Effect of Crown to Titanium Base Ratio and Force Angle on the Biomechanical Behavior of Dental Implants With CAD/CAM Zirconia and Lithium Disilicate Crowns: A Finite Element Analysis
Clinical and Experimental Dental Research · 2026
DOI: 10.1002/cre2.70296 ↗Kurz gefasst
Fachartikel in der Zeitschrift Clinical and Experimental Dental Research, 2026. Verfasst von Sara Abtahi und 4 weiteren Autor:innen. Beteiligte Einrichtung: School of Engineering, Faculty of Science and Engineering, Macquarie University, Sydney, New South Wales, Australia. Hauptthemen laut MEDLINE: Zirconium, Crowns, Titanium, Dental Porcelain. Die Arbeit nennt 39 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: the Research Center, Tehran University of Medical Sciences, Tehran, Iran.
Abstract (Originalsprache)
Objectives
This study investigates how titanium‐base (Ti‐base) abutment height, crown design, and force angulation affect biomechanics in an anterior single‐implant restoration, using finite element analysis. Material and Methods A three‐dimensional anterior maxilla model was constructed with linear elastic properties. Two Ti‐base heights (3.5, 5.5 mm) and two crown heights (8, 11 mm) were tested as monolithic zirconia or bilayer (zirconia core veneered with lithium disilicate). A 146 N load was applied at the cingulum at 45° or 65°. von Mises stress (VMS) was computed in the crown, abutment, and surrounding bone.
Results
A 3.5 mm Ti‐base with an 8 mm monolithic zirconia crown (SSZ) produced the lowest crown and abutment VMS. The highest crown VMS occurred in the 11 mm bilayer crown on a 5.5 mm Ti‐base (LLZEX) at 45°, while the highest abutment VMS occurred in the 3.5 mm Ti‐base with an 8 mm bilayer crown (SSZEX) at 65°. An increase in crown height raises crown stresses, whereas the impact of abutment height depends on configuration and angle. In bone, 45° loading increased VMS compared with 65° across all models.
Conclusions
In anterior single‐implant models, the lowest restoration stresses were achieved with a short crown on a 3.5‐mm Ti‐base and monolithic zirconia. Long crowns (11 mm) increased crown stresses, and abutment height should be tailored to material and anticipated loading direction rather than adjusted by a single rule. Oblique loading consistently raised bone stress compared with 65°, underscoring the need to optimize for axial force transmission.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- School of Engineering, Faculty of Science and Engineering, Macquarie University, Sydney, New South Wales, Australia.
- Department of Dental Biomaterials, School of Dentistry, Tehran University of Medical Sciences, Tehran, Iran.
- School of Dentistry, Tehran University of Medical Sciences, Tehran, Iran.
- Department of Orthodontics, School of Dentistry, Tehran University of Medical Sciences, Tehran, Iran.
- Department of Prosthodontics, School of Dentistry, Tehran University of Medical Sciences, Tehran, Iran.
ORCID-Profile: Sara Abtahi, Hakimeh Siadat
Förderung
- the Research Center, Tehran University of Medical Sciences, Tehran, Iran
Zitieren
APA
Abtahi, S., Mirzaboland, S., Geramy, A., Alikhasi, M., & Siadat, H. (2026). Effect of Crown to Titanium Base Ratio and Force Angle on the Biomechanical Behavior of Dental Implants With CAD/CAM Zirconia and Lithium Disilicate Crowns: A Finite Element Analysis. Clinical and Experimental Dental Research. https://doi.org/10.1002/cre2.70296
Vancouver
Abtahi S, Mirzaboland S, Geramy A, Alikhasi M, Siadat H. Effect of Crown to Titanium Base Ratio and Force Angle on the Biomechanical Behavior of Dental Implants With CAD/CAM Zirconia and Lithium Disilicate Crowns: A Finite Element Analysis. Clinical and Experimental Dental Research. 2026. doi:10.1002/cre2.70296
BibTeX
@article{Abtahi2026,
author = {Abtahi, Sara and Mirzaboland, Shamim and Geramy, Allahyar and Alikhasi, Marzieh and Siadat, Hakimeh},
title = {Effect of Crown to Titanium Base Ratio and Force Angle on the Biomechanical Behavior of Dental Implants With CAD/CAM Zirconia and Lithium Disilicate Crowns: A Finite Element Analysis},
journal = {Clinical and Experimental Dental Research},
year = {2026},
doi = {10.1002/cre2.70296}
}
EUVIMED