Fachartikel in der Zeitschrift Cytometry Part A, 2025. Verfasst von Amudhan Krishnaswamy-Usha und 2 weiteren Autor:innen. Beteiligte Einrichtung: Applied and Computational Mathematics Division, National Institute of Standards and Technology, Gaithersburg, Maryland, USA. Hauptthemen laut MEDLINE: Flow Cytometry, Cell Separation. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 24 Literaturangaben. Gefördert von: NIST DOC, National Institute of Standards and Technology.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

In cytometry, it is difficult to disentangle the contributions of population variance and instrument noise toward total measured variation. Fundamentally, this is due to the fact that one cannot measure the same particle multiple times. We propose a simple experiment that uses a cell sorter to distinguish instrument‐specific variation. For a population of beads whose intensities are distributed around a single peak, the sorter is used to collect beads whose measured intensities lie below some threshold. This subset of particles is then remeasured. If the variation in the measured values is only due to the sample, the second set of measurements should also lie entirely below our threshold. Any “spillover” is therefore due to instrument‐specific effects—we demonstrate how the distribution of the post‐sort measurements is sufficient to extract an estimate of the cumulative variability induced by the instrument. A distinguishing feature of our work is that we do not make any assumptions about the sources of said noise. We then show how “local affine transformations” let us transfer these estimates to cytometers not equipped with a sorter. We use our analysis to estimate noise for a set of three instruments and two bead types, across a range of sample flow rates. Lastly, we discuss the implications of instrument noise on optimal classification, as well as other applications.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Flow CytometryCell SeparationHumansUncertainty

Stichwörter der Autor:innen

FACSbeadsgatingsortinguncertainty quantification

Autor:innen und Institutionen

  • Applied and Computational Mathematics Division, National Institute of Standards and Technology, Gaithersburg, Maryland, USA.
  • Chakra Consulting Inc, Clarksburg, Maryland, USA.
  • Microsystems and Nanotechnology Division, National Institute of Standards and Technology, Gaithersburg, Maryland, USA.

ORCID-Profile: Amudhan Krishnaswamy-Usha

Förderung

  • NIST DOC
  • National Institute of Standards and Technology

Zitieren

APA

Krishnaswamy-Usha, A., Cooksey, G. A., & Patrone, P. N. (2025). Uncertainty Quantification in Flow Cytometry Using a Cell Sorter. Cytometry Part A. https://doi.org/10.1002/cyto.a.24925

Vancouver

Krishnaswamy-Usha A, Cooksey GA, Patrone PN. Uncertainty Quantification in Flow Cytometry Using a Cell Sorter. Cytometry Part A. 2025. doi:10.1002/cyto.a.24925

BibTeX
@article{KrishnaswamyUsha2025,
  author = {Krishnaswamy-Usha, Amudhan and Cooksey, Gregory A and Patrone, Paul N},
  title = {Uncertainty Quantification in Flow Cytometry Using a Cell Sorter},
  journal = {Cytometry Part A},
  year = {2025},
  doi = {10.1002/cyto.a.24925}
}
Kostenlos starten →