EUVIMEDObjevte EUVIMED
Open access · CC BY 4.0

Electrosurgical knife with the water‐jet function of tip‐type during endoscopic treatment injection

Satoshi Ono iD, Yasuko Kurihara, Fumiya Hirose, Hajime Aoki, Kyohei Mejima, Shun Ito, Shun Osumi, Seika Hatori, Kazushi Fukagawa iD, Shosuke Hosaka, Miho Matsukawa, Ryosuke Kobayashi and 2 more

DEN Open · 2023-04 · 3 (1)

DOI: 10.1002/deo2.165 ↗

At a glance

Journal article in DEN Open, published online on 2022-09-20. Written by Satoshi Ono and 13 co-authors. Institution: Department of Gastroenterology and Gastrointestinal Endoscopy Tokyo Metropolitan Geriatric Medical Center Tokyo Japan. Cited 4 times so far (Crossref); 14 references. The full text is freely available (open access, CC BY 4.0).

AbstraktSubjectsBibliografické údajeCiteRelated studies

Abstract (original language)

Objectives

This study aimed to objectively evaluate the water‐jet‐functioned electrosurgical knife injection performances in a desktop experiment.

Methods

Five types of water‐jet‐functioned electrosurgical knives, including two injection styles of sheath‐type (A: DualKnife J, KD‐655L; B: FlushKnife, DK2620‐J‐B20S; C: Splash M‐Knife, DN‐D2718B; D: ISSEN, SN1650‐20) and tip‐type (E: ORISE ProKnife, M00519361) were evaluated. These knives were compared with an injection needle (Control: SuperGrip 25G) as a control. The injection speed under constant pressure and the injection efficiency for each knife against prepared porcine stomach mucosa were evaluated. The additional clear gel injections using an injection needle were observed using an indigo blue‐colored gel to evaluate the difference between the locations of water‐jet holes.

Results

Four types of knives, except for A, showed significantly higher water‐jet speeds (A: 0.79 ± 0.03 g/20 s, B: 2.56 ± 0.05 g/20 s, C: 3.09 ± 0.06 g/20 s, D: 2.86 ± 0.05 g/20 s, and E: 1.79 ± 0.03 g/20 s) compared to that of the control (1.21 ± 0.03 g/20 s). Meanwhile, significantly higher efficacy of injection was found in the tip‐type water‐jet function knife, second to the injection needle (Control: 37.2% ± 35.5%, A: 20.9% ± 20.2%, B: 1.1% ± 2.2%, C: 6.2% ± 12.6%, D: 12.5% ± 15.6%, and E: 33.3% ± 32.2%). An additional injection experiment revealed that the injection with a piercing tip into the gel could achieve sufficient additional injection inside the stacked clear gel.

Conclusions

The tip‐type water‐jet function electrosurgical knife is preferable for effective submucosal injection during endoscopic treatments.

Subjects

Author keywords

desktop experimentelectrosurgical knifeendoscopic submucosal dissectioninjectionporcine model

Authors and institutions

  • Department of Gastroenterology and Gastrointestinal Endoscopy Tokyo Metropolitan Geriatric Medical Center Tokyo Japan.
  • Department of Gastroenterology Chiba-Nishi General Hospital Chiba Japan.
  • Department of Gastroenterology The University of Tokyo Hospital Tokyo Japan.
  • Department of Clinical Engineering Chiba-Nishi General Hospital Chiba Japan.
  • Department of Surgery Chiba-Nishi General Hospital Chiba Japan.

ORCID profiles: Satoshi Ono, Kazushi Fukagawa

Cite

APA

Ono, S., Kurihara, Y., Hirose, F., Aoki, H., Mejima, K., Ito, S., Osumi, S., Hatori, S., Fukagawa, K., Hosaka, S., Matsukawa, M., Kobayashi, R., Yamazaki, N., & Fujishiro, M. (2023). Electrosurgical knife with the water‐jet function of tip‐type during endoscopic treatment injection. DEN Open, 3(1). https://doi.org/10.1002/deo2.165

Vancouver

Ono S, Kurihara Y, Hirose F, Aoki H, Mejima K, Ito S, et al. Electrosurgical knife with the water‐jet function of tip‐type during endoscopic treatment injection. DEN Open. 2023;3(1). doi:10.1002/deo2.165

BibTeX
@article{Ono2023,
  author = {Ono, Satoshi and Kurihara, Yasuko and Hirose, Fumiya and Aoki, Hajime and Mejima, Kyohei and Ito, Shun and Osumi, Shun and Hatori, Seika and Fukagawa, Kazushi and Hosaka, Shosuke and Matsukawa, Miho and Kobayashi, Ryosuke and Yamazaki, Nobuyoshi and Fujishiro, Mitsuhiro},
  title = {Electrosurgical knife with the water‐jet function of tip‐type during endoscopic treatment injection},
  journal = {DEN Open},
  year = {2023},
  volume = {3},
  number = {1},
  doi = {10.1002/deo2.165}
}
Start for free →