Detection of seizures with ictal tachycardia, using heart rate variability and patient adaptive logistic regression machine learning methods: A hospital‐based validation study
Epileptic Disorders · 2024-04 · 26 (2) · 199-208
DOI: 10.1002/epd2.20196 ↗Kurz gefasst
Fachartikel in der Zeitschrift Epileptic Disorders, online erschienen am 09.02.2024. Verfasst von Jesper Jeppesen und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Clinical Neurophysiology, Aarhus University Hospital, Aarhus, Denmark. Hauptthemen laut MEDLINE: Seizures, Epilepsies, Partial. Die Arbeit wurde bisher 12-mal zitiert (Crossref) und nennt 28 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: Danmarks Frie Forskningsfond, The Danish Agency for Higher Education and Science (The International Network Programme).
Abstract (Originalsprache)
Objective
Automated seizure detection of focal epileptic seizures is needed for objective seizure quantification to optimize the treatment of patients with epilepsy. Heart rate variability (HRV)‐based seizure detection using patient‐adaptive threshold with logistic regression machine learning (LRML) methods has presented promising performance in a study with a Danish patient cohort. The objective of this study was to assess the generalizability of the novel LRML seizure detection algorithm by validating it in a dataset recorded from long‐term video‐EEG monitoring (LTM) in a Brazilian patient cohort.
Methods
Ictal and inter‐ictal ECG‐data epochs recorded during LTM were analyzed retrospectively. Thirty‐four patients had 107 seizures (79 focal, 28 generalized tonic–clonic [GTC] including focal‐to‐bilateral‐tonic–clonic seizures) eligible for analysis, with a total of 185.5 h recording. Because HRV‐based seizure detection is only suitable in patients with marked ictal autonomic change, patients with >50 beats/min change in heart rate during seizures were selected as responders. The patient‐adaptive LRML seizure detection algorithm was applied to all elected ECG data, and results were computed separately for responders and non‐responders.
Results
The patient‐adaptive LRML seizure detection algorithm yielded a sensitivity of 84.8% (95% CI: 75.6–93.9) with a false alarm rate of .25/24 h in the responder group (22 patients, 59 seizures). Twenty‐five of the 26 GTC seizures were detected (96.2%), and 25 of the 33 focal seizures without bilateral convulsions were detected (75.8%). Significance The study confirms in a new, independent external dataset the good performance of seizure detection from a previous study and suggests that the method is generalizable. This method seems useful for detecting both generalized and focal epileptic seizures. The algorithm can be embedded in a wearable seizure detection system to alert patients and caregivers of seizures and generate objective seizure counts helping to optimize the treatment of the patients.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Clinical Neurophysiology, Aarhus University Hospital, Aarhus, Denmark.
- Department of Clinical Medicine, Aarhus University, Aarhus, Denmark.
- Medical Sciences Post-graduate Program, Federal University of Santa Catarina (UFSC), Florianópolis, SC, Brazil.
- Neurology Division, Federal University of Santa Catarina (UFSC), Florianópolis, SC, Brazil.
- Center for Applied Neurosciences (CeNAp), Federal University of Santa Catarina (UFSC), Florianópolis, SC, Brazil.
- Graduate Program in Neuroscience, UFSC, Florianópolis, SC, Brazil.
ORCID-Profile: Jesper Jeppesen, Hiago Murilo Melo, Jefferson Luiz Brum Marques
Förderung
- Danmarks Frie Forskningsfond
- The Danish Agency for Higher Education and Science (The International Network Programme)
Zitieren
APA
Jeppesen, J., Lin, K., Melo, H. M., Pavei, J., Marques, J. L. B., Beniczky, S., & Walz, R. (2024). Detection of seizures with ictal tachycardia, using heart rate variability and patient adaptive logistic regression machine learning methods: A hospital‐based validation study. Epileptic Disorders, 26(2), 199-208. https://doi.org/10.1002/epd2.20196
Vancouver
Jeppesen J, Lin K, Melo HM, Pavei J, Marques JLB, Beniczky S, et al. Detection of seizures with ictal tachycardia, using heart rate variability and patient adaptive logistic regression machine learning methods: A hospital‐based validation study. Epileptic Disorders. 2024;26(2):199-208. doi:10.1002/epd2.20196
BibTeX
@article{Jeppesen2024,
author = {Jeppesen, Jesper and Lin, Katia and Melo, Hiago Murilo and Pavei, Jonatas and Marques, Jefferson Luiz Brum and Beniczky, Sándor and Walz, Roger},
title = {Detection of seizures with ictal tachycardia, using heart rate variability and patient adaptive logistic regression machine learning methods: A hospital‐based validation study},
journal = {Epileptic Disorders},
year = {2024},
volume = {26},
number = {2},
pages = {199--208},
doi = {10.1002/epd2.20196}
}