EUVIMEDEUVIMED entdecken
Crossref

Multi‐Site Quality Assurance of ADC and PDFF Measurements: A Multi‐Vendor Phantom Study

Ping Wang iD, Xiaodong Ma, Yongsheng Chen iD, Kezia Sharon Christopher, Isaac V Manzanera Esteve iD, Kirstianna R Lombardi iD, Kewei Chen iD, Wesley P Thayer, Mark A Mahan iD, Richard D Dortch iD

Journal of Magnetic Resonance Imaging · 09.11.2026

DOI: 10.1002/jmri.70620 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Magnetic Resonance Imaging, online erschienen am 09.11.2026. Verfasst von Ping Wang und 9 weiteren Autor:innen. Beteiligte Einrichtung: Neuroimaging Innovation Center, Barrow Neurological Institute, Phoenix, Arizona, USA. Die Arbeit nennt 47 Literaturangaben. Gefördert von: NINDS NIH HHS, NIH HHS, U.S. Department of Defense (DOD)/Peer Reviewed Medical Research Program (PRMRP).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Reliable quantitative imaging biomarkers are important for multi‐site studies. However, scanner‐ and site‐specific variations can compromise the clinical utility of many measures including apparent diffusion coefficient (ADC) and proton density fat fraction (PDFF). Phantom‐based quality assurance provides a foundation for multi‐site studies.

Purpose

To evaluate cross‐site reproducibility, intra‐session reliability, and accuracy of ADC and PDFF measurements across four sites. Study Type Prospective phantom study. Phantoms A NIST diffusion phantom (CaliberMRI, Model 128) and a fat phantom (Calimetrix, Model 300). Field Strength/Sequence 3 T/diffusion‐weighted spin‐echo echo‐planar imaging for ADC and 3D gradient‐echo imaging for PDFF. Assessment ADC and PDFF were measured to assess cross‐site reproducibility and intra‐session reliability. Accuracy was evaluated against reference values. Statistical Tests Intra‐session repeatability was assessed using the intraclass correlation coefficient (ICC). Intra‐session measurement variability was quantified using the coefficient of variation (CoV), and agreement between repeated measurements was assessed using Lin's concordance correlation coefficient (LCCC). NRMSE (normalized root mean square error) assessed accuracy relative to reference values. A linear mixed‐effects (LME) model assessed site differences in ADC and PDFF, with p < 0.05 considered statistically significant, while two one‐sided tests (TOST) assessed equivalence within prespecified margins of ±10% for ADC and ±5% for PDFF.

Results

ICC demonstrated excellent repeatability across repeated scans for both ADC and PDFF (ICC > 0.99). Mean NRMSE between measured and reference values was < 10.0% for ADC and < 7.0% for PDFF, suggesting high accuracy. Mean intra‐session CoV was < 1.0% for ADC and < 2.0% for PDFF, while LCCC exceeded 0.99 for both measurements, demonstrating strong agreement. LME models showed significant site effects for both ADC ( F (3, 36) = 6.05) and PDFF ( F (3, 33) = 5.07), while pairwise differences met the prespecified equivalence criteria for both measures. Data Conclusion Diffusion and fat phantoms provide reliable quality assurance tools for quantitative ADC and PDFF measurements across multiple MRI sites, supporting consistency despite statistically significant site effects. Evidence Level 4. Technical Efficacy Stage 1.

Schlagwörter

Stichwörter der Autor:innen

diffusion MRIfat fractionfat‐water imagingphantomquality assurance

Autor:innen und Institutionen

  • Neuroimaging Innovation Center, Barrow Neurological Institute, Phoenix, Arizona, USA.
  • Department of Translational Neuroscience, Barrow Neurological Institute, Phoenix, Arizona, USA.
  • Department of Radiology and Imaging Sciences, University of Utah, Salt Lake City, Utah, USA.
  • Department of Neurology, Wayne State University, Detroit, Michigan, USA.
  • Department of Plastic Surgery, Vanderbilt University Medical Center, Nashville, Tennessee, USA.
  • Department of Neurosurgery, University of Utah, Salt Lake City, Utah, USA.

ORCID-Profile: Ping Wang, Yongsheng Chen, Isaac V Manzanera Esteve, Kirstianna R Lombardi, Kewei Chen, Mark A Mahan, Richard D Dortch

Förderung

  • NINDS NIH HHS
  • NIH HHS
  • U.S. Department of Defense (DOD)/Peer Reviewed Medical Research Program (PRMRP)
  • U.S. Department of Defense
  • National Institutes of Health

Zitieren

APA

Wang, P., Ma, X., Chen, Y., Christopher, K. S., Manzanera Esteve, I. V., Lombardi, K. R., Chen, K., Thayer, W. P., Mahan, M. A., & Dortch, R. D. (2026). Multi‐Site Quality Assurance of ADC and PDFF Measurements: A Multi‐Vendor Phantom Study. Journal of Magnetic Resonance Imaging. https://doi.org/10.1002/jmri.70620

Vancouver

Wang P, Ma X, Chen Y, Christopher KS, Manzanera Esteve IV, Lombardi KR, et al. Multi‐Site Quality Assurance of ADC and PDFF Measurements: A Multi‐Vendor Phantom Study. Journal of Magnetic Resonance Imaging. 2026. doi:10.1002/jmri.70620

BibTeX
@article{Wang2026,
  author = {Wang, Ping and Ma, Xiaodong and Chen, Yongsheng and Christopher, Kezia Sharon and Manzanera Esteve, Isaac V and Lombardi, Kirstianna R and Chen, Kewei and Thayer, Wesley P and Mahan, Mark A and Dortch, Richard D},
  title = {Multi‐Site Quality Assurance of ADC and PDFF Measurements: A Multi‐Vendor Phantom Study},
  journal = {Journal of Magnetic Resonance Imaging},
  year = {2026},
  doi = {10.1002/jmri.70620}
}
Kostenlos starten →