Spatial sensitivity distribution assessment and M onte C arlo simulations for needle‐based bioimpedance imaging during venipuncture using the finite element method

Ömer Atmaca iD, Jan Liu iD, Toni J Ly iD, Flakë Bajraktari iD, Peter P Pott iD

International Journal for Numerical Methods in Biomedical Engineering · 2024-07 · 40 (7)

DOI: 10.1002/cnm.3831 ↗

Kurz gefasst

Fachartikel in der Zeitschrift International Journal for Numerical Methods in Biomedical Engineering, 2024-05. Verfasst von Ömer Atmaca und 4 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Medical Device Technology (IMT), University of Stuttgart, Baden-Württemberg, Germany. Hauptthemen laut MEDLINE: Finite Element Analysis, Monte Carlo Method, Electric Impedance, Needles. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 72 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Despite being among the most common medical procedures, needle insertions suffer from a high error rate. Impedance measurements using electrode‐equipped needles offer promise for improved tissue targeting and reduced errors. Impedance visualization usually requires an extensive pre‐measured impedance dataset for tissue differentiation and knowledge of the electric fields contributing to the resulting impedances. This work presents two finite element simulation approaches for both problems. The first approach describes the generation of a multitude of impedances with Monte Carlo simulations for both, homogeneous and inhomogeneous tissue to circumvent the need to rely on previously measured data. These datasets could be used for tissue discrimination. The second method describes the simulation of the spatial sensitivity distribution of an electrode layout. Two singularity analysis methods were employed to determine the bulk of the sensitivity within a finite volume, which in turn enables consistent 3D visualization. The modeled electrode layout consists of 12 electrodes radially placed around a hypodermic needle. Electrical excitation was simulated using two neighboring electrodes for current carriage and voltage pickup, which resulted in 12 distinct bipolar excitation states. Both, the impedance simulations and the respective singularity analysis methods were compared with each other. The results show that the statistical spread of impedances is highly dependent on the tissue type and its inhomogeneities. The bounded bulk of sensitivities of both methods are of similar extent and symmetry. Future models should incorporate more detailed tissue properties such as anisotropy or changing material properties due to tissue deformation to gain more accurate predictions.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Finite Element AnalysisMonte Carlo MethodElectric ImpedanceNeedlesHumansComputer SimulationElectrodes

Stichwörter der Autor:innen

Monte Carlofinite element methodimpedancesensitivitysimulationsingularitytissue discrimination

Autor:innen und Institutionen

  • Institute of Medical Device Technology (IMT), University of Stuttgart, Baden-Württemberg, Germany.
  • Institute of Applied Optics (ITO), University of Stuttgart, Stuttgart, Baden-Württemberg, Germany.
  • Fraunhofer Institute for Manufacturing Engineering and Automation (IPA), Stuttgart, Baden-Württemberg, Germany.

ORCID-Profile: Ömer Atmaca, Jan Liu, Toni J Ly, Flakë Bajraktari, Peter P Pott

Zitieren

APA

Atmaca, Ö., Liu, J., Ly, T. J., Bajraktari, F., & Pott, P. P. (2024). Spatial sensitivity distribution assessment and M onte C arlo simulations for needle‐based bioimpedance imaging during venipuncture using the finite element method. International Journal for Numerical Methods in Biomedical Engineering, 40(7). https://doi.org/10.1002/cnm.3831

Vancouver

Atmaca Ö, Liu J, Ly TJ, Bajraktari F, Pott PP. Spatial sensitivity distribution assessment and M onte C arlo simulations for needle‐based bioimpedance imaging during venipuncture using the finite element method. International Journal for Numerical Methods in Biomedical Engineering. 2024;40(7). doi:10.1002/cnm.3831

BibTeX
@article{Atmaca2024,
  author = {Atmaca, Ömer and Liu, Jan and Ly, Toni J and Bajraktari, Flakë and Pott, Peter P},
  title = {Spatial sensitivity distribution assessment and M onte C arlo simulations for needle‐based bioimpedance imaging during venipuncture using the finite element method},
  journal = {International Journal for Numerical Methods in Biomedical Engineering},
  year = {2024},
  volume = {40},
  number = {7},
  doi = {10.1002/cnm.3831}
}