EUVIMEDOpdag EUVIMED
Fallbericht

When the RotaWire Crosses but Cannot Advance: Dual‐Lumen Microcatheter–Assisted RotaWire Delivery

Kazumasa Kurogi iD, Masanobu Ishii, Nobuyasu Yamamoto, Kenichi Tsujita iD

Catheterization and Cardiovascular Interventions · 2026-11-06

DOI: 10.1002/ccd.70998 ↗

At a glance

Fallbericht in Catheterization and Cardiovascular Interventions, published online on 2026-11-06. Written by Kazumasa Kurogi and 3 co-authors. Institution: Department of Cardiology, Miyazaki Prefectural Nobeoka Hospital, Nobeoka, Japan. 5 references.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

RotaWire delivery can be challenging in severely calcified device‐uncrossable coronary lesions when conventional wire exchange through a microcatheter is not feasible. We report a 76‐year‐old man with a severely calcified mid‐left anterior descending coronary artery (LAD) lesion in which neither a microcatheter nor a 0.75 mm balloon could cross. A second channel was created using a penetration wire through a dual‐lumen microcatheter, allowing direct RotaWire crossing. However, the RotaWire could not be advanced sufficiently into the distal LAD. After intentional proximal pullback of the dual‐lumen microcatheter, the microcatheter and RotaWire were held together and advanced as a single unit, enabling adequate distal RotaWire positioning and successful rotational atherectomy. This case highlights a distinct challenge after successful direct RotaWire crossing and demonstrates coupled advancement with a dual‐lumen microcatheter as a potential bailout technique when distal RotaWire advancement remains difficult.

Subjects

Author keywords

RotaWirecalcified coronary lesiondevice‐uncrossable lesiondual‐lumen microcatheterrotational atherectomy

Authors and institutions

  • Department of Cardiology, Miyazaki Prefectural Nobeoka Hospital, Nobeoka, Japan.
  • Department of Cardiovascular Medicine, Graduate School of Medical Sciences, Kumamoto University, Kumamoto, Japan.

ORCID profiles: Kazumasa Kurogi, Kenichi Tsujita

Cite

APA

Kurogi, K., Ishii, M., Yamamoto, N., & Tsujita, K. (2026). When the RotaWire Crosses but Cannot Advance: Dual‐Lumen Microcatheter–Assisted RotaWire Delivery. Catheterization and Cardiovascular Interventions. https://doi.org/10.1002/ccd.70998

Vancouver

Kurogi K, Ishii M, Yamamoto N, Tsujita K. When the RotaWire Crosses but Cannot Advance: Dual‐Lumen Microcatheter–Assisted RotaWire Delivery. Catheterization and Cardiovascular Interventions. 2026. doi:10.1002/ccd.70998

BibTeX
@article{Kurogi2026,
  author = {Kurogi, Kazumasa and Ishii, Masanobu and Yamamoto, Nobuyasu and Tsujita, Kenichi},
  title = {When the RotaWire Crosses but Cannot Advance: Dual‐Lumen Microcatheter–Assisted RotaWire Delivery},
  journal = {Catheterization and Cardiovascular Interventions},
  year = {2026},
  doi = {10.1002/ccd.70998}
}
Start for free →