Multi‐Site Quality Assurance of ADC and PDFF Measurements: A Multi‐Vendor Phantom Study
Journal of Magnetic Resonance Imaging · 09.11.2026
DOI: 10.1002/jmri.70620 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Magnetic Resonance Imaging, online erschienen am 09.11.2026. Verfasst von Ping Wang und 9 weiteren Autor:innen. Beteiligte Einrichtung: Neuroimaging Innovation Center, Barrow Neurological Institute, Phoenix, Arizona, USA. Die Arbeit nennt 47 Literaturangaben. Gefördert von: NINDS NIH HHS, NIH HHS, U.S. Department of Defense (DOD)/Peer Reviewed Medical Research Program (PRMRP).
Abstract (Originalsprache)
Background
Reliable quantitative imaging biomarkers are important for multi‐site studies. However, scanner‐ and site‐specific variations can compromise the clinical utility of many measures including apparent diffusion coefficient (ADC) and proton density fat fraction (PDFF). Phantom‐based quality assurance provides a foundation for multi‐site studies.
Purpose
To evaluate cross‐site reproducibility, intra‐session reliability, and accuracy of ADC and PDFF measurements across four sites. Study Type Prospective phantom study. Phantoms A NIST diffusion phantom (CaliberMRI, Model 128) and a fat phantom (Calimetrix, Model 300). Field Strength/Sequence 3 T/diffusion‐weighted spin‐echo echo‐planar imaging for ADC and 3D gradient‐echo imaging for PDFF. Assessment ADC and PDFF were measured to assess cross‐site reproducibility and intra‐session reliability. Accuracy was evaluated against reference values. Statistical Tests Intra‐session repeatability was assessed using the intraclass correlation coefficient (ICC). Intra‐session measurement variability was quantified using the coefficient of variation (CoV), and agreement between repeated measurements was assessed using Lin's concordance correlation coefficient (LCCC). NRMSE (normalized root mean square error) assessed accuracy relative to reference values. A linear mixed‐effects (LME) model assessed site differences in ADC and PDFF, with p < 0.05 considered statistically significant, while two one‐sided tests (TOST) assessed equivalence within prespecified margins of ±10% for ADC and ±5% for PDFF.
Results
ICC demonstrated excellent repeatability across repeated scans for both ADC and PDFF (ICC > 0.99). Mean NRMSE between measured and reference values was < 10.0% for ADC and < 7.0% for PDFF, suggesting high accuracy. Mean intra‐session CoV was < 1.0% for ADC and < 2.0% for PDFF, while LCCC exceeded 0.99 for both measurements, demonstrating strong agreement. LME models showed significant site effects for both ADC ( F (3, 36) = 6.05) and PDFF ( F (3, 33) = 5.07), while pairwise differences met the prespecified equivalence criteria for both measures. Data Conclusion Diffusion and fat phantoms provide reliable quality assurance tools for quantitative ADC and PDFF measurements across multiple MRI sites, supporting consistency despite statistically significant site effects. Evidence Level 4. Technical Efficacy Stage 1.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Neuroimaging Innovation Center, Barrow Neurological Institute, Phoenix, Arizona, USA.
- Department of Translational Neuroscience, Barrow Neurological Institute, Phoenix, Arizona, USA.
- Department of Radiology and Imaging Sciences, University of Utah, Salt Lake City, Utah, USA.
- Department of Neurology, Wayne State University, Detroit, Michigan, USA.
- Department of Plastic Surgery, Vanderbilt University Medical Center, Nashville, Tennessee, USA.
- Department of Neurosurgery, University of Utah, Salt Lake City, Utah, USA.
ORCID-Profile: Ping Wang, Yongsheng Chen, Isaac V Manzanera Esteve, Kirstianna R Lombardi, Kewei Chen, Mark A Mahan, Richard D Dortch
Förderung
- NINDS NIH HHS
- NIH HHS
- U.S. Department of Defense (DOD)/Peer Reviewed Medical Research Program (PRMRP)
- U.S. Department of Defense
- National Institutes of Health
Zitieren
APA
Wang, P., Ma, X., Chen, Y., Christopher, K. S., Manzanera Esteve, I. V., Lombardi, K. R., Chen, K., Thayer, W. P., Mahan, M. A., & Dortch, R. D. (2026). Multi‐Site Quality Assurance of ADC and PDFF Measurements: A Multi‐Vendor Phantom Study. Journal of Magnetic Resonance Imaging. https://doi.org/10.1002/jmri.70620
Vancouver
Wang P, Ma X, Chen Y, Christopher KS, Manzanera Esteve IV, Lombardi KR, et al. Multi‐Site Quality Assurance of ADC and PDFF Measurements: A Multi‐Vendor Phantom Study. Journal of Magnetic Resonance Imaging. 2026. doi:10.1002/jmri.70620
BibTeX
@article{Wang2026,
author = {Wang, Ping and Ma, Xiaodong and Chen, Yongsheng and Christopher, Kezia Sharon and Manzanera Esteve, Isaac V and Lombardi, Kirstianna R and Chen, Kewei and Thayer, Wesley P and Mahan, Mark A and Dortch, Richard D},
title = {Multi‐Site Quality Assurance of ADC and PDFF Measurements: A Multi‐Vendor Phantom Study},
journal = {Journal of Magnetic Resonance Imaging},
year = {2026},
doi = {10.1002/jmri.70620}
}
EUVIMED