When the RotaWire Crosses but Cannot Advance: Dual‐Lumen Microcatheter–Assisted RotaWire Delivery
Catheterization and Cardiovascular Interventions · 06.11.2026
DOI: 10.1002/ccd.70998 ↗Kurz gefasst
Fallbericht in der Zeitschrift Catheterization and Cardiovascular Interventions, online erschienen am 06.11.2026. Verfasst von Kazumasa Kurogi und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Miyazaki Prefectural Nobeoka Hospital, Nobeoka, Japan. Die Arbeit nennt 5 Literaturangaben.
Abstract (Originalsprache)
RotaWire delivery can be challenging in severely calcified device‐uncrossable coronary lesions when conventional wire exchange through a microcatheter is not feasible. We report a 76‐year‐old man with a severely calcified mid‐left anterior descending coronary artery (LAD) lesion in which neither a microcatheter nor a 0.75 mm balloon could cross. A second channel was created using a penetration wire through a dual‐lumen microcatheter, allowing direct RotaWire crossing. However, the RotaWire could not be advanced sufficiently into the distal LAD. After intentional proximal pullback of the dual‐lumen microcatheter, the microcatheter and RotaWire were held together and advanced as a single unit, enabling adequate distal RotaWire positioning and successful rotational atherectomy. This case highlights a distinct challenge after successful direct RotaWire crossing and demonstrates coupled advancement with a dual‐lumen microcatheter as a potential bailout technique when distal RotaWire advancement remains difficult.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiology, Miyazaki Prefectural Nobeoka Hospital, Nobeoka, Japan.
- Department of Cardiovascular Medicine, Graduate School of Medical Sciences, Kumamoto University, Kumamoto, Japan.
ORCID-Profile: Kazumasa Kurogi, Kenichi Tsujita
Zitieren
APA
Kurogi, K., Ishii, M., Yamamoto, N., & Tsujita, K. (2026). When the RotaWire Crosses but Cannot Advance: Dual‐Lumen Microcatheter–Assisted RotaWire Delivery. Catheterization and Cardiovascular Interventions. https://doi.org/10.1002/ccd.70998
Vancouver
Kurogi K, Ishii M, Yamamoto N, Tsujita K. When the RotaWire Crosses but Cannot Advance: Dual‐Lumen Microcatheter–Assisted RotaWire Delivery. Catheterization and Cardiovascular Interventions. 2026. doi:10.1002/ccd.70998
BibTeX
@article{Kurogi2026,
author = {Kurogi, Kazumasa and Ishii, Masanobu and Yamamoto, Nobuyasu and Tsujita, Kenichi},
title = {When the RotaWire Crosses but Cannot Advance: Dual‐Lumen Microcatheter–Assisted RotaWire Delivery},
journal = {Catheterization and Cardiovascular Interventions},
year = {2026},
doi = {10.1002/ccd.70998}
}
EUVIMED