EUVIMEDScopri EUVIMED
Crossref

Percutaneous Stenting of LVAD Outflow Graft Obstruction: A Multimodality Imaging Case Series

Ahmed Ahmed iD, Fahad Eid, Mohamed Eid, Akshay Joshi, Farhad Abtahian, Abdallah Bitar, Ryan Lydonz, Scott Feitell, Soidjon Khodjaev

Catheterization and Cardiovascular Interventions · 2026-07 · 108 (1) · 26-32

DOI: 10.1002/ccd.70627 ↗

At a glance

Journal article in Catheterization and Cardiovascular Interventions, published online on 2026-04-14. Written by Ahmed Ahmed and 8 co-authors. Institution: Department of Medicine, Division of Cardiology, Rochester General Hospital, Rochester, New York, USA. Major MEDLINE topics: Stents, Endovascular Procedures, Graft Occlusion, Vascular, Heart-Assist Devices. 14 references.

AbstractSubjectsDati bibliograficiCiteRelated studies

Abstract (original language)

Background

Outflow graft obstruction (OGO) is an increasingly recognized complication of continuous‐flow left ventricular assist devices (CF‐LVADs), particularly the HeartMate 3. Surgical revision is often high risk, and percutaneous alternatives are emerging.

Aims

To evaluate the feasibility, safety, and outcomes of endovascular stenting for the treatment of LVAD OGO using a multimodality imaging approach.

Methods

We retrospectively reviewed all HeartMate 3 patients undergoing percutaneous treatment of OGO at our institution between September 2023 and April 2025. Clinical presentation, multimodality imaging findings, procedural characteristics, and outcomes were analyzed.

Results

Eight patients underwent endovascular intervention. All presented with persistent low‐flow alarms and variable syncope or neurologic symptoms. Technical success was achieved in all cases. Median LVAD flow improved from 2.8 L/min pre‐intervention to 4.4 L/min post‐intervention. Two access‐related bleeding complications and one hemothorax occurred, all managed successfully. At a median follow‐up of 7 months, no recurrent OGO was observed.

Conclusions

Endovascular stenting guided by multimodality imaging is a feasible and effective alternative to surgical revision for LVAD OGO in selected patients.

Subjects

MeSH terms (MEDLINE)

StentsEndovascular ProceduresGraft Occlusion, VascularHeart-Assist DevicesMultimodal ImagingVentricular Function, LeftHeart FailureHumansRetrospective StudiesMaleTreatment OutcomeFemaleFeasibility StudiesMiddle AgedTime FactorsPredictive Value of TestsAdultVascular PatencyAgedRecovery of Function

Author keywords

CTAIVUSLVADendovascular stentingoutflow graft obstruction

Authors and institutions

  • Department of Medicine, Division of Cardiology, Rochester General Hospital, Rochester, New York, USA.
  • Department of Medicine, Division of Advanced Heart Failure and Transplant Cardiology, NewYork-Presbyterian Hospital, Weill Cornell Medical Center, New York, New York, USA.
  • Department of Surgery, Division of Vascular Surgery, Rochester General Hospital, Rochester, New York, USA.

ORCID profiles: Ahmed Ahmed

Cite

APA

Ahmed, A., Eid, F., Eid, M., Joshi, A., Abtahian, F., Bitar, A., Lydonz, R., Feitell, S., & Khodjaev, S. (2026). Percutaneous Stenting of LVAD Outflow Graft Obstruction: A Multimodality Imaging Case Series. Catheterization and Cardiovascular Interventions, 108(1), 26-32. https://doi.org/10.1002/ccd.70627

Vancouver

Ahmed A, Eid F, Eid M, Joshi A, Abtahian F, Bitar A, et al. Percutaneous Stenting of LVAD Outflow Graft Obstruction: A Multimodality Imaging Case Series. Catheterization and Cardiovascular Interventions. 2026;108(1):26-32. doi:10.1002/ccd.70627

BibTeX
@article{Ahmed2026,
  author = {Ahmed, Ahmed and Eid, Fahad and Eid, Mohamed and Joshi, Akshay and Abtahian, Farhad and Bitar, Abdallah and Lydonz, Ryan and Feitell, Scott and Khodjaev, Soidjon},
  title = {Percutaneous Stenting of LVAD Outflow Graft Obstruction: A Multimodality Imaging Case Series},