MRI Safety Considerations for Permanent Magnet Implants in Muscle
Journal of Magnetic Resonance Imaging · 2026-02 · 63 (2) · 563-573
DOI: 10.1002/jmri.70126 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Magnetic Resonance Imaging, online erschienen am 17.09.2025. Verfasst von Cameron R Taylor und 11 weiteren Autor:innen. Beteiligte Einrichtung: K. Lisa Yang Center for Bionics, Massachusetts Institute of Technology, Cambridge, Massachusetts, USA. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Magnets, Prostheses and Implants, Muscle, Skeletal. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: K. Lisa Yang Center for Bionics at MIT.
Abstract (Originalsprache)
Background
Permanent magnet implants are used with several medical and assistive devices, such as cochlear implants, dental attachments, and prosthetic control, but raise caution for MR imaging. Previous work has evaluated several magnet implants for position and magnetization stability, as well as for image artifacts under MRI. Yet, the intramuscular magnets used for prosthetic control still require evaluation for potential MRI conditionality.
Purpose
To investigate the position and magnetization stability of and image artifacts from 3‐mm‐diameter spherical permanent magnets ( B r = 1.393 T, H ci = 1.637 MA/m) implanted within muscle. Study Type Prospective longitudinal study. Animal Model Porcine; one animal, eight muscles. Field Strength/Sequence 0.55‐T, 1.5‐T/SE, GRE. Assessment Permanent magnets and nonmagnetic controls were implanted into eight muscles and exposed to 1.5‐T MRI 36 days post‐implantation. All sites were examined histologically for evidence of implant migration (acute fibrotic response or fibrotic capsule disruption). Benchtop studies evaluated worst‐case demagnetization and image artifacts (artifact radius minus implant radius). The primary measure of position stability was histological examination interpreting characteristics of progressive skeletal muscle healing. Secondary position stability analysis was performed via CT imaging. Statistical Tests Unpaired one‐sided sign test with a significance level of 0.05. Demagnetization and imaging artifacts were summarized as maximums.
Results
Fibrotic capsules were similarly intact at permanent magnet and control sites (fibrotic capsule thicknesses: 20–550 μm [magnets], 20–220 μm [controls]). No effect of MRI exposure on implant migration was observed via secondary analysis ( p = 0.965 [0.55‐T], p = 0.996 [1.5‐T]). Maximum demagnetization was 2.1% under 0.55‐T exposure and 13.5% under 1.5‐T exposure, and maximum image artifact was 71 mm at both imaging strengths. Data Conclusion The permanent magnet implants used in this study were resistant to migration and substantial demagnetization under 0.55‐T and 1.5‐T MRI exposure and resulted in negligible image artifacts for critical organ imaging, suggesting that the presence of these implants does not preclude a patient from receiving MR imaging up to 1.5T. Evidence Level N/A. Technical Efficacy Stage 5: Improvements in patient care.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- K. Lisa Yang Center for Bionics, Massachusetts Institute of Technology, Cambridge, Massachusetts, USA.
- Lampe Joint Department of Biomedical Engineering, UNC-Chapel Hill and NC State, Chapel Hill, North Carolina, USA.
- MED Institute, West Lafayette, Indiana, USA.
- Cook Research Incorporated, West Lafayette, Indiana, USA.
ORCID-Profile: Cameron R Taylor, David C Gross
Förderung
- K. Lisa Yang Center for Bionics at MIT
Zitieren
APA
Taylor, C. R., Anttila, E. D., Charlebois, S. J., Gross, D. C., Taylor, A. F., Negron-Garcia, J. O., Suckow, C. E., Lyle, L. T., Hooten, S. R., Yeon, S. H., Shallal, C. C., & Herr, H. M. (2026). MRI Safety Considerations for Permanent Magnet Implants in Muscle. Journal of Magnetic Resonance Imaging, 63(2), 563-573. https://doi.org/10.1002/jmri.70126
Vancouver
Taylor CR, Anttila ED, Charlebois SJ, Gross DC, Taylor AF, Negron-Garcia JO, et al. MRI Safety Considerations for Permanent Magnet Implants in Muscle. Journal of Magnetic Resonance Imaging. 2026;63(2):563-573. doi:10.1002/jmri.70126
BibTeX
@article{Taylor2026,
author = {Taylor, Cameron R and Anttila, Eric D and Charlebois, Steven J and Gross, David C and Taylor, Amanda F and Negron-Garcia, Jose O and Suckow, Christopher E and Lyle, L Tiffany and Hooten, Scott R and Yeon, Seong Ho and Shallal, Christopher C and Herr, Hugh M},
title = {MRI Safety Considerations for Permanent Magnet Implants in Muscle},
journal = {Journal of Magnetic Resonance Imaging},
year = {2026},
volume = {63},
number = {2},
pages = {563--573},
doi = {10.1002/jmri.70126}
}
EUVIMED