Evaluation of the fully automated urine particle analyzer UF‐1500

Yoshifumi Morita iD, Rin Yokoyama, Masami Tanaka, Naru Nakatsuka, Takashi Hisasue, Yoshikazu Ono, Makoto Kurano

Journal of Clinical Laboratory Analysis · 2023

DOI: 10.1002/jcla.24993 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Clinical Laboratory Analysis, 2023. Verfasst von Yoshifumi Morita und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Clinical Laboratory, The University of Tokyo Hospital, Tokyo, Japan. Hauptthemen laut MEDLINE: Urinalysis, Microscopy. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 14 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Sysmex Corporation.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background and Aims This study primarily assessed the performance of the UF‐1500, the novel and compact model of the fully automated urine particle analyzer and evaluated its performance against the existing UF‐5000 instrument. Materials and Methods A total of 648 residual urine specimens were randomly collected and examined using both the UF‐1500 and UF‐5000 instruments as well as manual microscopy. For each parameter, the concordance rates and detection accuracy of the UF‐1500 against manual microscopy were compared with the UF‐5000.

Results

The concordance rates between the UF‐1500 and manual microscopy were 75.3%–98.5%. The UF‐1500 concordance rates within one group agreement were observed to be >90%, for all parameters except for YLCs. The differences within one group agreement between the UF‐1500 and manual microscopy were insignificant, in comparison to the UF‐5000, with exceptions noted for ECs and YLCs. The sensitivity and specificity of the UF‐1500 for RBCs, WBCs, Squa.ECs, and BACT exceeded 80%, while the positive predictive values of ECs and CASTs were below 70%.

Conclusion

The UF‐1500 exhibited a performance that was comparable to the existing instrument, the UF‐5000, and was suitable to be introduced in clinical practice. For the samples with suspected false‐positive or false‐negative results, a manual microscopic examination is required for accurate testing.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

UrinalysisMicroscopyHumansLeukocytesErythrocytesSensitivity and SpecificityUrineFlow Cytometry

Stichwörter der Autor:innen

UF-1500urineurine particle analyzerurine sediment

Autor:innen und Institutionen

  • Department of Clinical Laboratory, The University of Tokyo Hospital, Tokyo, Japan.
  • Department of Clinical Laboratory Medicine, Graduate School of Medicine, The University of Tokyo, Tokyo, Japan.

ORCID-Profile: Yoshifumi Morita

Förderung

  • Sysmex Corporation

Zitieren

APA

Morita, Y., Yokoyama, R., Tanaka, M., Nakatsuka, N., Hisasue, T., Ono, Y., & Kurano, M. (2023). Evaluation of the fully automated urine particle analyzer UF‐1500. Journal of Clinical Laboratory Analysis. https://doi.org/10.1002/jcla.24993

Vancouver

Morita Y, Yokoyama R, Tanaka M, Nakatsuka N, Hisasue T, Ono Y, et al. Evaluation of the fully automated urine particle analyzer UF‐1500. Journal of Clinical Laboratory Analysis. 2023. doi:10.1002/jcla.24993

BibTeX
@article{Morita2023,
  author = {Morita, Yoshifumi and Yokoyama, Rin and Tanaka, Masami and Nakatsuka, Naru and Hisasue, Takashi and Ono, Yoshikazu and Kurano, Makoto},
  title = {Evaluation of the fully automated urine particle analyzer UF‐1500},
  journal = {Journal of Clinical Laboratory Analysis},
  year = {2023},
  doi = {10.1002/jcla.24993}
}
Kostenlos starten →