Validierungsstudie in Catheterization and Cardiovascular Interventions, published online on 2026-08-05. Written by Mark M P van den Dorpel and 11 co-authors. Institution: Department of Cardiology, Cardiovascular Institute, Thoraxcenter, Erasmus Medical Center, Rotterdam, the Netherlands. Major MEDLINE topics: Multidetector Computed Tomography, Aortic Valve, Transcatheter Aortic Valve Replacement, Radiographic Image Interpretation, Computer-Assisted. 11 references. The full text is freely available (open access, CC BY-NC-ND 4.0).

AbstractSubjectsDati bibliograficiCiteRelated studies

Abstract (original language)

Background

Preprocedural planning with multislice computed tomography (MSCT) is an essential prerequisite for successful transcatheter aortic valve replacement (TAVR). However, preprocedural planning is time‐consuming and may contain clinically meaningful interobserver variability. Artificial intelligence (AI) based automated planning methods may provide anatomical MSCT measurements while reducing the time load required for the analysis. Previous studies on the validation of AI‐based TAVR planning methods excluded anatomical variants or lacked data on actual TAVR prosthesis size agreement. Mimics for TAVR (Materialise, Leuven, Belgium) is a new, AI‐driven TAVR planning platform based on patient‐specific 3D anatomical models and automated measurements through a web‐based user interface.

Aims

To validate the Mimics for TAVR planning tool in clinical practice.

Methods

Key planning assessments were performed on pre‐TAVR MSCT datasets using Mimics for TAVR and a commercially available semi‐automated software package. Agreement between the two planning methods was assessed using intraclass correlation coefficients and Bland‐Altman plots.

Results

A total of 126 consecutive patients (mean age 80 [74–85]; 56% male) who underwent MSCT as part of their TAVR work‐up were included. Mimics for TAVR derived measurements showcased high accuracy and excellent agreement for the area of the aortic annulus (mean error ± SD: −1.9 ± 3.3%, ICC: 0.976 [95% CI: 0.966–0.984]), left ventricular outflow (−2.7 ± 3.8%, ICC: 0.970 [0.957‐0.983]), sinotubular junction (−2.6 ± 4.2%, ICC: 0.953 [0.921–0.971]), coronary heights (LCA −3.5 ± 4.8%, ICC: 0.900 (0.859–0.928). RCA −3.9 ± 4.7%, ICC: 0.928 [0.898–0.949]) and membranous septum length (−5.0 ± 9.2%, ICC: 0.919 [0.880–0.940]). Mimics for TAVR assessment significantly reduced the time required for pre‐TAVR analysis (1.03 ± 0.15 min vs. 9.55 ± 3.16 min per case, respectively ( p = 0.001).

Conclusion

The automated AI‐driven Mimics for TAVR planner is an accurate alternative to existing semiautomated TAVR planning platforms and may enhance overall workflow efficiency.

Subjects

MeSH terms (MEDLINE)

Multidetector Computed TomographyAortic ValveTranscatheter Aortic Valve ReplacementRadiographic Image Interpretation, Computer-AssistedAortic Valve StenosisArtificial IntelligencePatient-Specific ModelingHumansPredictive Value of TestsReproducibility of ResultsAged, 80 and overMaleFemaleAgedObserver VariationHeart Valve ProsthesisProsthesis DesignAutomationTreatment Outcome

Authors and institutions

  • Department of Cardiology, Cardiovascular Institute, Thoraxcenter, Erasmus Medical Center, Rotterdam, the Netherlands.
  • Materialise, Leuven, Belgium.
  • Department of Radiology and Nuclear Medicine, Erasmus Medical Center, Rotterdam, the Netherlands.

ORCID profiles: Mark M P van den Dorpel, Lucas Uchoa de Assis, Joost Daemen

Cite

APA

van den Dorpel, M. M. P., Uchoa de Assis, L., van Niekerk, J., Deckx, S., Sherman, M., Schrot, J., Kozak, A., Kardys, I., Hirsch, A., Daemen, J., Nuis, R. J., & Van Mieghem, N. M. (2026). Validation of an Automated Artificial Intelligence‐Driven Planner for Transcatheter Aortic Valve Replacement. Catheterization and Cardiovascular Interventions. https://doi.org/10.1002/ccd.70785

Vancouver

van den Dorpel MMP, Uchoa de Assis L, van Niekerk J, Deckx S, Sherman M, Schrot J, et al. Validation of an Automated Artificial Intelligence‐Driven Planner for Transcatheter Aortic Valve Replacement. Catheterization and Cardiovascular Interventions. 2026. doi:10.1002/ccd.70785

BibTeX
@article{vandenDorpel2026,
  author = {van den Dorpel, Mark M P and Uchoa de Assis, Lucas and van Niekerk, Jenna and Deckx, Sophie and Sherman, Matthew and Schrot, Janelle and Kozak, Anna and Kardys, Isabella and Hirsch, Alexander and Daemen, Joost and Nuis, Rutger Jan and Van Mieghem, Nicolas M},
  title = {Validation of an Automated Artificial Intelligence‐Driven Planner for Transcatheter Aortic Valve Replacement},
  journal = {Catheterization and Cardiovascular Interventions},
  year = {2026},
  doi = {10.1002/ccd.70785}
}
Start for free →