Intra‐rater and inter‐rater reliability of robotic arthrometer DYNEELAX®
Journal of Experimental Orthopaedics · 2024-07 · 11 (3)
DOI: 10.1002/jeo2.70026 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Experimental Orthopaedics, online erschienen am 30.09.2024. Verfasst von Katja Mihalinec und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Physiotherapy Biomechanical Laboratory, Faculty of Health Sciences, University of Ljubljana Ljubljana Slovenia. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 33 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
The purpose of our study was to test the intra‐rater and inter‐rater reliability of the DYNEELAX® arthrometer in healthy subjects. Since rotational laxity will be measured for the first time in humans, indications for normative values will also be presented.
Methods
Knee anterior laxity and rotational laxity using a DYNEELAX® arthrometer were assessed in 73 subjects (39 females and 34 males). An intraclass correlation coefficient was calculated to evaluate the intra‐rater and inter‐rater reliability of the DYNEELAX®.
Results
An intraclass correlation coefficient for intra‐rater reliability ranges from 0.631 (95% confidence interval; [CI]: 0.47–0.75) to 0.699 (95% CI: 0.56–0.80) and from 0.916 (95% CI: 0.87–0.95) to 0.94 (95% CI: 0.91–0.96) for anterior knee laxity and rotational knee laxity, respectively. An intraclass correlation coefficient for inter‐rater reliability ranges from 0.578 (95% CI: 0.40–0.71) to 0.646 (95% CI: 0.44–0.78) and from 0.822 (95% CI: 0.57–0.91) to 0.933 (95% CI: 0.89–0.96) for anterior knee laxity and rotational knee laxity, respectively.
Conclusions
The DYNEELAX® arthrometer has good to excellent intra‐rater and inter‐rater reliability for rotational knee laxity and moderate intra‐rater reliability for anterior knee laxity in healthy subjects. Future studies should investigate the clinical significance of anterior and rotational laxity measured with the DYNEELAX® arthrometer in patients with knee pathology, as both laxities are critical for assessing the integrity of the intra‐articular structures of the knee in clinical practice. Level of evidence Level IV.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Physiotherapy Biomechanical Laboratory, Faculty of Health Sciences, University of Ljubljana Ljubljana Slovenia.
- Departamento de Fisioterapia Facultad de Medicina, Urbanización Montepríncipe, Universidad San Pablo-CEU, CEU Universities Boadilla del Monte Spain.
- Faculty of Medicine, Institute for Biostatistics and Medical Informatics University of Ljubljana Ljubljana Slovenia.
- Arthron, Institute for Joint and Sports Injuries, Ukmarjeva 2 Ljubljana Slovenia.
ORCID-Profile: Katja Mihalinec, Carmen B Martinez-Cepa, Juan C Zuil-Escobar, Nataša Kejžar, Renata Vauhnik
Zitieren
APA
Mihalinec, K., Martinez-Cepa, C. B., Zuil-Escobar, J. C., Kejžar, N., & Vauhnik, R. (2024). Intra‐rater and inter‐rater reliability of robotic arthrometer DYNEELAX®. Journal of Experimental Orthopaedics, 11(3). https://doi.org/10.1002/jeo2.70026
Vancouver
Mihalinec K, Martinez-Cepa CB, Zuil-Escobar JC, Kejžar N, Vauhnik R. Intra‐rater and inter‐rater reliability of robotic arthrometer DYNEELAX®. Journal of Experimental Orthopaedics. 2024;11(3). doi:10.1002/jeo2.70026
BibTeX
@article{Mihalinec2024,
author = {Mihalinec, Katja and Martinez-Cepa, Carmen B and Zuil-Escobar, Juan C and Kejžar, Nataša and Vauhnik, Renata},
title = {Intra‐rater and inter‐rater reliability of robotic arthrometer DYNEELAX®},
journal = {Journal of Experimental Orthopaedics},
year = {2024},
volume = {11},
number = {3},
doi = {10.1002/jeo2.70026}
}
EUVIMED