FallberichtOpen access · CC BY 4.0

Delayed brain and spine migration of a retained SEEG electrode fragment: An unexpected complication

Manel Krouma, Julia Makhalova iD, Gauthier Toutain, Francesca Pizzo, Stanislas Lagarde, Anne Lepine, Fabrice Bartolomei iD, Didier Scavarda

Epileptic Disorders · 2026-08 · 28 (4) · 1274-1280

DOI: 10.1002/epd2.70230 ↗

At a glance

Fallbericht in Epileptic Disorders, published online on 2026-03-31. Written by Manel Krouma and 7 co-authors. Institution: Department of Pediatric Neurosurgery, La Timone Hospital, Aix-Marseille University, Marseille, France. Major MEDLINE topics: Drug Resistant Epilepsy, Electrodes, Implanted, Foreign-Body Migration, Electroencephalography. 5 references. The full text is freely available (open access, CC BY 4.0).

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Background

Stereoelectroencephalography (SEEG) is a well‐established technique for localizing epileptogenic zones in patients with drug‐resistant epilepsy, including children. While considered safe, rare but serious complications can occur. Case Presentation We report the case of a girl with tuberous sclerosis complex and drug‐resistant epilepsy who underwent SEEG implantation as part of pre‐surgical evaluation. Electrode placement was uneventful. However, during SEEG monitoring, the signal was lost on two electrodes, raising suspicion of fracture. During explantation, these electrodes were not retrieved, as the distal ends were not visible. The removal of the remaining electrodes proceeded without complication. Nearly 3 years later, the patient developed progressive gait instability. Imaging revealed migration of one electrode fragment into the spinal canal at the L4–L5 level, in contact with the cauda equina, and another fragment into the left temporal horn of the lateral ventricle. The spinal fragment was successfully removed surgically without complication.

Conclusion

To the best of our knowledge, this is the first reported case of brain and spinal migration of a retained fragment of a SEEG electrode. This case highlights a rare but potentially significant complication of SEEG implantation and discuss the management of electrode fragments.

Subjects

MeSH terms (MEDLINE)

Drug Resistant EpilepsyElectrodes, ImplantedForeign-Body MigrationElectroencephalographyBrainHumansFemaleHardware Removal

Author keywords

SEEGchilddrug‐resistant epilepsy

Authors and institutions

  • Department of Pediatric Neurosurgery, La Timone Hospital, Aix-Marseille University, Marseille, France.
  • Epileptology Department, APHM, Timone Hospital, Marseille, France.
  • INS, Institute of Systems Neuroscience, Aix-Marseille University, INSERM, Marseille, France.
  • Department of Pediatric Neurology, La Timone Hospital, Aix-Marseille University, Marseille, France.

ORCID profiles: Julia Makhalova, Fabrice Bartolomei

Cite

APA

Krouma, M., Makhalova, J., Toutain, G., Pizzo, F., Lagarde, S., Lepine, A., Bartolomei, F., & Scavarda, D. (2026). Delayed brain and spine migration of a retained SEEG electrode fragment: An unexpected complication. Epileptic Disorders, 28(4), 1274-1280. https://doi.org/10.1002/epd2.70230

Vancouver

Krouma M, Makhalova J, Toutain G, Pizzo F, Lagarde S, Lepine A, et al. Delayed brain and spine migration of a retained SEEG electrode fragment: An unexpected complication. Epileptic Disorders. 2026;28(4):1274-1280. doi:10.1002/epd2.70230

BibTeX
@article{Krouma2026,
  author = {Krouma, Manel and Makhalova, Julia and Toutain, Gauthier and Pizzo, Francesca and Lagarde, Stanislas and Lepine, Anne and Bartolomei, Fabrice and Scavarda, Didier},
  title = {Delayed brain and spine migration of a retained SEEG electrode fragment: An unexpected complication},
  journal = {Epileptic Disorders},
  year = {2026},
  volume = {28},
  number = {4},
  pages = {1274--1280},
  doi = {10.1002/epd2.70230}
}
Start for free →