The Terumo Azur Vascular Plug for Occlusion of the Patent Ductus Arteriosus in Premature Neonates—A First Clinical Application
Catheterization and Cardiovascular Interventions · 2026-06 · 107 (7) · 2223-2229
DOI: 10.1002/ccd.70606 ↗Kurz gefasst
Beobachtungsstudie in der Zeitschrift Catheterization and Cardiovascular Interventions, online erschienen am 03.04.2026. Verfasst von Nicholas Fitzgerald und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Paediatrics, Division of Cardiology, The Labatt Family Heart Centre, The Hospital for Sick Children, Ontario, Canada. Hauptthemen laut MEDLINE: Ductus Arteriosus, Patent, Cardiac Catheterization, Infant, Premature. Die Arbeit nennt 19 Literaturangaben.
Abstract (Originalsprache)
Background
Transcatheter patent ductus arteriosus (PDA) occlusion outcomes in premature and low birth weight infants have improved with advances in procedural technique and availability of low‐profile devices. We describe our clinical experience with the Azur Vascular Plug (Terumo Medical Corporation, NJ, USA) for PDA occlusion.
Aims
To assess procedural success, safety and short‐term outcomes of the Azur Vascular Plug for PDA occlusion in premature and low birth weight infants.
Methods
Single‐center observational study of all preterm and low birth weight infants who underwent a transcatheter PDA occlusion procedure involving the Azur Vascular Plug between June 2024 and October 2025.
Results
The Azur Plug was used in 13 infants. The median (range) procedural age was 44 days (17−93 days) and weight 1344 g (855–3900 g). Successful implantation with microcatheter deployment occurred in 12/13 infants. In 11 cases, the 8 mm Azure device was deployed. The ratio of the unconstrained device to the minimal PDA diameter ranged from 2:1 to 4:1 and decreased with experience. The minimal ductal length for deployment was 9.0 mm (lateral angiography). Mean (range) follow‐up was 5.8 months (1.5–15.7 months). Complete ductal occlusion prior to 6‐month follow‐up occurred in 11/12 infants. No aortic obstruction, left pulmonary artery obstruction, or tricuspid valve injury was noted in follow‐up.
Conclusion
The Azur Vascular Plug is a safe and effective option for PDA device occlusion in premature and low birth weight infants. The sizing limits in premature ducts are under evolution. The Azur device appears suitable for ducts with a minimal length of 9.0 mm and, conservatively, a diameter up to 5.0 mm.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Paediatrics, Division of Cardiology, The Labatt Family Heart Centre, The Hospital for Sick Children, Ontario, Canada.
ORCID-Profile: Nicholas Fitzgerald, Lee Benson
Zitieren
APA
Fitzgerald, N., Howell, A., Morgan, C. T., Benson, L., & Chaturvedi, R. R. (2026). The Terumo Azur Vascular Plug for Occlusion of the Patent Ductus Arteriosus in Premature Neonates—A First Clinical Application. Catheterization and Cardiovascular Interventions, 107(7), 2223-2229. https://doi.org/10.1002/ccd.70606
Vancouver
Fitzgerald N, Howell A, Morgan CT, Benson L, Chaturvedi RR. The Terumo Azur Vascular Plug for Occlusion of the Patent Ductus Arteriosus in Premature Neonates—A First Clinical Application. Catheterization and Cardiovascular Interventions. 2026;107(7):2223-2229. doi:10.1002/ccd.70606
BibTeX
@article{Fitzgerald2026,
author = {Fitzgerald, Nicholas and Howell, Alison and Morgan, Conal T and Benson, Lee and Chaturvedi, Rajiv R},
title = {The Terumo Azur Vascular Plug for Occlusion of the Patent Ductus Arteriosus in Premature Neonates—A First Clinical Application},
journal = {Catheterization and Cardiovascular Interventions},
year = {2026},
volume = {107},
number = {7},
pages = {2223--2229},
doi = {10.1002/ccd.70606}
}