Analysis of drugs seized from amnesty bins at two major United Kingdom summer music festivals using two portable gas chromatography‐mass spectrometry (GC–MS) instruments
Drug Testing and Analysis · 2024-11 · 16 (11) · 1241-1253
DOI: 10.1002/dta.3629 ↗At a glance
Validierungsstudie in Drug Testing and Analysis, published online on 2024-01-23. Written by Anca Frinculescu and 5 co-authors. Institution: Department of Analytical, Environmental and Forensic Sciences, King's College London, London, UK. Major MEDLINE topics: Gas Chromatography-Mass Spectrometry, Illicit Drugs, Substance Abuse Detection, Ketamine. Cited 19 times so far (Crossref); 51 references. The full text is freely available (open access, CC BY 4.0).
Abstract (original language)
Globally, the number of drug users and the proportion of the drug using population has increased from 210 million in 2009 to 269 million in 2019. Several studies suggest that music festival attendees are more likely to abuse illicit substances and have a high‐risk profile. Consequently, it is crucial to develop robust field drug analysis methods that facilitate harm reduction and drug monitoring. The work presented in this report aimed at developing and validating qualitative analytical methods for 3,4‐methylenedioxymethamphetamine, 4‐bromo‐2,5‐dimethoxyphenethylamine (2C‐B), ketamine and N ‐ethylpentylone on two portable gas chromatography–mass spectrometry (GC–MS) systems: Griffin G510 (Teledyne FLIR, West Lafayette, IN) and Torion T‐9 (PerkinElmer, Shelton, CT). The diagnostic ability of the mobile GC–MS units was assessed on 200 samples in total, seized at two large summer music festivals in the United Kingdom. The method validation process included selectivity/specificity, limit of identification, carry‐over, ruggedness/robustness, and inter‐ and intra‐day precision (repeatability and reproducibility). The Griffin G510 demonstrated a limit of identification from 1 mg/mL for 2C‐B to 0.063 mg/mL for ketamine and good ruggedness and precision results. The precision for 2C‐B using the Torion T‐9 was poorer than for the Griffin G510, but equivalent for the other compounds tested. Correct identifications (versus benchtop GC–MS) for the two festivals were 85%–86% and 74%–83% for the Griffin G510 and the Torion T‐9, respectively. The two portable instruments were able to adequately cover current on‐site drug‐testing analytical gaps and proved to be a powerful addition to the on‐site drug analysis techniques.
Subjects
MeSH terms (MEDLINE)
Author keywords
Substances
Authors and institutions
- Department of Analytical, Environmental and Forensic Sciences, King's College London, London, UK.
- TICTAC Communications Ltd., St. George's University of London, London, UK.
- Analytical Services International, St. George's University of London, London, UK.
ORCID profiles: Anca Frinculescu, Lewis Couchman, Vincenzo Abbate
Cite
APA
Frinculescu, A., Shine, T., Ramsey, J., Couchman, L., Frascione, N., & Abbate, V. (2024). Analysis of drugs seized from amnesty bins at two major United Kingdom summer music festivals using two portable gas chromatography‐mass spectrometry (GC–MS) instruments. Drug Testing and Analysis, 16(11), 1241-1253. https://doi.org/10.1002/dta.3629
Vancouver
Frinculescu A, Shine T, Ramsey J, Couchman L, Frascione N, Abbate V. Analysis of drugs seized from amnesty bins at two major United Kingdom summer music festivals using two portable gas chromatography‐mass spectrometry (GC–MS) instruments. Drug Testing and Analysis. 2024;16(11):1241-1253. doi:10.1002/dta.3629
BibTeX
@article{Frinculescu2024,
author = {Frinculescu, Anca and Shine, Trevor and Ramsey, John and Couchman, Lewis and Frascione, Nunzianda and Abbate, Vincenzo},
title = {Analysis of drugs seized from amnesty bins at two major United Kingdom summer music festivals using two portable gas chromatography‐mass spectrometry (GC–MS) instruments},
journal = {Drug Testing and Analysis},
year = {2024},
volume = {16},
number = {11},
pages = {1241--1253},
doi = {10.1002/dta.3629}
}
EUVIMED