Predictors of Variability in Coronary Flow Reserve and Microvascular Resistance Measurement Using the Bolus Thermodilution Technique
Catheterization and Cardiovascular Interventions · 2026
DOI: 10.1002/ccd.70301 ↗At a glance
Multizentrische Studie in Catheterization and Cardiovascular Interventions, 2026. Written by Jason Galo and 12 co-authors. Institution: Medstar Washington Hospital Center, Washington, District of Columbia, USA. Major MEDLINE topics: Cardiac Catheterization, Coronary Artery Disease, Coronary Circulation, Coronary Vessels. Cited 2 times so far (Crossref); 15 references. Funded by: The FlowLab Study was sponsored by Abbott Vascular..
Abstract (original language)
Background
Bolus thermodilution is the most commonly used invasive assessment for coronary microvascular dysfunction (CMD). However, variability in thermodilution transit time measurements (TTM) affects calculated Coronary Flow Reserve (CFR) and Index of Microvascular Resistance (IMR) values. This study determined variables associated with transit time variability and quantified confidence intervals for the range of measured CFR and IMR values.
Methods
The CoroFlow Cardiovascular System (Coroventis, Sweden) and PressureWire™ X Guidewire (Abbott) were used to collect 2541 TTM from 259 vessels at 14 US sites included in the FlowLab Study. Correlations with baseline patient and procedural variables were determined by both linear and mixed linear models. Mixed linear models with the physician or vessel as the random effect were computed. Monte Carlo simulations of CFR and IMR were performed to determine 95% confidence intervals for thresholds of 2.5 for CFR and 25 for IMR.
Results
Mixed linear analysis with vessel as the random effect provided the best model fit ( R 2 = 0.948). In this model, RCA vessel location ( p < 0.001) and pressure sensor distance < 5 cm ( p < 0.005) were significantly associated with transit time. Monte Carlo simulations identified a “gray zone” range of 2.1–3.1 for CFR and 19–31 for IMR.
Conclusions
Minimizing variability in invasive CMD assessment is essential for improving diagnostic accuracy and guiding clinical decision‐making as testing becomes more widespread. In this study, target vessel choice (RCA) and pressure sensor placement were identified as key modifiable sources of transit time variability. Clinically applicable confidence intervals around CFR and IMR values were defined.
Subjects
MeSH terms (MEDLINE)
Authors and institutions
- Medstar Washington Hospital Center, Washington, District of Columbia, USA.
- Kansas Heart Hospital, Wichita, Kansas, USA.
- Section of Cardiovascular Medicine, Yale School of Medicine, New Haven, Connecticut, USA.
- Georgia Heart Institute, Gainesville, Georgia, USA.
- New York Presbyterian Brooklyn Methodist Hospital, New York, New York, USA.
- Cardiovascular Division, Massachusetts General Hospital, Harvard Medical School, Boston, Massachusetts, USA.
ORCID profiles: Jason Galo, Alok Sharma
Funding
- The FlowLab Study was sponsored by Abbott Vascular.
Cite
APA
Galo, J., Maksoud, A., Shah, S., Toleva, O., Kobayashi, Y., Fahed, A. C., Shlofmitz, E., Sharma, A., Buijs, J. O. D., Spinelli, J., Rapoza, R., Bergmark, B., & Hashim, H. (2026). Predictors of Variability in Coronary Flow Reserve and Microvascular Resistance Measurement Using the Bolus Thermodilution Technique. Catheterization and Cardiovascular Interventions. https://doi.org/10.1002/ccd.70301
Vancouver
Galo J, Maksoud A, Shah S, Toleva O, Kobayashi Y, Fahed AC, et al. Predictors of Variability in Coronary Flow Reserve and Microvascular Resistance Measurement Using the Bolus Thermodilution Technique. Catheterization and Cardiovascular Interventions. 2026. doi:10.1002/ccd.70301
BibTeX
@article{Galo2026,
author = {Galo, Jason and Maksoud, Aziz and Shah, Samit and Toleva, Olga and Kobayashi, Yuhei and Fahed, Akl C and Shlofmitz, Evan and Sharma, Alok and Buijs, Jorn Op den and Spinelli, Joia and Rapoza, Richard and Bergmark, Brian and Hashim, Hayder},
title = {Predictors of Variability in Coronary Flow Reserve and Microvascular Resistance Measurement Using the Bolus Thermodilution Technique},
journal = {Catheterization and Cardiovascular Interventions},
year = {2026},
doi = {10.1002/ccd.70301}
}
EUVIMED