Mechanical Performance of Two NiTi Rotary Files: Effects of Alloy and Design of Fatigue, Torsion, Bending, and Fracture Behavior
Microscopy Research and Technique · 2026-04 · 89 (4) · 565-577
DOI: 10.1002/jemt.70093 ↗Kurz gefasst
Fachartikel in der Zeitschrift Microscopy Research and Technique, online erschienen am 14.11.2025. Verfasst von José Aranguren und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Endodontics, Rey Juan Carlos University, Madrid, Spain. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 35 Literaturangaben.
Abstract (Originalsprache)
This study evaluated the mechanical performance of nickel–titanium (NiTi) endodontic instruments with identical tip diameters but different alloy compositions and designs. The study compared two geometric configurations of WaveOne Gold instruments (Respectively, WO I and WO II) manufactured using three different NiTi alloys: SE‐Wire (non‐heat‐treated superelastic NiTi), M‐Wire, and gold alloy. WO I instruments have a variable taper (25/0.08) and dual convex triangular cross‐section, while WO II instruments feature a reduced taper (25/0.07), square cross‐section, and shorter handle. Six groups were tested for cyclic fatigue resistance, torsional strength, bending behavior, deformation, and compressive load. Fracture patterns were examined via scanning electron microscopy (SEM). A two‐way ANOVA evaluated interactions between alloy type and instrument design. Alloy heat treatment and design significantly affected performance. Gold alloy instruments showed superior flexibility and the highest time‐to‐fracture (TTF) values, particularly in WO II. WO II files required less pressure to navigate canals and demonstrated better bending adaptability due to their reduced taper and square cross‐section. In contrast, SE‐Wire instruments had the lowest fatigue resistance and highest pressure loads, indicating their austenitic phase and lack of thermal treatment. WO I instruments had higher torsional strength but fractured at lower angular deflections. SEM revealed distinct fracture mechanisms based on alloy and generation. WO II instruments outperformed WO I in flexibility, cyclic fatigue resistance, and canal tracking, particularly with thermomechanically treated alloys like Gold. The relationship between alloy composition and instrument design is crucial for optimizing performance and minimizing procedural risks.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Endodontics, Rey Juan Carlos University, Madrid, Spain.
- Private Practice, Madrid, Spain.
- Nova Southeastern University College of Dental Medicine, Davie, Florida, USA.
- Department of Clinical Sciences and Stomatology, Università Politecnica delle Marche, Ancona, Italy.
- Department of Medical Biotechnology, Unit of Endodontics and Restorative Dentistry, University of Siena, Siena, Italy.
ORCID-Profile: Alejandro R Pérez, Giulia Malvicini
Zitieren
APA
Aranguren, J., Galvan, G., Bonilla, C., Tosco, V., Pérez, A. R., & Malvicini, G. (2026). Mechanical Performance of Two NiTi Rotary Files: Effects of Alloy and Design of Fatigue, Torsion, Bending, and Fracture Behavior. Microscopy Research and Technique, 89(4), 565-577. https://doi.org/10.1002/jemt.70093
Vancouver
Aranguren J, Galvan G, Bonilla C, Tosco V, Pérez AR, Malvicini G. Mechanical Performance of Two NiTi Rotary Files: Effects of Alloy and Design of Fatigue, Torsion, Bending, and Fracture Behavior. Microscopy Research and Technique. 2026;89(4):565-577. doi:10.1002/jemt.70093
BibTeX
@article{Aranguren2026,
author = {Aranguren, José and Galvan, Gema and Bonilla, Carmen and Tosco, Vincenzo and Pérez, Alejandro R and Malvicini, Giulia},
title = {Mechanical Performance of Two NiTi Rotary Files: Effects of Alloy and Design of Fatigue, Torsion, Bending, and Fracture Behavior},
journal = {Microscopy Research and Technique},
year = {2026},
volume = {89},
number = {4},
pages = {565--577},
doi = {10.1002/jemt.70093}
}
EUVIMED