Prospective Comparison of DWI ‐Derived Virtual MR Elastography and Conventional MR Elastography in Metabolic Dysfunction‐Associated Steatotic Liver Disease and Healthy Volunteers
Journal of Magnetic Resonance Imaging · 2025
DOI: 10.1002/jmri.70192 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Journal of Magnetic Resonance Imaging, 2025. Verfasst von Anton Volniansky und 13 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Radiation Oncology and Nuclear Medicine, Université de Montréal, Montréal, Canada. Hauptthemen laut MEDLINE: Elasticity Imaging Techniques, Diffusion Magnetic Resonance Imaging, Fatty Liver. Die Arbeit nennt 33 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: Fonds de Recherche du Quebec - Sante, Fonds de recherche du Québec en Santé (FRQS) and Fondation de l'association des radiologistes du Québec (FARQ) Senior Clinical Research Scholarship, Health Research (CIHR)-Institute of Nutrition, Metabolism and Diabetes.
Abstract (Originalsprache)
Background
Virtual MR elastography (VMRE) and MRE have been proposed for liver fibrosis staging in metabolic dysfunction‐associated steatotic liver disease (MASLD), but VMRE's diagnostic performance remains debated.
Purpose
To assess the inter‐visit and inter‐reader reproducibility of fat‐uncorrected and fat‐corrected diffusion‐weighted imaging (DWI)‐based VMRE and to compare their diagnostic performance with MRE for liver fibrosis staging in the MASLD population. Study Type Prospective. Population Fifty four participants were enrolled: 43 with biopsy‐proven MASLD (age: 57.0 ± 9.0 years; 26 males) and 11 healthy volunteers (age: 31.0 ± 15.0 years; 4 males). Field Strength/Sequence 3.0T, DWI ( b ‐values of 0, 200, and 1500 s/mm 2 ) for VMRE and phase‐contrast MRE at 60 Hz was performed. Assessment VMRE‐derived shifted apparent diffusion coefficients (sADC) reproducibility and diagnostic performance; MRE‐derived stiffness diagnostic performance. Statistical Tests Reproducibility was evaluated using intraclass correlation coefficients (ICC), within‐subject coefficient of variation (wCV), and bias and limits of agreement (LOA) in Bland–Altman analysis. Diagnostic performance was assessed with areas under the receiver operating characteristic curve (AUC) and compared with DeLong's test. p < 0.05 was considered statistically significant.
Results
For inter‐visit agreement, the ICC of fat‐uncorrected and fat‐corrected sADC were 0.88 and 0.83; wCV were 0.120 ± 0.30 and 0.141 ± 0.31; bias and 95% LOA were (−0.03 ± 0.18) × 10 −3 mm 2 /s and (−0.05 ± 0.33) × 10 −3 mm 2 /s, respectively. For inter‐reader agreement, the ICC of fat‐uncorrected and fat‐corrected VMRE were 0.99 and 0.99; wCV were 0.028 ± 0.011 and 0.039 ± 0.012, respectively; bias and 95% LOA were (−0.01 ± 0.03) × 10 −3 mm 2 /s and (−0.02 ± 0.05) × 10 −3 mm 2 /s, respectively. AUC of fat‐uncorrected, fat‐corrected sADC, and MRE‐derived stiffness for distinguishing fibrosis stages F0 versus ≥ F1 were 0.70 ± 0.17, 0.56 ± 0.18, and 0.87 ± 0.10; ≤ F1 versus ≥ F2 were 0.61 ± 0.16, 0.49 ± 0.17, and 0.86 ± 0.10; ≤ F2 versus ≥ F3 were 0.54 ± 0.16, 0.50 ± 0.16, and 0.89 ± 0.09; and ≤ F3 versus F4 were 0.58 ± 0.16, 0.55 ± 0.17, and 0.85 ± 0.11, respectively. MRE had significantly higher diagnostic performance than fat‐uncorrected and fat‐corrected VMRE for all fibrosis stages. Data Conclusion VMRE has good reproducibility, but has lower fibrosis staging accuracy than MRE. Evidence Level 1. Technical Efficacy Stage 2.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiology, Radiation Oncology and Nuclear Medicine, Université de Montréal, Montréal, Canada.
- Centre de recherche du Centre hospitalier de l'Université de Montréal (CRCHUM), Montréal, Canada.
- Department of Physics, University of Cambridge, Cambridge, UK.
- Cancer Research UK Cambridge Institute, University of Cambridge, Cambridge, UK.
- Department of Ophthalmology & Visual Sciences, McGill University, Montréal, Canada.
- MR Clinical Science, Philips Healthcare Canada, Mississauga, Canada.
ORCID-Profile: Anton Volniansky, Guillaume Gilbert, Justine Racette
Förderung
- Fonds de Recherche du Quebec - Sante
- Fonds de recherche du Québec en Santé (FRQS) and Fondation de l'association des radiologistes du Québec (FARQ) Senior Clinical Research Scholarship
- Health Research (CIHR)-Institute of Nutrition, Metabolism and Diabetes
- Fondation de l'Association des radiologistes du Québec
- Canadian Institutes of Health Research
Zitieren
APA
Volniansky, A., Lefebvre, T. L., Kulbay, M., Gilbert, G., Fan, B., Racette, J., Montagnon, E., Olivié, D., Sebastiani, G., Giard, J. M., Sylvestre, M. P., Nguyen, B. N., Cloutier, G., & Tang, A. (2025). Prospective Comparison of DWI ‐Derived Virtual MR Elastography and Conventional MR Elastography in Metabolic Dysfunction‐Associated Steatotic Liver Disease and Healthy Volunteers. Journal of Magnetic Resonance Imaging. https://doi.org/10.1002/jmri.70192
Vancouver
Volniansky A, Lefebvre TL, Kulbay M, Gilbert G, Fan B, Racette J, et al. Prospective Comparison of DWI ‐Derived Virtual MR Elastography and Conventional MR Elastography in Metabolic Dysfunction‐Associated Steatotic Liver Disease and Healthy Volunteers. Journal of Magnetic Resonance Imaging. 2025. doi:10.1002/jmri.70192
BibTeX
@article{Volniansky2025,
author = {Volniansky, Anton and Lefebvre, Thierry L and Kulbay, Merve and Gilbert, Guillaume and Fan, Boyan and Racette, Justine and Montagnon, Emmanuel and Olivié, Damien and Sebastiani, Giada and Giard, Jeanne-Marie and Sylvestre, Marie-Pierre and Nguyen, Bich Ngoc and Cloutier, Guy and Tang, An},
title = {Prospective Comparison of DWI ‐Derived Virtual MR Elastography and Conventional MR Elastography in Metabolic Dysfunction‐Associated Steatotic Liver Disease and Healthy Volunteers},
journal = {Journal of Magnetic Resonance Imaging},
year = {2025},
doi = {10.1002/jmri.70192}
}
EUVIMED