Performance Evaluation of Automated Urit 5160 Haematology Analyser at the Bamenda Regional Hospital, Cameroon

Victor N Fondoh, Loveline L Niba, Mary B Atanga

Journal of Clinical Laboratory Analysis · 2026-07 · 40 (13)

DOI: 10.1002/jcla.70256 ↗

Kurz gefasst

Übersichtsarbeit in der Zeitschrift Journal of Clinical Laboratory Analysis, online erschienen am 14.05.2026. Verfasst von Victor N Fondoh und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Public Health, Faculty of Health Sciences, The University of Bamenda, Bambili, Cameroon. Hauptthemen laut MEDLINE: Hematologic Tests, Automation, Laboratory. Die Arbeit nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The Urit 5160 is a five‐part leucocyte automated haematology analyser routinely used to improve workflow, patient management, and research. It is essential to confirm the instrument's performance by examining objective evidence that the laboratory can achieve the manufacturer‐provided performance characteristics. Surprisingly, some analysers are used without evaluation. Our objectives were to evaluate the performance characteristics of Urit 5160 using blank count, precision, carryover, linearity, and method‐comparison experiments to improve workflow, patient management, and research. Material and Methods This was a cross‐sectional prospective study conducted between October 2024 and March 2026 at the Bamenda Regional Hospital Laboratory, Cameroon. The parameters examined included: White Blood Cell Count, Haemoglobin, Red Blood Cell Count, Haematocrit, and Platelet Count. Excellent agreement was considered if the results from various experiments matched the manufacturer‐provided specifications. Method comparison was performed by analysing 100 specimens in parallel on Urit 5160 and 3000 plus (reference instrument). Bland–Altman plots and Passing‐Bablok regression analysis were used to evaluate the slopes, intercepts, and bias relative to the total allowable error (TEa) limits.

Results

Our results demonstrated acceptable blank count (0.0%), imprecision (%CV ≤ 3.9), carry‐over (≤ 0.6%), and linearity ( r ≥ 0.999) and the comparison study across all parameters. Bland–Altman analysis and Passing‐Bablok regression showed acceptable correlation and minimal bias.

Conclusion

The instrument demonstrated acceptable blank count, imprecision, carry‐over, linearity, and method comparison within the constraints of the comparator instrument—Urit 3000 plus across all parameters. Thus, laboratories should ensure that manufacturer‐provided characteristics are verified before instruments are routinely used.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Hematologic TestsAutomation, LaboratoryHumansCameroonCross-Sectional StudiesReproducibility of ResultsProspective StudiesLeukocyte Count

Stichwörter der Autor:innen

Urit 5160carryoverevaluation performancelaboratory method verificationlinearitymethod comparisonprecision

Autor:innen und Institutionen

  • Department of Public Health, Faculty of Health Sciences, The University of Bamenda, Bambili, Cameroon.
  • Bamenda Regional Hospital Laboratory, Regional Hospital Bamenda, Bamenda, Cameroon.
  • Department of Medical Laboratory Science, School of Health and Biomedical Science, Catholic University of Cameroon, Bamenda, Cameroon.
  • Department of Nursing and Midwifery, Faculty of Health Sciences, The University of Bamenda, Bambili, Cameroon.

Zitieren

APA

Fondoh, V. N., Niba, L. L., & Atanga, M. B. (2026). Performance Evaluation of Automated Urit 5160 Haematology Analyser at the Bamenda Regional Hospital, Cameroon. Journal of Clinical Laboratory Analysis, 40(13). https://doi.org/10.1002/jcla.70256

Vancouver

Fondoh VN, Niba LL, Atanga MB. Performance Evaluation of Automated Urit 5160 Haematology Analyser at the Bamenda Regional Hospital, Cameroon. Journal of Clinical Laboratory Analysis. 2026;40(13). doi:10.1002/jcla.70256

BibTeX
@article{Fondoh2026,
  author = {Fondoh, Victor N and Niba, Loveline L and Atanga, Mary B},
  title = {Performance Evaluation of Automated Urit 5160 Haematology Analyser at the Bamenda Regional Hospital, Cameroon},
  journal = {Journal of Clinical Laboratory Analysis},
  year = {2026},
  volume = {40},
  number = {13},
  doi = {10.1002/jcla.70256}
}
Kostenlos starten →