Journal article in Catheterization and Cardiovascular Interventions, published online on 2025-12-12. Written by Rayyan Hemetsberger and 11 co-authors. Institution: Department of Cardiology, Internal Medicine II, Medical University of Vienna, Vienna, Austria. Major MEDLINE topics: Drug-Eluting Stents, Coronary Vessels, Percutaneous Coronary Intervention, Angioplasty, Balloon, Coronary. Cited 1 times so far (Crossref); 12 references. The full text is freely available (open access, CC BY-NC-ND 4.0). Funded by: Biotronik, Abbott, and Novomed.
Abstract (original language)
Background
Management of coronary perforation in bifurcation lesions with covered stents remains challenging.
Aims
We aimed to describe a proper practice and toolbox for “jailed‐branch” management after covered stent deployment.
Methods
Using a bench model, we tested various wires with a 120° angulated microcatheter for their ability to penetrate covered stents to access jailed side branches (SB). Next, four PCI techniques were assessed: Test #1 (main branch [MB] rupture simulation): MB covered stent, graft penetration, and kissing balloon inflation (KBI) ( n = 7). Test #2: Test #1 completed with Culotte using a drug‐eluting stent (DES) ( n = 3). Test #3: Culotte with two covered stents simulating MB and SB rupture ( n = 3). Test #4: Test #3 plus final DES in MB ( n = 3).
Results
High tip‐load wires ( > 12 g) successfully penetrated covered stents within 14 s. All procedures were completed successfully. Test #1 showed lumen area reductions of –8% (MB) and –25% (SB), with graft overhang at the SB. Conversion to Culotte with DES (Test #2) improved SB lumen reduction to –8%. Test #3 showed reductions of –7% (MB) and –15% (SB), with graft overhang at MB, corrected in Test #4 by DES implantation, resulting in +9% lumen gain (MB) and improved SB lumen reduction (–11%).
Conclusion
In this bifurcation perforation in‐bench model, high tip‐load wires effectively penetrated covered stents toward excluded branches after crossover covered stent implantation. Conversion to Culotte, using either a DES or a second covered stent, was feasible. Overhanging graft material following SB fenestration could be addressed by overstenting with a DES.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Department of Cardiology, Internal Medicine II, Medical University of Vienna, Vienna, Austria.
- University Heart Center Graz, Medical University Graz, Graz, Austria.
- Cardiology Unit, Sohar Hospital, Sohar, Oman.
- Cardiology Department, Al-Azhar University, Cairo, Egypt.
- Cardiology Department, Amsterdam University Medical Center, University of Amsterdam, Amsterdam, the Netherlands.
- Lenox Hill Hospital, Northwell Health, Donald and Barbara Zucker School of Medicine at Hofstra/Northwell, New York, New York, USA.
ORCID profiles: Anass Maaroufi, Judit Andreka, Mohammad Abdelghani
Funding
- Biotronik, Abbott, and Novomed
Cite
APA
Hemetsberger, R., Maaroufi, A., Hamzaraj, K., Andreka, J., Abdelghani, M., Farhan, S., Lootz, D., Gollmer, J., Mankerious, N., Hengstenberg, C., Johnson, T. W., & Toth, G. G. (2026). Bench Testing Analysis of Perforation Management in the Setting of Bifurcation Coronary Intervention. Catheterization and Cardiovascular Interventions, 107(4), 888-897. https://doi.org/10.1002/ccd.70406
Vancouver
Hemetsberger R, Maaroufi A, Hamzaraj K, Andreka J, Abdelghani M, Farhan S, et al. Bench Testing Analysis of Perforation Management in the Setting of Bifurcation Coronary Intervention. Catheterization and Cardiovascular Interventions. 2026;107(4):888-897. doi:10.1002/ccd.70406
BibTeX
@article{Hemetsberger2026,
author = {Hemetsberger, Rayyan and Maaroufi, Anass and Hamzaraj, Kevin and Andreka, Judit and Abdelghani, Mohammad and Farhan, Serdar and Lootz, Daniel and Gollmer, Johannes and Mankerious, Nader and Hengstenberg, Christian and Johnson, Tom W and Toth, Gabor G},
title = {Bench Testing Analysis of Perforation Management in the Setting of Bifurcation Coronary Intervention},
journal = {Catheterization and Cardiovascular Interventions},
year = {2026},
volume = {107},
number = {4},
pages = {888--897},
doi = {10.1002/ccd.70406}
}
EUVIMED