Electrosurgical knife with the water‐jet function of tip‐type during endoscopic treatment injection
DEN Open · 2023-04 · 3 (1)
DOI: 10.1002/deo2.165 ↗Kurz gefasst
Fachartikel in der Zeitschrift DEN Open, online erschienen am 20.09.2022. Verfasst von Satoshi Ono und 13 weiteren Autor:innen. Beteiligte Einrichtung: Department of Gastroenterology and Gastrointestinal Endoscopy Tokyo Metropolitan Geriatric Medical Center Tokyo Japan. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 14 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Objectives
This study aimed to objectively evaluate the water‐jet‐functioned electrosurgical knife injection performances in a desktop experiment.
Methods
Five types of water‐jet‐functioned electrosurgical knives, including two injection styles of sheath‐type (A: DualKnife J, KD‐655L; B: FlushKnife, DK2620‐J‐B20S; C: Splash M‐Knife, DN‐D2718B; D: ISSEN, SN1650‐20) and tip‐type (E: ORISE ProKnife, M00519361) were evaluated. These knives were compared with an injection needle (Control: SuperGrip 25G) as a control. The injection speed under constant pressure and the injection efficiency for each knife against prepared porcine stomach mucosa were evaluated. The additional clear gel injections using an injection needle were observed using an indigo blue‐colored gel to evaluate the difference between the locations of water‐jet holes.
Results
Four types of knives, except for A, showed significantly higher water‐jet speeds (A: 0.79 ± 0.03 g/20 s, B: 2.56 ± 0.05 g/20 s, C: 3.09 ± 0.06 g/20 s, D: 2.86 ± 0.05 g/20 s, and E: 1.79 ± 0.03 g/20 s) compared to that of the control (1.21 ± 0.03 g/20 s). Meanwhile, significantly higher efficacy of injection was found in the tip‐type water‐jet function knife, second to the injection needle (Control: 37.2% ± 35.5%, A: 20.9% ± 20.2%, B: 1.1% ± 2.2%, C: 6.2% ± 12.6%, D: 12.5% ± 15.6%, and E: 33.3% ± 32.2%). An additional injection experiment revealed that the injection with a piercing tip into the gel could achieve sufficient additional injection inside the stacked clear gel.
Conclusions
The tip‐type water‐jet function electrosurgical knife is preferable for effective submucosal injection during endoscopic treatments.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Gastroenterology and Gastrointestinal Endoscopy Tokyo Metropolitan Geriatric Medical Center Tokyo Japan.
- Department of Gastroenterology Chiba-Nishi General Hospital Chiba Japan.
- Department of Gastroenterology The University of Tokyo Hospital Tokyo Japan.
- Department of Clinical Engineering Chiba-Nishi General Hospital Chiba Japan.
- Department of Surgery Chiba-Nishi General Hospital Chiba Japan.
ORCID-Profile: Satoshi Ono, Kazushi Fukagawa
Zitieren
APA
Ono, S., Kurihara, Y., Hirose, F., Aoki, H., Mejima, K., Ito, S., Osumi, S., Hatori, S., Fukagawa, K., Hosaka, S., Matsukawa, M., Kobayashi, R., Yamazaki, N., & Fujishiro, M. (2023). Electrosurgical knife with the water‐jet function of tip‐type during endoscopic treatment injection. DEN Open, 3(1). https://doi.org/10.1002/deo2.165
Vancouver
Ono S, Kurihara Y, Hirose F, Aoki H, Mejima K, Ito S, et al. Electrosurgical knife with the water‐jet function of tip‐type during endoscopic treatment injection. DEN Open. 2023;3(1). doi:10.1002/deo2.165
BibTeX
@article{Ono2023,
author = {Ono, Satoshi and Kurihara, Yasuko and Hirose, Fumiya and Aoki, Hajime and Mejima, Kyohei and Ito, Shun and Osumi, Shun and Hatori, Seika and Fukagawa, Kazushi and Hosaka, Shosuke and Matsukawa, Miho and Kobayashi, Ryosuke and Yamazaki, Nobuyoshi and Fujishiro, Mitsuhiro},
title = {Electrosurgical knife with the water‐jet function of tip‐type during endoscopic treatment injection},
journal = {DEN Open},
year = {2023},
volume = {3},
number = {1},
doi = {10.1002/deo2.165}
}
EUVIMED