), complementing the volume CT dose index (CTDIvol). This study compared SSDE calculated using patient's lateral size from scan projection radiograph (SPR) with SSDE calculated using water equivalent diameter ( D w ) from tomographic images in adult chest high‐resolution CT (HRCT).
Methods
In a single‐centre study, the CTDIvol and dose‐length product (DLP) were recorded from HRCT dose reports of adult patients. Lateral width (SLat), at the centre of the scan range, from the SPR was measured and the SSDE (SSDE R ) was calculated using conversion factors related to SLat. Average CT number, area of the slice, and lateral size of the patient (AxLat) were measured on the middle slice. The D w and SSDE from D w (SSDE W ) were calculated. SSDE R and SSDE W were compared using Wilcoxon signed rank test. Correlation between patient size and dosimetry parameters were investigated using Spearman Correlation test with statistical significance at P < 0.05. Bland–Altman plot was also used to test agreement between the two SSDE values.
Results
Median CTDIvol, DLP, SSDE R and SSDE W were 11.0 mGy, 372 mGy.cm, 11.6 mGy and 12.9 mGy, respectively. Small but statistically significant differences ( P < 0.03) were found between SLat and AxLat as well as between SSDE R and SSDE W . Bland–Altman analysis resulted in borderline agreement between SSDE values. Moderate correlations were observed between dosimetry quantities and patient size measurements ( ρ > 0.640; P < 0.001). SSDEw showed statistically significant correlation ( ρ = 0.587 and P < 0.001) with SSDE R .
Conclusion
SSDE R may be used to assess patients' absorbed radiation dose, before the scan, in adult chest HRCT. The median value of SSDE R was about 10% lower than the median value SSDE W . However, the SSDE W should be used after the scan to establish effective dose and radiation risk to the patient.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiologic Sciences, Kuwait University, Sulaibikhat, Kuwait.
ORCID-Profile: Ajit Brindhaban
Zitieren
APA
Brindhaban, A. (2025). Size‐specific dose estimates calculated using patient size measurements from scanned projection radiograph in high‐resolution chest computed tomography. Journal of Medical Radiation Sciences. https://doi.org/10.1002/jmrs.830
Vancouver
Brindhaban A. Size‐specific dose estimates calculated using patient size measurements from scanned projection radiograph in high‐resolution chest computed tomography. Journal of Medical Radiation Sciences. 2025. doi:10.1002/jmrs.830
BibTeX
@article{Brindhaban2025,
author = {Brindhaban, Ajit},
title = {Size‐specific dose estimates calculated using patient size measurements from scanned projection radiograph in high‐resolution chest computed tomography},
journal = {Journal of Medical Radiation Sciences},
year = {2025},
doi = {10.1002/jmrs.830}
}
EUVIMED