Fallbericht

Balloon Over‐the‐Wire Opposing Support (BOOST) Technique: The Sawah‐Sabry Approach for Retrieval of a Stuck Retrograde Microcatheter in Chronic Total Occlusion PCI—A Case Report

Ahmed Mohsen Elsawah iD, Mohamed Sabry Elhadainy

Catheterization and Cardiovascular Interventions · 2026

DOI: 10.1002/ccd.70320 ↗

At a glance

Fallbericht in Catheterization and Cardiovascular Interventions, 2026. Written by Ahmed Mohsen Elsawah and 1 co-authors. Institution: Cardiology Department, National Heart Institute, Giza, Egypt. Major MEDLINE topics: Coronary Occlusion, Cardiac Catheters, Angioplasty, Balloon, Coronary, Device Removal. 10 references. Funded by: The authors received no specific funding for this work..

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Chronic total occlusion (CTO) percutaneous coronary intervention (PCI) remains one of the most technically demanding areas in interventional cardiology, and although the retrograde approach has improved procedural success, rare complications such as retrograde microcatheter (MC) entrapment present a serious challenge with limited bailout options. We report the case of a 67‐year‐old male farmer with diabetes mellitus and hypertension who presented with refractory typical chest pain and a failed right coronary artery CTO PCI 3 months earlier. His J‐CTO score was 4, with a prior failed attempt, blunt stump, and severe angulation. A retrograde Gaia 2nd wire successfully crossed the distal cap and entered the antegrade guiding catheter; however, the retrograde MC became entrapped at a heavily calcified bend, forming a complete LAD–RCA loop and resisting conventional retrieval maneuvers. A novel bailout strategy, the Balloon Over‐the‐wire Opposing Support (BOOST) Technique—Sawah‐Sabry Approach—was applied using an antegrade 2.5 × 20 mm balloon advanced within the MC to generate forward counterforce, combined with minimal retrograde traction and GuideLiner support. The MC was gradually released and retrieved safely, allowing successful implantation of three drug‐eluting stents with restoration of TIMI III flow.

Subjects

MeSH terms (MEDLINE)

Coronary OcclusionCardiac CathetersAngioplasty, Balloon, CoronaryDevice RemovalPercutaneous Coronary InterventionHumansMaleAgedChronic DiseaseTreatment OutcomeEquipment DesignCoronary Angiography

Author keywords

BOOST techniquechronic total occlusionentrapped microcatheterpercutaneous coronary interventionretrograde approach

Authors and institutions

  • Cardiology Department, National Heart Institute, Giza, Egypt.

ORCID profiles: Ahmed Mohsen Elsawah

Funding

  • The authors received no specific funding for this work.

Cite

APA

Elsawah, A. M., & Elhadainy, M. S. (2026). Balloon Over‐the‐Wire Opposing Support (BOOST) Technique: The Sawah‐Sabry Approach for Retrieval of a Stuck Retrograde Microcatheter in Chronic Total Occlusion PCI—A Case Report. Catheterization and Cardiovascular Interventions. https://doi.org/10.1002/ccd.70320

Vancouver

Elsawah AM, Elhadainy MS. Balloon Over‐the‐Wire Opposing Support (BOOST) Technique: The Sawah‐Sabry Approach for Retrieval of a Stuck Retrograde Microcatheter in Chronic Total Occlusion PCI—A Case Report. Catheterization and Cardiovascular Interventions. 2026. doi:10.1002/ccd.70320

BibTeX
@article{Elsawah2026,
  author = {Elsawah, Ahmed Mohsen and Elhadainy, Mohamed Sabry},
  title = {Balloon Over‐the‐Wire Opposing Support (BOOST) Technique: The Sawah‐Sabry Approach for Retrieval of a Stuck Retrograde Microcatheter in Chronic Total Occlusion PCI—A Case Report},
  journal = {Catheterization and Cardiovascular Interventions},
  year = {2026},
  doi = {10.1002/ccd.70320}
}
Start for free →