Management of Acurate Neo 2 Valve Protrusion Into the Left Ventricle
Catheterization and Cardiovascular Interventions · 2025
DOI: 10.1002/ccd.31708 ↗At a glance
Fallbericht in Catheterization and Cardiovascular Interventions, 2025. Written by Giorgio Quadri and 3 co-authors. Institution: Interventional Cardiology Unit, A.O. Ordine Mauriziano, Turin, Italy. Major MEDLINE topics: Aortic Valve Stenosis, Heart Valve Prosthesis, Aortic Valve Insufficiency, Transcatheter Aortic Valve Replacement. 12 references.
Abstract (original language)
We report the case of an 81‐year‐old woman with severe aortic stenosis and moderate aortic regurgitation who underwent transcatheter aortic valve implantation (TAVR) with an Acurate Neo 2 valve. During the procedure, the valve protruded into the left ventricle, rendering simple retrieval impossible. A novel “balloon‐assisted” technique was successfully employed to centralize and retract the valve into the ascending aorta, allowing for proper deployment without complications. This case highlights a unique complication and its innovative management strategy.
Subjects
MeSH terms (MEDLINE)
Authors and institutions
- Interventional Cardiology Unit, A.O. Ordine Mauriziano, Turin, Italy.
ORCID profiles: Giorgio Quadri, Gianmarco Annibali
Cite
APA
Quadri, G., Scrocca, I., Annibali, G., & Musumeci, G. (2025). Management of Acurate Neo 2 Valve Protrusion Into the Left Ventricle. Catheterization and Cardiovascular Interventions. https://doi.org/10.1002/ccd.31708
Vancouver
Quadri G, Scrocca I, Annibali G, Musumeci G. Management of Acurate Neo 2 Valve Protrusion Into the Left Ventricle. Catheterization and Cardiovascular Interventions. 2025. doi:10.1002/ccd.31708
BibTeX
@article{Quadri2025,
author = {Quadri, Giorgio and Scrocca, Innocenzo and Annibali, Gianmarco and Musumeci, Giuseppe},
title = {Management of Acurate Neo 2 Valve Protrusion Into the Left Ventricle},
journal = {Catheterization and Cardiovascular Interventions},
year = {2025},
doi = {10.1002/ccd.31708}
}