Kurz gefasst

Fachartikel in der Zeitschrift Engineering Reports, online erschienen am 21.05.2026. Verfasst von Changhui Li und 2 weiteren Autor:innen. Die Arbeit nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Natural Science Foundation of China.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Oral inhalation therapy has been widely adopted for treating patients with respiratory diseases. Among different inhalation devices, nebulizers are widely used to transform liquid drug formulations into micrometer‐scale droplets. However, the structure–performance relationships of portable pneumatic nebulizers remain insufficiently defined, especially with respect to how local geometric features influence aerosolization characteristics. In this study, nebulizers with different nozzle apertures, gap distances, and outlet lengths were comparatively evaluated to examine the effects of key structural parameters on aerosolization performance. The FPD and MMAD of emitted droplets were sensitive to structural variation, particularly to the gap distance between the nozzle and baffle. Within the tested design space, increasing the outlet length was generally beneficial for improving aerosolization rate and reducing residual liquid in the nebulizer cup. Considering multiple performance indicators together, the 0.4–0.7–5 mm configuration (aperture–gap–outlet) showed a favorable overall balance of FPF, aerosolization rate, and low residual volume among the tested cases. Overall, these findings clarify the structure–performance relationships of the proposed portable pneumatic nebulizer and identify a promising design candidate for further refinement, rather than a fully optimized final geometry.

Förderung

  • National Natural Science Foundation of China

Zitieren

APA

Li, C., Shi, J., & Tong, Z. (2026). Development of Novel Nebulizers: Effects of Key Device Design Parameters on Aerosolization Performance. Engineering Reports, 8(5). https://doi.org/10.1002/eng2.70833

Vancouver

Li C, Shi J, Tong Z. Development of Novel Nebulizers: Effects of Key Device Design Parameters on Aerosolization Performance. Engineering Reports. 2026;8(5). doi:10.1002/eng2.70833