Using the Sysmex UF ‐4000 urine flow cytometer for rapid diagnosis of urinary tract infection in the clinical microbiological laboratory
Journal of Clinical Laboratory Analysis · 2024
DOI: 10.1002/jcla.25004 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Clinical Laboratory Analysis, 2024. Verfasst von Koos Korsten und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Microbiology and Infection Control, Amsterdam UMC, location AMC, Amsterdam, The Netherlands. Hauptthemen laut MEDLINE: Saccharomyces cerevisiae, Urinary Tract Infections. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
Background
Urinary tract infections are responsible for a significant worldwide disease burden. Performing urine culture is time consuming and labor intensive. Urine flow cytometry might provide a quick and reliable method to screen for urinary tract infection.
Methods
We analyzed routinely collected urine samples received between 2020 and 2022 from both inpatients and outpatients. The UF‐4000 urine flow cytometer was implemented with an optimal threshold for positivity of ≥100 bacteria/μL. We thereafter validated the prognostic value to detect the presence of urinary tract infection (UTI) based on bacterial (BACT), leukocyte (WBC), and yeast‐like cell (YLC) counts combined with the bacterial morphology (UF gram‐flag).
Results
In the first phase, in 2019, the UF‐4000 was implemented using 970 urine samples. In the second phase, between 2020 and 2022, the validation was performed in 42,958 midstream urine samples. The UF‐4000 screen resulted in a 37% ( n = 15,895) decrease in performed urine cultures. Uropathogens were identified in 18,673 (69%) positively flagged urine samples. BACT > 10.000/μL combined with a gram‐negative flag had a >90% positive predictive value for the presence of gram‐negative uropathogens. The absence of gram‐positive flag or YLC had high negative predictive values (99% and >99%, respectively) and are, therefore, best used to rule out the presence of gram‐positive bacteria or yeast. WBC counts did not add to the prediction of uropathogens.
Conclusion
Implementation of the UF‐4000 in routine practice decreased the number of cultured urine samples by 37%. Bacterial cell counts were highly predictive for the presence of UTI, especially when combined with the presence of a gram‐negative flag.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Medical Microbiology and Infection Control, Amsterdam UMC, location AMC, Amsterdam, The Netherlands.
- Department of Medical Microbiology, Jeroen Bosch Hospital, 's Hertogenbosch, The Netherlands.
ORCID-Profile: Koos Korsten
Zitieren
APA
Korsten, K., de Gier, A., Leenders, A., Wever, P. C., & Kolwijck, E. (2024). Using the Sysmex UF ‐4000 urine flow cytometer for rapid diagnosis of urinary tract infection in the clinical microbiological laboratory. Journal of Clinical Laboratory Analysis. https://doi.org/10.1002/jcla.25004
Vancouver
Korsten K, de Gier A, Leenders A, Wever PC, Kolwijck E. Using the Sysmex UF ‐4000 urine flow cytometer for rapid diagnosis of urinary tract infection in the clinical microbiological laboratory. Journal of Clinical Laboratory Analysis. 2024. doi:10.1002/jcla.25004
BibTeX
@article{Korsten2024,
author = {Korsten, Koos and de Gier, Astrid and Leenders, Alexander and Wever, Peter C and Kolwijck, Eva},
title = {Using the Sysmex UF ‐4000 urine flow cytometer for rapid diagnosis of urinary tract infection in the clinical microbiological laboratory},
journal = {Journal of Clinical Laboratory Analysis},
year = {2024},
doi = {10.1002/jcla.25004}
}
EUVIMED