A Comparative Study of Three Systems for Liver Magnetic Resonance Elastography
Journal of Magnetic Resonance Imaging · 2024
DOI: 10.1002/jmri.29335 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Journal of Magnetic Resonance Imaging, 2024. Verfasst von Runke Wang und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Ruijin Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai Jiao Tong University, Shanghai, China. Hauptthemen laut MEDLINE: Elasticity Imaging Techniques, Phantoms, Imaging, Liver. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 39 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Shanghai Municipal Science and Technology Major Project, National Natural Science Foundation of China, Science and Technology Commission of Shanghai Municipality.
Abstract (Originalsprache)
Background
Different MR elastography (MRE) systems may produce different stiffness measurements, making direct comparison difficult in multi‐center investigations.
Purpose
To assess the repeatability and reproducibility of liver stiffness measured by three typical MRE systems. Study Type Prospective. Population/Phantoms Thirty volunteers without liver disease history (20 males, aged 21–28)/5 gel phantoms. Field Strength/Sequence 3.0 T United Imaging Healthcare (UIH), 1.5 T Siemens Healthcare, 3.0 T General Electric Healthcare (GE)/Echo planar imaging‐based MRE sequence. Assessment Wave images of volunteers and phantoms were acquired by three MRE systems. Tissue stiffness was evaluated by two observers, while phantom stiffness was assessed automatically by code. The reproducibility across three MRE systems was quantified based on the mean stiffness of each volunteer and phantom. Statistical Tests Intraclass correlation coefficients (ICC), coefficients of variation (CV), and Bland–Altman analyses were used to assess the interobserver reproducibility, the interscan repeatability, and the intersystem reproducibility. Paired t ‐tests were performed to assess the interobserver and interscan variation. Friedman tests with Dunn's multiple comparison correction were performed to assess the intersystem variation. P values less than 0.05 indicated significant difference.
Results
The reproducibility of stiffness measured by the two observers demonstrated consistency with ICC > 0.92, CV < 4.32%, Mean bias < 2.23%, and P > 0.06. The repeatability of measurements obtained using the electromagnetic system for the liver revealed ICC > 0.96, CV < 3.86%, Mean bias < 0.19%, P > 0.90. When considering the range of reproducibility across the three systems for liver evaluations, results ranged with ICCs from 0.70 to 0.87, CVs from 6.46% to 10.99%, and Mean biases between 1.89% and 6.30%. Phantom studies showed similar results. The values of measured stiffness differed across all three systems significantly. Data Conclusion Liver stiffness values measured from different MRE systems can be different, but the measurements across the three MRE systems produced consistent results with excellent reproducibility. Evidence Level 1 Technical Efficacy Stage 2
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiology, Ruijin Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai Jiao Tong University, Shanghai, China.
- School of Biomedical Engineering, Shanghai Jiao Tong University, Shanghai, China.
- Institute of Medical Robotics, Shanghai Jiao Tong University, Shanghai, China.
- National Engineering Research Center of Advanced Magnetic Resonance Technologies for Diagnosis and Therapy (NERC-AMRT), Shanghai Jiao Tong University, Shanghai, China.
- Faculty of Medical Imaging Technology, College of Health Science and Technology, Shanghai Jiao Tong University School of Medicine, Shanghai, China.
ORCID-Profile: Runke Wang, Fuhua Yan, Ruokun Li, Yuan Feng
Förderung
- Shanghai Municipal Science and Technology Major Project
- National Natural Science Foundation of China
- Science and Technology Commission of Shanghai Municipality
- National Key Research and Development Program of China
- Natural Science Foundation of Shanghai Municipality
- Medicine-Engineering Interdisciplinary Grant at SJTU
Zitieren
APA
Wang, R., Wang, Y., Qiu, S., Ma, S., Yan, F., Yang, G. Z., Li, R., & Feng, Y. (2024). A Comparative Study of Three Systems for Liver Magnetic Resonance Elastography. Journal of Magnetic Resonance Imaging. https://doi.org/10.1002/jmri.29335
Vancouver
Wang R, Wang Y, Qiu S, Ma S, Yan F, Yang GZ, et al. A Comparative Study of Three Systems for Liver Magnetic Resonance Elastography. Journal of Magnetic Resonance Imaging. 2024. doi:10.1002/jmri.29335
BibTeX
@article{Wang2024,
author = {Wang, Runke and Wang, Yikun and Qiu, Suhao and Ma, Shengyuan and Yan, Fuhua and Yang, Guang-Zhong and Li, Ruokun and Feng, Yuan},
title = {A Comparative Study of Three Systems for Liver Magnetic Resonance Elastography},
journal = {Journal of Magnetic Resonance Imaging},
year = {2024},
doi = {10.1002/jmri.29335}
}
EUVIMED