The effect of Xylocaine spray on suture material degradation

Pantea Jannati iD, Charlotte Arp Sørensen, Ditte Gommesen, Marianne Glavind-Kristensen, Peggy Seehafer, Sara Fevre Kindberg, Sarah Hjorth

International Journal of Gynecology & Obstetrics · 2024

DOI: 10.1002/ijgo.15377 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift International Journal of Gynecology & Obstetrics, 2024. Verfasst von Pantea Jannati und 6 weiteren Autor:innen. Beteiligte Einrichtung: Hospital Pharmacy Central Denmark Region, Clinical Pharmacy, Aarhus University Hospital, Aarhus N, Denmark. Hauptthemen laut MEDLINE: Tensile Strength, Sutures, Anesthetics, Local, Lidocaine. Die Arbeit nennt 13 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: Gynzone Denmark.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

To compare the tensile strength of fast absorbable Polyglactin 910 suture material when impregnated with various agents for local anesthesia and to investigate whether the presence of ethanol in Xylocaine spray could explain a potential reduction in tensile strength after use of Xylocaine spray.

Methods

In all, 120 suture samples of Polyglactin 910 were divided into four groups of 30. These four groups were randomly impregnated with isotonic sodium chloride, isotonic sodium chloride plus Xylocaine spray, isotonic sodium chloride plus Xylocaine gel, or isotonic sodium chloride plus ethanol. After impregnation, the sutures were stored in sealed glass tubes in a heating cabinet at 37°C for 72 h. Thereafter, the tensile strength of these 120 samples was assessed by a universal tensile testing machine. The maximal force needed to break the suture material was recorded in newtons (N).

Results

Fast absorbable Polyglactin 910 suture material impregnated with Xylocaine spray or ethanol showed weakened tensile strength (mean values 11.40 and 11.86 N, respectively), whereas the specimens impregnated with Xylocaine gel or sodium chloride retained their tensile strength better (mean values 13.81 and 13.28 N, respectively; mean difference between Xylocaine gel and Xylocaine spray −2.41 N, P < 0.001).

Conclusion

In this in vitro experiment, ethanol and Xylocaine spray weakened the tensile strength of fast absorbable Polyglactin 910 sutures. Use of Xylocaine spray, which contains ethanol, for local anesthesia might lead to early breakdown of the suture material and wound rupture. The authors suggest caution when using Xylocaine spray in combination with fast absorbable Polyglactin 910 suture.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Tensile StrengthSuturesAnesthetics, LocalLidocainePolyglactin 910EthanolMaterials TestingHumansSodium Chloride

Stichwörter der Autor:innen

Polyglactin 910ethanolsuture breakagesuture materialtensile strength

Substanzen

Anesthetics, LocalLidocainePolyglactin 910EthanolSodium Chloride

Autor:innen und Institutionen

  • Hospital Pharmacy Central Denmark Region, Clinical Pharmacy, Aarhus University Hospital, Aarhus N, Denmark.
  • Hospital Pharmacy Central Denmark Region, Research & Development Department and Clinical Pharmacy, Aarhus University Hospital, Aarhus N, Denmark.
  • Institute of Clinical Research, University of Southern Denmark, Odense, Denmark.
  • Department of Gynecology and Obstetrics, Aarhus University Hospital, Aarhus N, Denmark.
  • AnthropologenKontor, Hamburg, Germany.
  • GynZone Denmark, Aarhus C, Denmark.

ORCID-Profile: Pantea Jannati

Förderung

  • Gynzone Denmark

Zitieren

APA

Jannati, P., Sørensen, C. A., Gommesen, D., Glavind-Kristensen, M., Seehafer, P., Kindberg, S. F., & Hjorth, S. (2024). The effect of Xylocaine spray on suture material degradation. International Journal of Gynecology & Obstetrics. https://doi.org/10.1002/ijgo.15377

Vancouver

Jannati P, Sørensen CA, Gommesen D, Glavind-Kristensen M, Seehafer P, Kindberg SF, et al. The effect of Xylocaine spray on suture material degradation. International Journal of Gynecology & Obstetrics. 2024. doi:10.1002/ijgo.15377

BibTeX
@article{Jannati2024,
  author = {Jannati, Pantea and Sørensen, Charlotte Arp and Gommesen, Ditte and Glavind-Kristensen, Marianne and Seehafer, Peggy and Kindberg, Sara Fevre and Hjorth, Sarah},
  title = {The effect of Xylocaine spray on suture material degradation},
  journal = {International Journal of Gynecology & Obstetrics},
  year = {2024},
  doi = {10.1002/ijgo.15377}
}
Kostenlos starten →